APPLICATION OF A SYNTHETIC DIAMOND DETECTOR FOR THE MEASUREMENT OF RELATIVE DOSE DISTRIBUTIONS IN RADIOTHERAPY

被引:0
|
作者
Kazuchits, N. [1 ]
Rusetskyi, M. [1 ]
Shulenkov, A. [2 ]
Kukushkin, V. [3 ]
Martynov, S. [3 ]
Khrunov, V. [3 ]
Tuzov, J. [3 ]
Gatskevich, G. [4 ]
Kandybovich, D. [4 ]
机构
[1] Belarusian State Univ, Minsk 220030, BELARUS
[2] UP Minskii NII Radiomat, Minsk 220024, BELARUS
[3] Inst Phys & Tech Problems, Dubna 141980, Moscow Region, Russia
[4] NN Alexandrov Natl Canc Ctr Belarus, Minsk, BELARUS
来源
MEDICAL PHYSICS IN THE BALTIC STATES: PROCEEDINGS OF THE 7TH INTERNATIONAL CONFERENCE ON MEDICAL PHYSICS | 2009年
关键词
Diamond; detector; sensitivity; gamma-rays; dose rate;
D O I
暂无
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The suitability of a synthetic diamond for the measurement of relative dose distributions in selected radiotherapy applications was studied. The performance of the synthetic diamond detector was analyzed by comparison with that of natural diamond detector produced by a Physikalisch-Technische Werkstatten (Freiburg, Germany) (PTW). The synthetic diamond detector is concluded to be suitable for relative dose measurements, especially when high response and high spatial resolution are required.
引用
收藏
页码:76 / +
页数:2
相关论文
共 50 条
  • [21] Clinical investigations of a CVD diamond detector for radiotherapy dosimetry
    Betzel, G. T.
    Lansley, S. P.
    Baluti, F.
    Reinisch, L.
    Meyer, J.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02): : 144 - 152
  • [22] Detector-Grade CVD Diamond for Radiotherapy Dosimetry
    Lansley, S.
    Betzel, G.
    Mckay, D.
    Meyer, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3738 - 3738
  • [23] Silicon substrate diamond film detector for gamma dose rate measurement in a high radiation environment
    Xue, Jinlong
    Hou, Qingfeng
    Niu, Liumin
    Ma, Zongmin
    Shi, Yunbo
    Tang, Jun
    Liu, Jun
    DIAMOND AND RELATED MATERIALS, 2022, 121
  • [24] Application of a diamond detector to brachytherapy dosimetry
    Rustgi, SN
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08): : 2085 - 2094
  • [25] Correlating Areas of Radiation Toxicity with Relative Biological Effectiveness-Weighted Dose Distributions in Proton Radiotherapy
    Peeler, C.
    Mirkovic, D.
    Grosshans, D.
    Titt, U.
    Mohan, R.
    MEDICAL PHYSICS, 2012, 39 (06) : 3994 - 3994
  • [26] SYNTHETIC DIAMOND-BASED DETECTOR OF IONIZED RADIATION
    ROMANKO, LA
    GONTAR, AG
    DIAMOND AND RELATED MATERIALS, 1994, 3 (03) : 259 - 263
  • [27] A new radiotherapy surface dose detector: The MOSFET
    Butson, MJ
    Rozenfeld, A
    Mathur, JN
    Carolan, M
    Wong, TPY
    Metcalfe, PE
    MEDICAL PHYSICS, 1996, 23 (05) : 655 - 658
  • [28] APPLICATION OF ARTIFICIAL NEURAL NETWORK FOR OPTIMISATION OF DOSE DISTRIBUTIONS IN STEREOTACTIC RADIOTHERAPY OF INTRACRANIAL CASES
    Skrobala, A.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S66 - S66
  • [29] Characterisation of a MOSFET-based detector for dose measurement under megavoltage electron beam radiotherapy
    Jong, W. L.
    Ung, N. M.
    Tiong, A. H. L.
    Rosenfeld, A. B.
    Wong, J. H. D.
    RADIATION PHYSICS AND CHEMISTRY, 2018, 144 : 76 - 84
  • [30] Development of a synthetic diamond radiation detector with a boron doped CVD diamond contact
    Kaneko, J
    Katagiri, M
    Ikeda, Y
    Nishitani, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3): : 211 - 215