Label-free detection of nosocomial bacteria using a nanophotonic interferometric biosensor

被引:1
|
作者
Maldonado, Jesus [1 ,2 ]
Estevez, M. -Carmen [1 ,2 ]
Fernandez-Gavela, Adrian [1 ,2 ]
Jose Gonzalez-Lopez, Juan [3 ]
Belen Gonzalez-Guerrero, Ana [1 ,2 ]
Lechuga, Laura M. [1 ,2 ]
机构
[1] Catalan Inst Nanosci & Nanotechnol ICN2, CSIC, Nanobiosensors & Bioanalyt Applicat Grp, CIBER,BBN, Campus UAB, Barcelona 08193, Spain
[2] BIST, Campus UAB, Barcelona 08193, Spain
[3] Univ Autonoma Barcelona, Vall dHebron Inst Recerca VHIR, Hosp Univ Vall dHebron, Dept Clin Microbiol, Barcelona 08035, Spain
关键词
RESISTANT STAPHYLOCOCCUS-AUREUS; PSEUDOMONAS-AERUGINOSA; MRSA; DIAGNOSIS; SILICON; GENE; MECA;
D O I
10.1039/c9an01485c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Nosocomial infections are a major concern at the worldwide level. Early and accurate identification of nosocomial pathogens is crucial to provide timely and adequate treatment. A prompt response also prevents the progression of the infection to life-threatening conditions, such as septicemia or generalized bloodstream infection. We have implemented two highly sensitive methodologies using an ultrasensitive photonic biosensor based on a bimodal waveguide interferometer (BiMW) for the fast detection of Pseudomonas aeruginosa and methicillin-resistant Staphylococcus aureus (MRSA), two of the most prevalent bacteria associated with nosocomial infections. For that, we have developed a biofunctionalization strategy based on the use of a PEGylated silane (silane-PEG-COOH) which provides a highly resistant and bacteria-repelling surface, which is crucial to specifically detect each bacterium. Two different biosensor assays have been set under standard buffer conditions: one based on a specific direct immunoassay employing polyclonal antibodies for the detection of P. aeruginosa and another one employing aptamers for the direct detection of MRSA. The biosensor immunoassay for P. aeruginosa is fast (it only takes 12 min) and specific and has experimentally detected concentrations down to 800 cfu mL(-1) (cfu: colony forming unit). The second one relies on the use of an aptamer that specifically detects penicillin-binding protein 2a (PBP2a), a protein only expressed in the MRSA mutant, providing a photonic biosensor with the ability to identify the resistant pathogen MRSA and differentiate it from methicillin-susceptible S. aureus (MSSA). Direct, label-free, and selective detection of whole MRSA bacteria has been achieved, making possible the direct detection of also 800 cfu mL(-1). According to the signal-to-noise (S/N) ratio of the device, a theoretical limit of detection (LOD) of around 49 and 29 cfu mL(-1) was estimated for P. aeruginosa and MRSA, respectively. Both results obtained under standard conditions reveal the great potential this interferometric biosensor device has as a versatile and specific tool for bacterial detection and quantification, providing a rapid method for the identification of nosocomial pathogens within the clinical requirements of sensitivity for the diagnosis of infections.
引用
收藏
页码:497 / 506
页数:10
相关论文
共 50 条
  • [41] Nanophotonic interferometric immunosensors for label-free and real-time monitoring of chemical contaminants in marine environment
    Chocarro-Ruiz, B.
    Herranz, S.
    Fernandez Gavela, A.
    Lechuga, L. M.
    ADVANCED ENVIRONMENTAL, CHEMICAL, AND BIOLOGICAL SENSING TECHNOLOGIES XIV, 2017, 10215
  • [42] Single Walled Carbon Nanotube-Based Electrical Biosensor for the Label-Free Detection of Pathogenic Bacteria
    Yoo, Seung Min
    Baek, Youn-Kyoung
    Shin, SunHaeRa
    Kim, Ju-Hyun
    Jung, Hee-Tae
    Choi, Yang-Kyu
    Lee, Sang Yup
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2016, 16 (06) : 6520 - 6525
  • [43] Discrimination and detection of bacteria with a label-free impedimetric biosensor based on self-assembled lectin monolayer
    Xi, Fengna
    Gao, Jingqing
    Wang, Jine
    Wang, Zhenxin
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2011, 656 (1-2) : 252 - 257
  • [44] A Label-Free Electrochemical Detection of Cadmium Ions Using Aptamer-Based Biosensor
    Zaras, Izabela
    Kralka, Ewa
    Olszewski, Marcin
    Jarczewska, Marta
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2024, 171 (11)
  • [45] Label-Free DNA Hybridization Detection Using a Highly Sensitive Fiber Microcavity Biosensor
    Wu, Yao
    Wang, Guiyu
    Yu, Xiujuan
    Fan, Yuanji
    Chen, Xuefeng
    Liu, Shengchun
    SENSORS, 2024, 24 (01)
  • [46] Plasmon-Based Label-Free Biosensor Using Gold Nanosphere for Dengue Detection
    Mahmood, Hafiz Zeeshan
    Jilani, Asim
    Farooq, Sajid
    Javed, Yasir
    Jamil, Yasir
    Iqbal, Javed
    Ullah, Sami
    Wageh, Swelm
    CRYSTALS, 2021, 11 (11)
  • [47] Label-Free Biosensor Detection of Endocrine Disrupting Compounds Using Engineered Estrogen Receptors
    La Spina, Rita
    Ferrero, Valentina E. V.
    Aiello, Venera
    Pedotti, Mattia
    Varani, Luca
    Lettieri, Teresa
    Calzolai, Luigi
    Haasnoot, Willem
    Colpo, Pascal
    BIOSENSORS-BASEL, 2018, 8 (01):
  • [48] Label-free detection of alanine aminotransferase using a graphene field-effect biosensor
    Her, Jim-Long
    Pan, Tung-Ming
    Lin, Wan-Ying
    Wang, Kai-Sheng
    Li, Lain-Jong
    SENSORS AND ACTUATORS B-CHEMICAL, 2013, 182 : 396 - 400
  • [49] Label-free detection of DNA hybridization using nanopillar arrays based optical biosensor
    Chen, Jem-Kun
    Zhou, Gang-Yan
    Chang, Chi-Jung
    Cheng, Chih-Chia
    SENSORS AND ACTUATORS B-CHEMICAL, 2014, 194 : 10 - 18
  • [50] Label-Free and Selective Bacteria Detection Using a Film with Transferred Bacterial Configuration
    Tokonami, Shiho
    Nakadoi, Yu
    Takahashi, Mari
    Ikemizu, Mugihei
    Kadoma, Tetsuya
    Saimatsu, Kenta
    Le Quynh Dung
    Shiigi, Hiroshi
    Nagaoka, Tsutomu
    ANALYTICAL CHEMISTRY, 2013, 85 (10) : 4925 - 4929