The accuracy of blood pressure measured by arterial line and non-invasive cuff in critically ill children

被引:30
|
作者
Joffe, Rachel [1 ]
Duff, Jonathan [2 ]
Guerra, Gonzalo Garcia [2 ]
Pugh, Jodie [2 ]
Joffe, Ari R. [2 ,3 ]
机构
[1] Univ Alberta, Fac Sci, Edmonton, AB, Canada
[2] Univ Alberta, Dept Pediat, Edmonton, AB, Canada
[3] 4-546 Edmonton Clin Hlth Acad,11405 87 Ave, Edmonton, AB T6G 1C9, Canada
来源
CRITICAL CARE | 2016年 / 20卷
关键词
Arterial line; Blood pressure; Children; Intensive care units; Pediatric; Monitoring; Physiologic; NEWBORNS; MONITOR;
D O I
10.1186/s13054-016-1354-x
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The accuracy of arterial lines (AL) using the flush test or stopcock test has not been described in children, nor has the difference between invasive arterial blood pressure (IABP) versus non-invasive cuff (NIBP) blood pressure. Methods: After ethics approval and consent, we performed the flush test and stopcock test on AL (to determine over damping, under damping, and optimal damping), and determined the difference (NIBP-IABP) in systolic, diastolic, and mean blood pressure (Delta SBP, Delta DBP, and Delta MAP). The primary outcome was incidence (95 % CI) of optimally damped AL. Predictors of Delta BP (effect size (95 % CI)) were determined using multiple linear regression. Results: There were 147 AL tests in 100 enrolled patients with mean age 44.7 (SD 56) months, weight 16.8 (SD 18. 3) kg, male 59 %, postoperative-cardiovascular 52 %, peripheral-AL 78 %, inotropes 29 %, vasodilators 15 %, and ventilated 73 %. The flush test performed in 66 patients (45 %) showed optimal damping in 30 (46 %; 95 % CI 34, 57 %), over damping in 25 (38 %) and under damping in 11 patients (17 %). The stopcock test was over-damped in 128/146 patients (88 %), with the same damping as the flush test in 24/64 (38 %). In optimally damped (flush test) AL, Delta SBP, Delta DBP, and Delta MAP were 0.8 (SD 12.2), -5.2 (SD 8.7), and -4.9 (7.6) respectively. A second set of AL tests was done 2 h later on the same day in 62 patients; AL damping often changed (10/28 flush tests) and Delta BPs correlated poorly (r = 0.31-0.55). Predictors (effect size) of Delta DBP were vasodilator infusion (15.6 (2.9 to 28.3); p = 0.016) and optimal damping (-7.2 (-12.2 to 2.2); p = 0.005); and of Delta MAP were vasodilator infusion (10.0 (-0.3 to 20.4); p = 0.057) and optimal damping (-4.0 (-8 to 0.1); p = 0.058). There were no independent predictors of damping category (n = 66 flush tests). Conclusions: Optimally damped AL occur in half of critically ill children, and this is not predictable. There is much variability in Delta BP between NIBP and the gold standard IABP, and this varies even in the same patient on the same day, and is not easily predictable. In critically ill children, NIBP may not be accurate enough to guide management, and more attention to ensuring the AL is optimally damped is needed.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] The Mean Prehospital Machine; accurate prehospital non-invasive blood pressure measurement in the critically ill patient
    Sandy Muecke
    Andrew Bersten
    John Plummer
    Journal of Clinical Monitoring and Computing, 2010, 24 : 191 - 202
  • [32] A comparison of invasive and non-invasive arterial blood pressure measurements in the neonate
    Shea, J. E. O'
    Dempsey, E. M.
    ACTA PAEDIATRICA, 2007, 96 : 190 - 191
  • [33] A COMPARISON OF INVASIVE AND NON-INVASIVE ARTERIAL BLOOD PRESSURE MEASUREMENTS IN THE NEONATE
    O'Shea, J.
    Dempsey, E.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2010, 179 : S174 - S174
  • [34] Complication of a non-invasive blood pressure cuff during prone positioning
    Kempen, PM
    Meng, L
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2006, 53 (01): : 106 - 106
  • [35] NICBPM: Non-Invasive Cuff-less Blood Pressure Monitor
    Aboughaly, Ali A.
    Iqbal, Danyal
    Abd El Ghany, Mohamed A.
    Hofmann, Klaus
    2017 29TH INTERNATIONAL CONFERENCE ON MICROELECTRONICS (ICM), 2017, : 121 - 124
  • [36] Complication of a non-invasive blood pressure cuff during prone positioning
    Paul M. Kempen
    Li Meng
    Canadian Journal of Anesthesia, 2006, 53 : 106 - 106
  • [37] A New Non-Invasive Cuff-Less Blood Pressure Sensor
    Tu, Tse-Yi
    Chao, Paul C. -P.
    Lee, Yung-Pin
    2013 IEEE SENSORS, 2013, : 857 - 860
  • [38] ACCURACY OF A NOVEL NON-INVASIVE TRANSDERMAL CONTINUOUS GLUCOSE MONITOR IN CRITICALLY ILL PATIENTS
    Nasraway, Stanley
    England, Michael
    Ehsan, Afshin
    Melanson, Ann Marie
    Menzie, Wayne
    Trieu, My-Quyen
    Berlin, Jason
    Hurley, James
    Krystyniak, Keith
    Segal, Scott
    CRITICAL CARE MEDICINE, 2012, 40 (12) : U88 - U89
  • [39] Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children
    Garnier, R. P.
    van der Spoel, A. G. E.
    Sibarani-Ponsen, R.
    Markhorst, D. G.
    Boer, C.
    BRITISH JOURNAL OF ANAESTHESIA, 2012, 109 (04) : 609 - 615
  • [40] Comparison of three non-invasive hemodynamic monitoring methods in critically ill children
    Chaiyakulsil, Chanapai
    Chantra, Marut
    Katanyuwong, Poomiporn
    Khositseth, Anant
    Anantasit, Nattachai
    PLOS ONE, 2018, 13 (06):