Wear analysis in anatomical and reversed shoulder prostheses

被引:8
|
作者
Ribeiro, N. S. [1 ]
Folgado, J. [1 ]
Fernandes, P. R. [1 ]
Monteiro, J. [2 ]
机构
[1] Univ Tecn Lisboa, IDMEC IST, P-1049001 Lisbon, Portugal
[2] Univ Lisbon, Sch Med, P-1649028 Lisbon, Portugal
关键词
total shoulder prostheses; polyethylene wear; shoulder joint; finite element models; POLYETHYLENE WEAR; GLENOID COMPONENT; BONE; ARTHROPLASTY; COMPLICATIONS; OSTEOLYSIS; STRENGTH; BEHAVIOR; GEOMETRY; CONTACT;
D O I
10.1080/10255842.2010.499868
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper describes the development of a computational model to calculate wear rates in total shoulder prostheses, for a 5-150 degrees arm abduction. Anatomical keeled and pegged prosthesis as well as reversed prosthesis were the studied implants. The bone models were built based on computed tomography (CT) images and using a computer aided design-based modelling pipeline. The finite element method was used to solve the contact problem between the surface of the polyethylene (PE) components and the corresponding articular component. The aim of this work was to determine linear and volumetric PE wear, for several radial mismatches, in conditions of pathological (rheumatoid arthritis) and non-pathological bone. Results showed that contact pressures and linear wear developed in anatomical prosthesis were higher than those visualised in reversed prosthesis. However, anatomical prosthesis exhibited a better volumetric wear performance. Moreover, our findings indicated higher values of volumetric wear in higher congruent models and on pathological bone conditions.
引用
收藏
页码:883 / 892
页数:10
相关论文
共 50 条
  • [1] Comparison of polyethylene wear in anatomical and reversed shoulder prostheses
    Terrier, A.
    Merlini, F.
    Pioletti, D. P.
    Farron, A.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (07): : 977 - 982
  • [2] Computational analysis of polyethylene wear in anatomical and reverse shoulder prostheses
    Quental, C.
    Folgado, J.
    Fernandes, P. R.
    Monteiro, J.
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2015, 53 (02) : 111 - 122
  • [3] Computational analysis of polyethylene wear in anatomical and reverse shoulder prostheses
    C. Quental
    J. Folgado
    P. R. Fernandes
    J. Monteiro
    Medical & Biological Engineering & Computing, 2015, 53 : 111 - 122
  • [4] Design of a wear simulator for in vitro shoulder prostheses testing
    van der Pijl A.J.
    Świěszkowski W.
    Bersee H.E.N.
    Experimental Techniques, 2004, 28 (5) : 45 - 48
  • [5] Design of a wear simulator for in vitro shoulder prostheses testing
    van der Pijl, AJ
    Swieszkowski, W
    Bersee, HEN
    EXPERIMENTAL TECHNIQUES, 2004, 28 (05) : 45 - 48
  • [6] Differential indication of anatomic and reversed shoulder prostheses in fracture sequelae
    Irlenbusch, U.
    Fuhrmann, U.
    Gebhardt, K.
    Rott, O.
    ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2008, 146 (04): : 478 - 485
  • [7] Simulated joint and muscle forces in reversed and anatomic shoulder prostheses
    Terrier, A.
    Reist, A.
    Merlini, F.
    Farron, A.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (06): : 751 - 756
  • [8] Comparison of anatomical shoulder prostheses and the proximal humeri of Chinese people
    School of Mechanical Engineering, Shanghai Jiao Tong University, 800 Dongchuan Road, Shanghai 200240, China
    Proc. Inst. Mech. Eng. Part H J. Eng. Med., 2007, 8 (921-927):
  • [9] Comparison of anatomical shoulder prostheses and the proximal humeri of Chinese people
    Zhang, L.
    Yuan, B.
    Wang, C.
    Liu, Z.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H8) : 921 - 927
  • [10] Inversion of bearing materials in reversed shoulder prostheses. Is there any difference?
    Irlenbusch, Ulrich
    Rott, Olaf
    Braunke, Mike
    OBERE EXTREMITAET-SCHULTER-ELLENBOGEN-HAND-UPPER EXTREMITY-SHOULDER ELBOW HAND, 2015, 10 (03): : 173 - 179