Left Ventricular Filling Pressures Contribute to Exercise Limitation in Patients with Continuous Flow Left Ventricular Assist Devices

被引:5
|
作者
Koshy, Aaron [1 ]
Bouzas-Cruz, Noelia [1 ,2 ]
Okwose, Nduka C. [3 ]
Fernandez, Oscar Gonzalez [1 ]
Green, Thomas [1 ]
Woods, Andrew [1 ]
Robinson-Smith, Nicola [1 ]
Tovey, Sian [1 ]
McDiarmid, Adam [1 ]
Parry, Gareth [1 ]
Schueler, Stephan [1 ]
Jakovljevic, Djordje G. [3 ]
MacGowan, Guy A. [1 ,3 ]
机构
[1] Freeman Rd Hosp, Dept Cardiol & Cardiothorac Surg, Newcastle Upon Tyne, Tyne & Wear, England
[2] Univ Santiago Compostela, Galicia, Spain
[3] Newcastle Univ, Inst Cellular Med & Genet Med, Newcastle Upon Tyne, Tyne & Wear, England
关键词
exercise capacity; left ventricular assist device; hemodynamics; PUMP SPEED; CARDIAC-OUTPUT; HEART-FAILURE; DOUBLE-BLIND; IMPACT; PERFORMANCE; AFTERLOAD; PRELOAD; SUPPORT;
D O I
10.1097/MAT.0000000000001073
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We sought to determine hemodynamic mechanisms of exercise intolerance in a group of patients with the HeartWare ventricular assist device (VAD) compared to a group of heart failure patients. Twenty VAD and 22 heart failure patients underwent symptom-limited active straight leg raising exercise during right heart catheterization with thermodilution (TD), and upright cycling cardiopulmonary stress testing with cardiac output measurement by inert gas rebreathing (IGR) method. The TD and IGR exercise cardiac indexes were higher in VAD compared with heart failure group (both P < 0.05), although there was only a borderline increase in peak exercise oxygen consumption (VO2) (P = 0.06). Baseline and exercise right heart catheterization pressures were not significantly different between the two groups. The only significant independent predictors of peak VO2 in the heart failure group were exercise heart rate and cardiac index (both P < 0.05). In contrast, for the VAD group only, resting pulmonary arterial wedge and pulmonary arterial mean pressures were independently related to peak VO2 (both P < 0.05). Thus, in heart failure, exercise cardiac index is an important limitation to exercise capacity, and VADs increase exercise cardiac index. However, in VAD patients, this only produces limited benefits as increased pulmonary and pulmonary wedge pressures limit increases in exercise capacity.
引用
收藏
页码:247 / 252
页数:6
相关论文
共 50 条
  • [41] Catheter Ablation for Ventricular Tachyarrhythmias in Patients Supported by Continuous-Flow Left Ventricular Assist Devices
    Garan, Arthur R.
    Iyer, Vivek
    Whang, William
    Mody, Kanika P.
    Yuzefpolskaya, Melana
    Colombo, Paolo C.
    Te-Frey, Rosie
    Takayama, Hiroo
    Naka, Yoshifumi
    Garan, Hasan
    Jorde, Ulrich P.
    Uriel, Nir
    ASAIO JOURNAL, 2014, 60 (03) : 311 - 316
  • [42] Electrophysiologic Changes and Their Effects on Ventricular Arrhythmias in Patients with Continuous-Flow Left Ventricular Assist Devices
    Celik, Mehmet
    Emiroglu, Mehmet Yunus
    Bayram, Zubeyde
    Izci, Servet
    Karagoz, Ali
    Akbal, Ozgur Yasar
    Kahyaoglu, Muzaffer
    Kup, Ayhan
    Yilmaz, Yusuf
    Kirali, Mehmet Kaan
    Ozdemir, Nihal
    ASAIO JOURNAL, 2022, 68 (03) : 341 - 348
  • [43] Gastrointestinal bleeding with continuous-flow left ventricular assist devices
    Amer S.
    Shah P.
    Hassan S.
    Clinical Journal of Gastroenterology, 2015, 8 (2) : 63 - 67
  • [44] Third-Generation Continuous Flow Left Ventricular Assist Devices
    Nguyen, Duc Q.
    Thourani, Vinod H.
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2010, 5 (04) : 250 - 258
  • [45] Physiologic effects of continuous-flow left ventricular assist devices
    Healy, Aaron H.
    McKellar, Stephen H.
    Drakos, Stavros G.
    Koliopoulou, Antigoni
    Stehlik, Josef
    Selzman, Craig H.
    JOURNAL OF SURGICAL RESEARCH, 2016, 202 (02) : 363 - 371
  • [46] Stroke after implantation of continuous flow left ventricular assist devices
    Tsiouris, Athanasios
    Heliopoulos, Ioannis
    Mikroulis, Dimitrios
    Mitsias, Panayiotis D.
    JOURNAL OF CARDIAC SURGERY, 2019, 34 (07) : 541 - 548
  • [47] Neurologic Events in Continuous-Flow Left Ventricular Assist Devices
    Kadakkal, Ajay
    Najjar, Samer S.
    CARDIOLOGY CLINICS, 2018, 36 (04) : 531 - +
  • [48] Long Term Survival in Continuous Flow Left Ventricular Assist Devices
    Gosev, I.
    Joseph, S. M.
    Eckman, P.
    Kilic, A.
    Uriel, N.
    Rich, J. D.
    Katz, J. N.
    Cowger, J.
    Lima, B.
    McGurk, S.
    Patel, C. B.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2015, 34 (04): : S81 - S81
  • [49] Hematologic disorders and continuous-flow left ventricular assist devices
    Connors, Jean M.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (11): : 1114 - 1116
  • [50] A Decade of Experience With Continuous-Flow Left Ventricular Assist Devices
    John, Ranjit
    Holley, Christopher T.
    Eckman, Peter
    Roy, Samit S.
    Cogswell, Rebecca
    Harvey, Laura
    Shumway, Sara
    Liao, Kenneth
    SEMINARS IN THORACIC AND CARDIOVASCULAR SURGERY, 2016, 28 (02) : 363 - 374