Calibration of radiation protection area monitoring instruments in Sudan

被引:3
|
作者
Suliman, I. I. [1 ]
Youssif, B. E. [1 ]
Beineen, A. A. [1 ]
Hassan, M. [1 ]
机构
[1] Sudan Atom Energy Commiss, Radiat Safety Inst, Khartoum, Sudan
关键词
Calibrations; Radiation dosimetry; Ambient dose equivalent;
D O I
10.1016/j.radmeas.2010.06.005
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
This article presents results of measurements for the calibration of radiation protection area monitoring instruments carried out during the period 2006-2008 at Secondary Standard Dosimetry Laboratory of Sudan. The work performed included quality assurance measurements, measurements for the dosimeter calibrations and uncertainty analysis. Calibrations were performed using (137)Cs gamma ray sources produced by OB 85 and OB 34/1 gamma calibrators producing air kerma rate that ranged from 10 mu Gy/h to 50 mGy/h. Area monitoring instruments were calibrated in terms of ambient dose equivalent, H*(10) derived using air kerma to ambient dose equivalent conversion coefficients. Results are presented for 78 area monitoring instruments representing most commonly used types in Sudan. Radioactive check source measurements for the reference chamber showed deviation within 1% limit. The accuracy in the beam output measurements was within 5% internationally considered as acceptable. The results highlighted the importance of radiation protection calibrations. Regulations are further need to ensure safety aspect really meet the required international standards. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1578 / 1581
页数:4
相关论文
共 50 条
  • [1] New IAEA recommendations on the calibration of radiation protection monitoring instruments
    Bohm, J
    Thompson, IMG
    Griffith, R
    [J]. IRPA9 - 1996 INTERNATIONAL CONGRESS ON RADIATION PROTECTION / NINTH INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, PROCEEDINGS, VOL 4, 1996, : D186 - D188
  • [2] Latin American and Caribbean intercomparison of radiation protection area monitoring instruments
    Ramos, MMO
    da Cunha, PG
    Suárez, RC
    [J]. RADIATION PROTECTION DOSIMETRY, 2002, 102 (04) : 315 - 322
  • [3] BETA CALIBRATION OF RADIATION SURVEY INSTRUMENTS AT PROTECTION LEVELS
    OWEN, B
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (02): : 175 - &
  • [4] A SECONDARY STANDARD DOSIMETRY SYSTEM FOR CALIBRATION OF RADIATION PROTECTION INSTRUMENTS
    DUFTSCHMID, KE
    HIZO, J
    [J]. ACTA PHYSICA ACADEMIAE SCIENTIARUM HUNGARICAE, 1982, 52 (3-4): : 275 - 282
  • [5] The radiation field characteristics of a 137Cs source used for calibration of radiation protection instruments
    Fernandes, E
    Freire, D
    de Freltas, AC
    deAlmeida, CE
    [J]. APPLIED RADIATION AND ISOTOPES, 2004, 61 (06) : 1425 - 1430
  • [6] THERMAL-NEUTRON IRRADIATION FIELD FOR ROUTINE CALIBRATION OF RADIATION PROTECTION INSTRUMENTS
    ASANO, Y
    UESAWA, T
    MINAMI, K
    [J]. HEALTH PHYSICS, 1985, 49 (02): : 375 - 375
  • [7] CALIBRATION OF SATELLITE RADIATION INSTRUMENTS
    HOUGHTON, JT
    [J]. APPLIED OPTICS, 1977, 16 (02): : 295 - 296
  • [8] TEMPORAL DEPENDENCE OF THE RADIATION MONITORING INSTRUMENTS FOR AREA MONITORING USED AT RADIODIAGNOSTIC AND INTERVENTION FACILITIES
    Leyton, Fernando
    Macedo, Eric M.
    Ferreira, Mario J.
    Navarro, Valeria C. C.
    Garcia, Igor F. M.
    Pereira, Lara C. S.
    Navarro, Marcus V. T.
    [J]. RADIATION PROTECTION DOSIMETRY, 2016, 171 (04) : 509 - 519
  • [9] Recent developments at the Calibration Laboratory for radiation protection instruments and dosemeters of the Paul Scherrer Institute
    Kasprzak, Malgorzata
    Geser, Federico A.
    Sliz, Malgorzata
    Yukihara, Eduardo
    Mayer, Sabine
    [J]. RADIATION PROTECTION DOSIMETRY, 2023, 199 (15-16) : 1710 - 1715
  • [10] Traceability of radiation protection instruments
    Hino, Y.
    Kurosawa, T.
    [J]. METROLOGIA, 2007, 44 (04) : S146 - S152