Validation of the A&D UM-211 device for office blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010

被引:4
|
作者
Fania, Claudio [1 ]
Albertini, Federica [1 ]
Palatini, Paolo [1 ]
机构
[1] Univ Padua, Dept Med, Via Giustiniani 2, I-35128 Padua, Italy
关键词
cuff; device; office blood pressure; self-measurement; validation;
D O I
10.1097/MBP.0000000000000266
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The aim of this study was to define the accuracy of UM-211, an automated oscillometric device for office use coupled to several cuffs for different arm sizes, according to the International Protocol of the European Society of Hypertension. The validation was performed in 33 individuals. Their mean age was 59.6 +/- 12.9 years, systolic blood pressure (BP) was 144.3 +/- 21.5mmHg (range: 96-184mmHg), diastolic BP was 86.8 +/- 18.5mmHg (range: 48-124 mmHg), and arm circumference was 30.2 +/- 4.3 cm (range: 23-39 cm). Four sequential readings were taken by observers 1 and 2 using a double-headed stethoscope and a mercury sphygmomanometer, whereas three BP readings were taken by the supervisor using the test instrument. The differences between the readings provided by the device and the mean observer measurements were calculated. Therefore, each device measurement was compared with the previous and the next mean observer measurement. The validation results fulfilled all the 2010 European Society of Hypertension revision Protocol criteria for the general population and passed all validation grades. On average, the device overestimated systolic BP by 1.7 +/- 2.4mmHg and diastolic BP by 1.7 +/- 2.5mmHg. These data show that the UM-211 device coupled to several cuffs for different ranges of arm circumference met the requirements for validation according to the International Protocol and can be recommended for clinical use in the adult population. However, these results mainly apply to the use of the 22-32 and the 31-45 cm cuffs. Copyright (C) 2017 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:302 / 305
页数:4
相关论文
共 50 条
  • [31] Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (05) : 204 - 206
  • [32] Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Liu, Chang-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (04) : 230 - 233
  • [33] Validation of noninvasive oscillometric blood pressure 2020 up pressure upper arm blood pressure monitoring technology according to the European Society of Hypertension International Protocol revision 2010
    Reshetnik, Alexander
    Gohlisch, Christopher
    Abou-Dakn, Michael
    Toelle, Markus
    Zidek, Walter
    van der Giet, Markus
    BLOOD PRESSURE MONITORING, 2019, 24 (02) : 99 - 101
  • [34] Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010
    Abou-Dakn, Michael
    Wenzel, Sabine
    JOURNAL OF HUMAN HYPERTENSION, 2018, 32 (11) : 770 - 774
  • [35] Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010
    Michael Abou-Dakn
    Sabine Wenzel
    Journal of Human Hypertension, 2018, 32 : 770 - 774
  • [36] Validation of the Microlife Watch BP Office professional device for office blood pressure measurement according to the International protocol
    Stergiou, George S.
    Tzamouranis, Dimitris
    Protogerou, Athanasios
    Nasothimiou, Efthimia
    Kapralos, Christos
    BLOOD PRESSURE MONITORING, 2008, 13 (05) : 299 - 303
  • [37] Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Yi, Jun
    Wan, Yi
    Pan, Feng
    Yu, Xiaorong
    Zhao, Huadong
    Shang, Fujun
    Xu, Yongyong
    BLOOD PRESSURE MONITORING, 2011, 16 (04) : 197 - 199
  • [38] Validation of the iHealth BP5 wireless upper arm blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Shang, Fujun
    Zhu, Yizheng
    Zhu, Zhenlai
    Liu, Lei
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2013, 18 (05) : 278 - 281
  • [39] Validation of the AVITA BPM15S wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Zeng, Wei-Fang
    Zhang, Lu
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2014, 19 (03) : 183 - 186
  • [40] Validation of Transtek TMB-1776 according to European Society of Hypertension International Protocol revision 2010 in adults
    Liu, Zhonghua
    Chen, Liyi
    BLOOD PRESSURE MONITORING, 2019, 24 (06) : 319 - 322