Performance Evaluation of the FLEX Triumph X-PET Scanner Using the National Electrical Manufacturers Association NU-4 Standards

被引:0
|
作者
Prasad, Rameshwar [1 ]
Ratib, Osman [1 ]
Zaidi, Habib [1 ,2 ]
机构
[1] Univ Hosp Geneva, Div Nucl Med, CH-1211 Geneva 4, Switzerland
[2] Univ Geneva, Geneva Neurosci Ctr, Geneva, Switzerland
基金
瑞士国家科学基金会;
关键词
PET; preclinical imaging; performance evaluation; NEMA; rodents; SMALL-ANIMAL PET; IMAGE-RECONSTRUCTION; RESOLUTION; MICROPET; DESIGN; SENSITIVITY; DETECTOR; CAMERA;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work was to evaluate the performance characteristics of the preclinical X-PET subsystem of the FLEX Triumph PET/CT scanner based on the NU 4-2008 standards of the National Electrical Manufacturers Association (NEMA). Methods: The performance parameters evaluated include the spatial resolution, scatter fraction, count losses and random coincidences, sensitivity, and image-quality characteristics. The PET detector array consisted of 11,520 individual bismuth germanate crystals arranged in 48 rings and 180 blocks, with an axial field of view (FOV) of 11.6 cm and a inner ring diameter of 16.5 cm. The spatial resolution was measured with a small Na-22 point source (diameter, 0.25 mm) at different radial offsets from the center. Sensitivity was calculated using the same source by stepping the source axially through the axial FOV of the scanner. Scatter fraction and counting-rate performances were determined using a mouse-and rat-sized phantom with an F-18 line source insert. The NEMA image-quality phantom and rodent imaging were also performed to access the overall imaging capabilities of the scanner. Results: Tangential spatial resolution in terms of full width at half maximum varied between 2.2 mm at the center of the FOV and 2.3 mm at a radial offset of 2.5 cm. The radial spatial resolution varied between 2.0 at the center and 4.4 mm at a radial offset of 2.3 cm. The peak system absolute sensitivity was 5.9% at the center of the FOV. The absolute system sensitivity was 0.67 counts/s/Bq, and the relative total system sensitivity was 73.9%. The scatter fraction for the mouse-sized phantom was 7.9%, with a peak true counting rate of 168 kilocounts per second (kcps) at 0.3 MBq/mL and a peak noise-equivalent counting rate of 106 kcps at 0.17 MBq/ mL. The rat-sized phantom had a scatter fraction of 21%, with a peak true counting rate of 93 kcps at 0.034 MBq/mL and a peak noise-equivalent counting rate of 49 kcps at 0.02 MBq/mL. Recovery coefficients for the image-quality phantom ranged from 0.13 to 0.88. Conclusion: The performance of the X-PET scanner based on the NEMA NU 4-2008 standards was fully characterized. The overall performance demonstrates that the X-PET system is suitable for preclinical research.
引用
收藏
页码:1608 / 1615
页数:8
相关论文
共 40 条
  • [31] NEMA NU4 performance evaluation of the SynchroPET human arterial PET (ArterialPET) scanner for standalone blood input function derivation
    Karakatsanis, Nicolas
    Fung, Edward
    Akerele, Mercy
    Pollenz, Louis
    Gross, Robert
    Sinelnikov, Yegor
    Mariner, Tom
    Alessi, Marc
    Babich, John
    Nehmeh, Sadek
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [32] Performance evaluation of a novel high performance pinhole array detector module using NEMA NU-4 image quality phantom for four head SPECT Imaging
    Rahman, Tasneem
    Tahtali, Murat
    Pickering, Mark R.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [33] Technical Note: Performance evaluation of a small-animal PET/CT system based on NEMA NU 4-2008 standards
    Liu, Qiong
    Li, Chaofan
    Liu, Jiguo
    Krish, Kishore
    Fu, Xinlei
    Zhao, Jie
    Chen, Jyh-Cheng
    MEDICAL PHYSICS, 2021, 48 (09) : 5272 - 5282
  • [34] Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner
    Disselhorst, Jonathan A.
    Brom, Maarten
    Laverman, Peter
    Slump, Cornelius H.
    Boerman, Otto C.
    Oyen, Wim J. G.
    Gotthardt, Martin
    Visser, Eric P.
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (04) : 610 - 617
  • [35] Performance Evaluation of a PET of 7T Bruker Micro-PET/MR Based on NEMA NU 4-2008 Standards (vol 11, 2194, 2022)
    Doss, Kishore Krishnagiri Manoj
    Mion, Pei En
    Kao, Yu-Chieh Jill
    Kuo, Tsung-Ter
    Chen, Jyh-Cheng
    ELECTRONICS, 2022, 11 (19)
  • [36] Image quality as a function of reconstruction algorithms and their parameter settings for the Siemens Inveon small-animal PET scanner evaluated using the NEMA NU4 2008 standards
    Visser, E. P.
    Disselhorst, J. A.
    van Lier, M. G. J. T.
    Laverman, P.
    Gotthardt, M.
    Oyen, W. J. G.
    Boerman, O. C.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 : S171 - S171
  • [37] Characterization and optimization of image quality as a function of reconstruction algorithms and parameter settings in a Siemens Inveon small-animal PET scanner using the NEMA NU 4-2008 standards
    Visser, Eric P.
    Disselhorst, Jonathan A.
    van Lier, Monique G. J. T. B.
    Laverman, Peter
    de Jong, Gabie M.
    Oyen, Wim J. G.
    Boerman, Otto C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 629 (01): : 357 - 367
  • [38] Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
    Jha, Ashish Kumar
    Mithun, Sneha
    Puranik, Ameya D.
    Purandare, Nilendu C.
    Shah, Sneha
    Agrawal, Archi
    Rangarajan, Venkatesh
    WORLD JOURNAL OF NUCLEAR MEDICINE, 2019, 18 (04) : 351 - 360
  • [39] Optimizing the Performance evaluation tests of Gemini TF64 PET/CT scanner using 68Ga isotope adhering NEMANU-2 2007 standards
    Rawat, B.
    Sawant, V.
    Jha, A. K.
    Mithun, S.
    Rangarajan, V.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S599 - S599
  • [40] Partial Volume Correction in Small Animal PET imaging: performance evaluation of a reconstruction based algorithm using the NEMA NU4 2008 image quality phantom
    D'Ambrosio, D.
    Zagni, F.
    Cicoria, G.
    Boschi, S.
    Fanti, S.
    Marengo, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S273 - S273