Source model tuning for a 6 MV photon beam used in radiotherapy

被引:0
|
作者
Hirschi, Lukas A. [1 ,2 ,3 ]
Siebers, Jeffrey V. [3 ]
Fix, Michael K. [1 ,2 ]
机构
[1] Inselspital Bern, Div Med Radiat Phys, Bern, Switzerland
[2] Univ Bern, CH-3012 Bern, Switzerland
[3] Virginia Commonwealth Univ, Richmond, VA USA
关键词
D O I
10.1088/1742-6596/74/1/012008
中图分类号
O59 [应用物理学];
学科分类号
摘要
The purpose of this work was to do a feasibility study on a procedure for tuning the mean energy T) and the spatial spread (sigma(R)) of the initial electron beam hitting the bremsstrahlung target as a part of a multicomponent source model, so that dose distributions predicted by Monte Carlo simulations (MC) will dosimetrically match measurements. The dose in a water phantom is considered to be a sum of doses produced by different components of a linear accelerator, namely target (TGT), flattening filter (1717), and primary collimator (PC). A histogram-based source model is used to model the phase space output of the 6 MV linear accelerator (VARIAN CL21EX). The phantom dose contribution from the target subsource has the greatest sensitivity to the initial beam parameters ((E) over bar sigma(R)), whereas the FF and PC subsource dose distributions axe only weakly dependent on them. In order to create a source model that could be tuned and used for a continuous range of (E) over bar and sigma(R), a method to interpolate across the histograms created with this data and to determine the sampling weights is developed, i.e. energy and radial-dependent interpolative polynomial fits to the histograms. The simulated dose distribution is then iteratively tuned to measured dose by changing the beam parameters, computing the corresponding dose distribution and compare simulation with measurements. This process is repeated until simulated dose matches measurements. The method ensures that the relative weights of radiation subsources are consistent with those that would be derived from full MC simulations of the treatment head. For comparison, the depth dose and the lateral dose profiles at various depths of the 10 x 10 cm(2) 20 x 20 cm(2), and 30 x 30 cm(2) field sizes are used. This study showed that a general source model tuning is feasible.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] Surface dose measurements in chest wall postmastectomy radiotherapy to achieve optimal dose delivery with 6 MV photon beam
    Lobo, Dilson
    Banerjee, Sourjya
    Srinivas, Challapalli
    Athiyamaan, M. S.
    Reddy, Shreyas
    Sunny, Johan
    Ravichandran, Ramamoorthy
    Kotian, Himani
    Prakash Saxena, P. U.
    JOURNAL OF MEDICAL PHYSICS, 2021, 46 (04) : 324 - 333
  • [22] Cellular analysis on the radiation induced bystander effects due to bismuth oxide nanoparticles with 6 MV photon beam radiotherapy
    Zainudin, Nur Hamizah Mohd
    Sisin, Noor Nabilah Talik
    Ab Rashid, Raizulnasuha
    Jamil, Amirah
    Anuar, Muhammad Afiq Khairil
    Razak, Khairunisak Abdul
    Abdullah, Reduan
    Rahman, Wan Nordiana
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2022, 15 (03) : 318 - 325
  • [23] CARDIAC EVENTS AFTER RADIOTHERAPY OF THE INTERNAL MAMMARY CHAIN BY AN EXCLUSIVE 6 MV PHOTON BEAM: A COHORT FROM EPINAL
    Simon, J. M.
    Renoult, F.
    Admant, P.
    Marchesi, V.
    Noel, A.
    El Hajj, L.
    Levitchi, M.
    Marchal, C.
    Eschwege, F.
    Peiffert, D.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S129 - S129
  • [24] Characterization of wedge factors and dose distributions in radiotherapy with symmetric and asymmetric physical wedged beams of 6 MV photon beam
    Zabihzadeh M.
    Fadaei M.
    Hoseini S.M.
    Arvandi S.
    Tahmasbi M.
    Iranian Journal of Medical Physics, 2020, 17 (03) : 213 - 219
  • [25] Dosimetric properties of a beam quality-matched 6 MV unflattened photon beam
    Huang, Yunfei
    Siochi, R. Alfredo
    Bayouth, John E.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (04): : 71 - 81
  • [26] Evaluation of the dosimetric characteristics of 6 MV flattened and unflattened photon beam
    Mohammed, Maged
    Chakir, E.
    Boukhal, H.
    Mroan, S.
    El Bardouni, T.
    JOURNAL OF KING SAUD UNIVERSITY SCIENCE, 2017, 29 (03) : 371 - 379
  • [27] Characterization of GaN dosimetry for 6 MV photon beam in clinical conditions
    Chaikh, A.
    Balosso, J.
    Giraud, J. -Y.
    Wang, R.
    Pittet, P.
    Lu, G. -N.
    RADIATION MEASUREMENTS, 2014, 71 : 392 - 395
  • [28] Investigation of the RBE of radiation in the penumbra of a 6mv photon beam
    Syme, A.
    Kirkby, C.
    Mirzayans, R.
    MacKenzie, M.
    Field, C.
    Fallone, G.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S78 - S78
  • [29] Investigation of Lower Energy (< 6 MV) Photon Beams for Cancer Radiotherapy
    Zhang, Y.
    Feng, Y.
    Ming, X.
    Zhou, L.
    Ahmad, M.
    Nguyen, K.
    Griffin, M.
    Deng, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 274 - 274
  • [30] Distribution of secondary electrons in the photon beam of a 6-MV and 15-MV linear accelerator
    Wiezorek, T.
    Scheithauer, M.
    Salz, H.
    Meininger, G.
    Harder, D.
    Zeitschrift fur Medizinische Physik, 2000, 10 (04): : 265 - 273