A JFET-CMOS fast preamplifier for segmented germanium detectors

被引:15
|
作者
Pullia, A. [1 ]
Zocca, F. [1 ]
Riboldi, S. [1 ]
机构
[1] Univ Milan, Ist Nazl Fis Nucl, Dept Phys, I-20133 Milan, Italy
关键词
CMOS preamplifier; charge-sensitive preamplifier; gamma ray spectroscopy;
D O I
10.1109/TNS.2007.914022
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A JFET-CMOS fast charge-sensitive preamplifier for germanium detectors, able to operate at cryogenic temperatures, has been designed, realized, and characterized. The monolithic part of the circuit is realized in a mature 5 V 0.8 mu m Si CMOS technology, which yields better performances than scaled technologies in this case. The input transistor is an external Si JFET, which can be easily replaced if necessary. The charge-to-voltage gain and the fall-time are as well set through an external RC,network. The circuit works in the wide temperature range of -196 to 55 degrees C and is able to drive a terminated coaxial cable with an exceptionally fast and clean transition. Namely, with a detector capacitance of 16 pF and a negative power supply of -3 V it is able to provide a -2.4 V pulse onto a 100 Omega load in less than 13 ns with no ringing. The static power consumption is 8 mW excluding the JFET. The area occupancy of the integrated circuit is as little as 366 x 275 mu m(2). The noise performance with a 16 pF detector capacitance is 110 r.m.s. electrons both at room temperature and at -196 degrees C, at a quasi-Gaussian shaping time of 10 mu s. The obtained performance is adequate for gamma-ray spectroscopy and pulse-shape analysis with bulky HPGe segmented detectors.
引用
收藏
页码:591 / 594
页数:4
相关论文
共 50 条
  • [21] A fast current preamplifier for silicon detectors
    E. V. Atkin
    A. G. Voronin
    A. D. Klyuev
    M. M. Merkin
    A. Yu. Pakhomov
    Instruments and Experimental Techniques, 2010, 53 : 86 - 88
  • [22] JFET-PMOS TECHNOLOGY IN THE DESIGN OF MONOLITHIC PREAMPLIFIER SYSTEMS FOR MULTIELECTRODE DETECTORS
    BUTTLER, W
    VOGT, H
    LUTZ, G
    MANFREDI, PF
    SPEZIALI, V
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1991, 38 (02) : 69 - 76
  • [23] Low noise, dual gain preamplifier with built in spectroscopic pulser for highly segmented high-purity germanium detectors
    Pascovici, G.
    Pullia, A.
    Zocca, F.
    Bruyneel, B.
    Bazzacco, D.
    WSEAS Transactions on Circuits and Systems, 2008, 7 (06): : 470 - 481
  • [24] Low-noise fully differential amplifiers using JFET-CMOS integration technology for smart sensors
    Takao, Hidekuni
    Vatedka, Radhakrishna
    Ito, Yoshiaki
    Komakine, Fumihito
    Serge, Kolelas
    Sawada, Kazuaki
    Ishida, Makoto
    IEEJ TRANSACTIONS ON ELECTRICAL AND ELECTRONIC ENGINEERING, 2008, 3 (03) : 274 - 280
  • [25] Test of a Fully Integrated CMOS Preamplifier for HPGe Detectors
    Riboldi, S.
    Pullia, A.
    Zocca, F.
    Budjas, D.
    D'Andragora, A.
    Cattadori, C.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 1346 - +
  • [26] Segmented high-purity germanium detectors
    Yu. B. Gurov
    K. N. Gusev
    S. L. Katulina
    V. G. Sandukovsky
    D. Borowicz
    J. Yurkowski
    Instruments and Experimental Techniques, 2007, 50 : 757 - 760
  • [27] FAST PREAMPLIFIER WITH SEPARATION OF SIGNAL FOR SEMICONDUCTING DETECTORS
    LANKO, EV
    DOMBROVSKAYA, GS
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES-USSR, 1968, (06): : 1376 - +
  • [28] Segmented high-purity germanium detectors
    Gurov, Yu. B.
    Gusev, K. N.
    Katualina, S. L.
    Sandukovsky, V. G.
    Borowicz, D.
    Yurkowski, J.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2007, 50 (06) : 757 - 760
  • [29] A FAST COINCIDENCE PREAMPLIFIER FOR SOLID STATE RADIATION DETECTORS
    WAHL, H
    NUCLEAR INSTRUMENTS & METHODS, 1964, 25 (02): : 247 - 252
  • [30] Coulomb excitation of exotic nuclei and segmented germanium detectors
    Glasmacher, T
    ENAM 98: EXOTIC NUCLEI AND ATOMIC MASSES, 1998, 455 : 204 - 208