Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures

被引:111
|
作者
Efstathopoulos, E. P. [1 ]
Pantos, I. [1 ]
Andreou, M. [1 ]
Gkatzis, A. [1 ]
Carinou, E. [2 ]
Koukorava, C. [2 ]
Kelekis, N. L. [1 ]
Brountzos, E. [1 ]
机构
[1] Univ Athens, Dept Radiol 2, Sch Med, GR-12462 Athens, Greece
[2] GAEC, Aghia Paraskevi, Greece
来源
BRITISH JOURNAL OF RADIOLOGY | 2011年 / 84卷 / 997期
关键词
SURGICAL PLACEMENT; STAFF DOSIMETRY; MEDICAL STAFF; PATIENT; EXPOSURE; CATHETERS; HANDS;
D O I
10.1259/bjr/83222759
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The aim of this study was to determine occupational dose levels in interventional radiology and cardiology procedures. Methods: The study covered a sample of 25 procedures and monitored occupational dose for all laboratory personnel. Each individual wore eight thermoluminescent dosemeters next to the eyes, wrists, fingers and legs during each procedure. Radiation protection shields used in each procedure were recorded. Results: The highest doses per procedure were recorded for interventionists at the left wrist (average 485 mu Sv, maximum 5239 mu Sv) and left finger (average 324 mu Sv, maximum 2877 mu Sv), whereas lower doses were recorded for the legs (average 124 mu Sv, maximum 1959 mu Sv) and the eyes (average 64 mu Sv, maximum 1129 mu Sv). Doses to the assisting nurses during the intervention were considerably lower; the highest doses were recorded at the wrists (average 26 mu Sv, maximum 41 mu Sv) and legs (average 18 mu Sv, maximum 22 mu Sv), whereas doses to the eyes were minimal (average 4 mu Sv, maximum 16 mu Sv). Occupational doses normalised to kerma area product (KAP) ranged from 11.9 to 117.3 mu Sv/1000 cGy cm(2) and KAP was poorly correlated to the interventionists' extremity doses. Conclusion: Calculation of the dose burden for interventionists considering the actual number of procedures performed annually revealed that dose limits for the extremities and the lenses of the eyes were not exceeded. However, there are cases in which high doses have been recorded and this can lead to exceeding the dose limits when bad practices are followed and the radiation protection tools are not properly used.
引用
收藏
页码:70 / 77
页数:8
相关论文
共 50 条
  • [21] EXTREMITY AND EYE LENS DOSES IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY PROCEDURES: FIRST RESULTS OF THE ORAMED PROJECT
    Domienik, J.
    Brodecki, M.
    Carinou, E.
    Donadille, L.
    Jankowski, J.
    Koukorava, C.
    Krim, S.
    Nikodemova, D.
    Ruiz-Lopez, N.
    Sans-Merce, M.
    Struelens, L.
    Vanhavere, F.
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 442 - 447
  • [22] Influence of patient thickness and operation modes on occupational and patient radiation doses in interventional cardiology
    Vano, E.
    Gonzalez, L.
    Fernandez, J. M.
    Prieto, C.
    Guibelalde, E.
    RADIATION PROTECTION DOSIMETRY, 2006, 118 (03) : 325 - 330
  • [23] Occupational radiation doses in interventional cardiology:: a 15-year follow-up
    Vaño, E
    Gonzalez, L
    Fernandez, JM
    Alfonso, F
    Macaya, C
    BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (941): : 383 - 388
  • [24] PATIENT RADIATION DOSES IN THE MOST COMMON INTERVENTIONAL CARDIOLOGY PROCEDURES IN CROATIA: FIRST RESULTS
    Brnic, Z.
    Krpan, T.
    Faj, D.
    Kubelka, D.
    Ramac, J. Popic
    Posedel, D.
    Steiner, R.
    Vidjak, V.
    Brnic, V.
    Viskovic, K.
    Baraban, V.
    RADIATION PROTECTION DOSIMETRY, 2010, 138 (02) : 180 - 186
  • [25] Assessment of Occupational Radiation Doses in Different Diagnostic and Interventional Radiology and Molecular Imaging Services
    Bouchareb, Y.
    Al-Maskery, I.
    Al-Zeheimi, H.
    Al-Rasbi, A.
    Al-Makhmari, N.
    Al-Haji, A.
    Al-Dhuhli, H.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S851 - S851
  • [26] COMPARING MEASURED AND CALCULATED DOSES IN INTERVENTIONAL CARDIOLOGY PROCEDURES
    Oliveira da Silva, M. W.
    Canevaro, L. V.
    Hunt, J.
    Rodrigues, B. B. D.
    RADIATION PROTECTION DOSIMETRY, 2017, 176 (04) : 439 - 443
  • [27] Patient Radiation Doses in Interventional Cardiology Procedure
    Hiswara, E.
    Kartikasari, D.
    Nuraeni, N.
    Sofyan, H.
    Sandy, K. Y. P.
    ATOM INDONESIA, 2022, 48 (02) : 81 - 85
  • [28] A COMPARISON OF OCCUPATIONAL DOSES IN CONVENTIONAL AND INTERVENTIONAL RADIOLOGY IN IRAN
    Mianji, Fereidoun
    Gheshlaghi, Fariba
    Darabi, Majid
    Kardan, Mohammad Reza
    Hosseinipanah, Mirshahram
    Zakeri, Farideh
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (01) : 37 - 47
  • [29] Variation in Radiation Doses in Pediatric and Adult Interventional Radiology Procedures: The Need for Dose Optimization
    Al-Haj, A.
    Lobriguito, A.
    Chantziantoniou, K.
    Lagarde, C.
    Iqelan, N.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [30] Estimating Radiation Doses to the Skin from Interventional Radiology Procedures for a Patient Population with Cancer
    Dauer, Lawrence T.
    Thornton, Raymond
    Erdi, Yusuf
    Ching, Hung
    Hamacher, Klaus
    Boylan, Daniel C.
    Williamson, Matthew J.
    Balter, Stephen
    Germain, Jean St.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (06) : 782 - 788