A high-sensitivity large volume cryogenic detector for radon in gas

被引:7
|
作者
Wojcik, M. [1 ]
Zuzel, G. [2 ]
机构
[1] M Smoluchowski Inst Phys, PL-30059 Krakow, Poland
[2] Max Planck Inst Nucl Phys, D-69117 Heidelberg, Germany
关键词
Radon; Stainless Steel Plate; Minimum Detectable Activity; 222Rn Concentration; Cold Plate;
D O I
10.1007/s10967-008-0730-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new detector for Rn-222 measurements in gases is described. It is based on the low temperature collection of radon on a stainless steel plate and subsequent counting of emitted alpha-particles using an ion-implanted silicon detector. A very low background of 25 counts/day for the Rn-222 energy window was achieved. High detection efficiency of 32% (2 pi geometry) and the large 65 liter volume of the detector allow radon measurements at the level of 12 mBq/m(3) with 30% accuracy (relative error). Behavior of the alpha-emitting radon daughters born on the cooled stainless steel plate, Po-218 and Po-214, was also analyzed.
引用
收藏
页码:199 / 205
页数:7
相关论文
共 50 条
  • [1] A high-sensitivity large volume cryogenic detector for radon in gas
    M. Wojcik
    G. Zuzel
    [J]. Journal of Radioanalytical and Nuclear Chemistry, 2008, 277 : 199 - 205
  • [2] Radon concentration measurement with a high-sensitivity radon detector at the Yemilab
    Seo, Kyung Min
    Kim, Hyunsoo
    Kim, Yeong Duk
    Lee, Hye Young
    Lee, Jaison
    Lee, Moo Hyun
    So, Jungho
    Yoon, Sang Cheol
    Yoon, Young Soo
    [J]. JOURNAL OF INSTRUMENTATION, 2024, 19 (05):
  • [3] A HIGH-SENSITIVITY SCINTILLATION CHAMBER FOR RADON IN GAS
    WOJCIK, M
    WLAZLO, W
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 345 (02): : 351 - 355
  • [4] Development of a high-sensitivity 80 L radon detector for purified gases
    Hosokawa, K.
    Murata, A.
    Nakano, Y.
    Onishi, Y.
    Sekiya, H.
    Takeuchi, Y.
    Tasaka, S.
    [J]. PROGRESS OF THEORETICAL AND EXPERIMENTAL PHYSICS, 2015, 2015 (03):
  • [5] CHARACTERISTICS OF A LARGE AREA HIGH-SENSITIVITY PARTICLE DETECTOR
    NAKASHIMA, Y
    ICHIMURA, M
    MIYOSHI, S
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 1985, 24 (09): : L699 - L701
  • [6] LARGE SCINTILLATION CELLS FOR HIGH-SENSITIVITY RADON CONCENTRATION MEASUREMENTS
    COHEN, BL
    ELGANAYNI, M
    COHEN, ES
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 212 (1-3): : 403 - 412
  • [7] A novel low background cryogenic detector for radon in gas
    Wojcik, Marcin
    Zuzel, Grzegorz
    [J]. TOPICAL WORSHOP ON LOW RADIOACTIVITY TECHNIQUES, 2007, 897 : 39 - +
  • [8] Review of high-sensitivity Radon studies
    Wojcik, M.
    Zuzel, G.
    Simgen, H.
    [J]. INTERNATIONAL JOURNAL OF MODERN PHYSICS A, 2017, 32 (30):
  • [9] HIGH-SENSITIVITY FREQUENCY DETECTOR
    ZENKOVIC.AV
    [J]. TELECOMMUNICATIONS AND RADIO ENGINEER-USSR, 1972, (12): : 35 - &
  • [10] A HIGH-SENSITIVITY SUPERCONDUCTING DETECTOR
    ZIMMERMAN, JE
    SILVER, AH
    [J]. JOURNAL OF APPLIED PHYSICS, 1968, 39 (06) : 2679 - +