Segmentation of the lumen and media-adventitial borders in intravascular ultrasound images using a geometric deformable model

被引:14
|
作者
Lee, Ju Hwan [1 ]
Hwang, Yoo Na [2 ]
Kim, Ga Young [2 ]
Min, Kim Sung [1 ,2 ]
机构
[1] Dongguk Univ Seoul, Dept Med Devices Ind, 04620 30,Pildong Ro 1 Gil, Seoul, South Korea
[2] Dongguk Univ, Dept Med Biotechnol, Bio Medi Campus 10326 32, Goyang Si, Gyeonggi Do, South Korea
关键词
medical image processing; image segmentation; biomedical ultrasonics; catheters; image denoising; evolutionary computation; set theory; media-adventitial borders; lumen segmentation; intravascular ultrasound images; geometric deformable model; intima segmentation; sequential intravascular ultrasound images; sequential IVUS image frames; human coronary arteries; vessel border estimation; border initialisation; edge preservation; noise reduction; dead zone preservation; local binary pattern-based mask initialisation; modified distance regularised level set evolution model; correlation coefficients; vessel perimeter; maximum vessel diameter; maximum lumen diameter; linear regression analysis; frequency; 20; MHz; 45; FAST-MARCHING METHOD; INTRACORONARY ULTRASOUND; AUTOMATIC SEGMENTATION; CORONARY-ARTERIES; CONTOUR-DETECTION; ACTIVE CONTOURS; IVUS IMAGES; IN-VIVO; PLAQUES; WALL;
D O I
10.1049/iet-ipr.2017.1143
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
This study presents a geometric deformable model-based segmentation approach to segmentation of the intima and media-adventitial (MA) borders in sequential intravascular ultrasound (IVUS) images. The initial estimation of the vessel borders was done manually only for the first frame of each sequence. After the border initialisation, pre-processing including edge preservation, noise reduction, and dead zone preservation was successively performed on each IVUS frame. To improve segmentation performance, the image masks were determined preliminarily by local binary pattern-based mask initialisation. Then, the inner and outer borders were approximated using a modified distance regularised level set evolution model. The results showed superior performance of the suggested approach for estimating intima and MA layers from the IVUS images. The corresponding correlation coefficients of area, vessel perimeter, maximum vessel diameter, and maximum lumen diameter were r=0.782, r=0.716, r=0.956, and r=0.874 for the 20MHz images, respectively, and r=0.990, r=0.995, r=0.989, and r=0.996 for the 45MHz images, respectively. In addition, linear regression analysis indicated that the manual segmentation had significantly high similarity at r>0.967 and r>0.993 for 20 and 45MHz images, respectively.
引用
收藏
页码:1881 / 1891
页数:11
相关论文
共 50 条
  • [11] Fully automated media and lumen boundary detection in intravascular ultrasound images
    Vandenberg, J
    Liersch, G
    Hanna, H
    Cameron, J
    [J]. PROCEEDINGS OF THE IEEE SOUTHWEST SYMPOSIUM ON IMAGE ANALYSIS AND INTERPRETATION, 1996, : 71 - 75
  • [12] Segmentation of interest region in medical volume images using geometric deformable model
    Lee, Myungeun
    Cho, Wanhyun
    Kim, Sunworl
    Park, Soonyoung
    Kim, Jong Hyo
    [J]. COMPUTERS IN BIOLOGY AND MEDICINE, 2012, 42 (05) : 523 - 537
  • [13] Supervised machine learning for coronary artery lumen segmentation in intravascular ultrasound images
    Cui, Hengfei
    Xia, Yong
    Zhang, Yanning
    [J]. INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2020, 36 (07)
  • [14] Validation of a new semi-automatic method for lumen and media-adventitia borders detection in sequential intravascular ultrasound images
    Giannoglou, G. D.
    Hatzizisis, I. S.
    Koutkias, V. G.
    Kompatsiaris, I.
    Diamantopoulos, P.
    Maglaveras, N.
    Parcharidis, G. E.
    Louridas, G. E.
    [J]. EUROPEAN HEART JOURNAL, 2005, 26 : 539 - 539
  • [15] AUTOMATIC DETECTION OF ARTERIAL LUMEN AND MEDIO-ADVENTITIAL BOUNDARIES IN INTRAVASCULAR ULTRASOUND IMAGES USING TEXTURAL OPERATORS AND MATHEMATICAL MORPHOLOGY
    MOJSILOVIC, A
    AMODA, N
    POPOVIC, M
    BABIC, R
    OSTOJIC, M
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, : A3 - A3
  • [16] A novel 3D segmentation method of the lumen from intravascular ultrasound images
    Alexandrescu, Ionut
    Cheriet, Farida
    Delorme, Sebastien
    [J]. IMAGE ANALYSIS AND RECOGNITION, PROCEEDINGS, 2007, 4633 : 949 - +
  • [17] AUTOMATED IDENTIFICATION OF LUMINAL AND MEDIAL-ADVENTITIAL BORDERS IN INTRAVASCULAR ULTRASOUND IMAGES OF HUMAN CORONARY-ARTERIES
    MEIER, DS
    COTHREN, RM
    TUZCU, EM
    MASSEROLI, M
    THOMAS, JD
    NISSEN, SE
    CORNHILL, JF
    [J]. CIRCULATION, 1995, 92 (08) : 370 - 370
  • [18] Segmentation in virtual colonoscopy using a geometric deformable model
    Wyatt, CL
    Ge, YR
    Vining, DJ
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2006, 30 (01) : 17 - 30
  • [19] Segmentation in virtual colonoscopy using a Geometric Deformable Model
    Wyatt, CL
    Ge, YR
    Vining, DJ
    [J]. MEDICAL IMAGING 2004: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, 2004, 5 (23): : 268 - 279
  • [20] Quantitative analysis for lumen and media-adventitia border detection in intravascular ultrasound images
    Wang, Ling
    Yu, Daoyin
    Chen, Xiaodong
    An, Zhiyong
    [J]. 2008 INTERNATIONAL CONFERENCE ON OPTICAL INSTRUMENTS AND TECHNOLOGY: OPTICAL SYSTEMS AND OPTOELECTRONIC INSTRUMENTS, 2009, 7156