Non-invasive intraocular pressure monitoring with a contact lens engineered with a nanostructured polymeric sensing film

被引:47
|
作者
Laukhin, Vladimir [1 ,2 ]
Sanchez, Irene [3 ,4 ]
Moya, Ana [4 ,5 ]
Laukhina, Elena [1 ,4 ]
Martin, Raul [3 ,4 ,6 ]
Ussa, Fernando [3 ]
Rovira, Concepcio [1 ,4 ]
Guimera, Anton [4 ,5 ]
Villa, Rosa [4 ,5 ]
Aguilo, Jordi [4 ,5 ]
Pastor, Jose-Carlos [3 ]
Veciana, Jaume [1 ,4 ]
机构
[1] CSIC, Inst Ciencia Mat Barcelona, E-08193 Barcelona, Spain
[2] ICREA, Catalunya, Spain
[3] Univ Valladolid, IOBA Eye Inst, Valladolid, Spain
[4] Networking Res Ctr Bioengn Biomat & Nanomed CIBER, Zaragoza, Spain
[5] CSIC, Inst Microelect Barcelona IMB CNM, E-08193 Barcelona, Spain
[6] Univ Valladolid, Sch Optometry, Dept Phys TAO, Valladolid, Spain
关键词
Glaucoma; Pressure sensor; Nanostructured polymer composite; Contact lens; Intraocular pressure; In vitro experiments; OPEN-ANGLE GLAUCOMA; CORNEAL CURVATURE;
D O I
10.1016/j.sna.2011.05.021
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A new prototype of contact lens sensor (CLS) for monitoring the intraocular pressure (IOP) variations in a non-invasive way has been developed. The CLS has as a key element a thin film, comprised of a new all-organic flexible highly piezo-resistive sensor, that is glued to the central hole of a doughnut-shaped hard contact lens. A few CLSs were fabricated and tested with an eye phantom and with enucleated pig eyes by means of applying pressure with a low pressure transducer and recording the electrical changes in the sensor with a portable recorder. The developed CLSs are biocompatible permitting to transmit the changes in cornea curvature induced by the IOP variations, directly to the flexible conducting polymeric sensor embedded in the CL The electrical response of the CLS to pressure changes reveals a high linearity as well as a good reproducibility having the proper sensitivity to perform continuous monitoring of LOP. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:36 / 43
页数:8
相关论文
共 50 条
  • [21] Non-invasive intraocular pressure monitoring system based on LC resonance
    Chen L.
    Li H.
    Liu X.
    Huazhong Keji Daxue Xuebao (Ziran Kexue Ban)/Journal of Huazhong University of Science and Technology (Natural Science Edition), 2024, 52 (05): : 1 - 7
  • [22] Noninvasive Intraocular Pressure Monitorization With a Sensing Contact Lens
    Martin, R., Sr.
    Sanchez, I.
    Ussa, F.
    Pastor, J.
    Veciana, J.
    Rovira, C.
    Laukhin, V.
    Laukhina, E.
    Villa, R.
    Aguilo, J.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [23] A capacitor-based sensor and a contact lens sensing system for intraocular pressure monitoring
    Chiou, Jin-Chern
    Huang, Yu-Chieh
    Yeh, Guan-Ting
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2016, 26 (01)
  • [24] Contact lens as an emerging platform for non-invasive bio-sensing: A review
    Shetty, Kiran H.
    Desai, Ditixa T.
    Patel, Hetal P.
    Shah, Dinesh O.
    Willcox, Mark D. P.
    Maulvi, Furqan A.
    SENSORS AND ACTUATORS A-PHYSICAL, 2024, 376
  • [25] Wearable contact lens sensor for intraocular pressure monitoring
    Fan, Yanyan
    Tu, Hailing
    Zhao, Hongbin
    Wei, Feng
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2021, 128 : 22 - 22
  • [26] Non-invasive Intraocular Pressure Measurements in Zebrafish
    Cameron, D. Joshua
    Davey, Pinakin
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [27] NON-INVASIVE DIGITAL PRESSURE MONITORING
    MELENDEZ, J
    WEISSMAN, C
    DELPHIN, E
    DODDS, T
    WALD, A
    CRITICAL CARE MEDICINE, 1988, 16 (04) : 427 - 427
  • [28] LC contact lens sensor for ultrasensitive intraocular pressure monitoring
    An, Hongbin
    Wang, Xian
    Liao, Zhikang
    Zhang, Liang
    Zhao, Hui
    Yang, Yingyun
    Song, Jizhou
    Ma, Yinji
    NPJ FLEXIBLE ELECTRONICS, 2024, 8 (01)
  • [29] Non-Invasive Intracranial Pressure Monitoring
    Mueller, Sebastian Johannes
    Henkes, Elina
    Gounis, Matthew J.
    Felber, Stephan
    Ganslandt, Oliver
    Henkes, Hans
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (06)
  • [30] Highly Transparent and Sensitive Graphene Sensors for Continuous and Non-invasive Intraocular Pressure Monitoring
    Xu, Jiandong
    Cui, Tianrui
    Hirtz, Thomas
    Qiao, Yancong
    Li, Xiaoshi
    Zhong, Fanhui
    Han, Xiaolin
    Yang, Yi
    Zhang, Sheng
    Ren, Tian-Ling
    ACS APPLIED MATERIALS & INTERFACES, 2020, 12 (16) : 18375 - 18384