Test-Retest Reliability of an Automated Infrared-Assisted Trunk Accelerometer-Based Gait Analysis System

被引:11
|
作者
Hsu, Chia-Yu [1 ,2 ]
Tsai, Yuh-Show [2 ]
Yau, Cheng-Shiang [2 ]
Shie, Hung-Hai [3 ]
Wu, Chu-Ming [1 ]
机构
[1] Ten Chan Gen Hosp, Dept Rehabil Med, 155 Yanping Rd, Taoyuan 320, Taiwan
[2] Chung Yuan Christian Univ, Dept Biomed Engn, 200 Zhongbei Rd, Taoyuan 320, Taiwan
[3] Ten Chan Gen Hosp, Dept Physiotherapy, 155 Yanping Rd, Taoyuan 320, Taiwan
关键词
accelerometry; gait; infrared; trunk acceleration; reliability; MINIMAL DETECTABLE CHANGE; ELDERLY-PEOPLE; HUMAN WALKING; ACCELERATION; PARAMETERS; DISORDERS; PREVALENCE; LOCATIONS; RECOVERY; DISEASE;
D O I
10.3390/s16081156
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The aim of this study was to determine the test-retest reliability of an automated infrared-assisted, trunk accelerometer-based gait analysis system for measuring gait parameters of healthy subjects in a hospital. Thirty-five participants (28 of them females; age range, 23-79 years) performed a 5-m walk twice using an accelerometer-based gait analysis system with infrared assist. Measurements of spatiotemporal gait parameters (walking speed, step length, and cadence) and trunk control (gait symmetry, gait regularity, acceleration root mean square (RMS), and acceleration root mean square ratio (RMSR)) were recorded in two separate walking tests conducted 1 week apart. Relative and absolute test-retest reliability was determined by calculating the intra-class correlation coefficient (ICC3,1) and smallest detectable difference (SDD), respectively. The test-retest reliability was excellent for walking speed (ICC = 0.87, 95% confidence interval = 0.74-0.93, SDD = 13.4%), step length (ICC = 0.81, 95% confidence interval = 0.63-0.91, SDD = 12.2%), cadence (ICC = 0.81, 95% confidence interval = 0.63-0.91, SDD = 10.8%), and trunk control (step and stride regularity in anterior-posterior direction, acceleration RMS and acceleration RMSR in medial-lateral direction, and acceleration RMS and stride regularity in vertical direction). An automated infrared-assisted, trunk accelerometer-based gait analysis system is a reliable tool for measuring gait parameters in the hospital environment.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Test-retest reliability of trunk accelerometric gait analysis
    Henriksen, M
    Lund, H
    Moe-Nilssen, R
    Bliddal, H
    Danneskiod-Samsoe, B
    GAIT & POSTURE, 2004, 19 (03) : 288 - 297
  • [2] Test-Retest Reliability and Concurrent Validity of a Single Tri-Axial Accelerometer-Based Gait Analysis in Older Adults with Normal Cognition
    Byun, Seonjeong
    Han, Ji Won
    Kim, Tae Hui
    Kim, Ki Woong
    PLOS ONE, 2016, 11 (07):
  • [3] Test-Retest Reliability of PODOSmart(R) Gait Analysis Insoles
    Loukovitis, Andreas
    Ziagkas, Efthymios
    Zekakos, Dimitrios Xypolias
    Petrelis, Alexandros
    Grouios, George
    SENSORS, 2021, 21 (22)
  • [4] Test-retest reliability of 3-dimensional computerized gait analysis
    Westhoff, B
    Hirsch, MA
    Hefter, H
    Wild, A
    Krauspe, R
    SPORTVERLETZUNG-SPORTSCHADEN, 2004, 18 (02) : 76 - 79
  • [5] Test-Retest Reliability of the MotionMetrix Software for the Analysis of Walking and Running Gait Parameters
    Jaen-Carrillo, Diego
    Ruiz-Alias, Santiago A.
    Chicano-Gutierrez, Jose M.
    Ruiz-Malagon, Emilio J.
    Roche-Seruendo, Luis E.
    Garcia-Pinillos, Felipe
    SENSORS, 2022, 22 (09)
  • [6] Test-retest reliability of quantitative gait analysis in people with Mild Cognitive Impairment
    Montero-Odasso, M. M.
    Casas, A.
    Hansen, K. T.
    Gutmanis, I.
    Wells, J.
    Bilski, P.
    Borrie, M.
    JOURNAL OF THE AMERICAN GERIATRICS SOCIETY, 2008, 56 (04) : S153 - S153
  • [7] Test-Retest Reliability of Sensor-Based Telemonitoring of Parkinsonian Gait at Home
    Kostikis, N.
    Rigas, G.
    Tachos, N.
    Tsamis, K.
    Kakalou, D.
    Konitsiotis, S.
    Fotiadis, D.
    MOVEMENT DISORDERS, 2020, 35 : S657 - S657
  • [8] Test-retest reliability of daily life gait speed as measured by smartphone global positioning system
    Obuchi, Shuichi P.
    Tsuchiya, Shuichi
    Kawai, Hisashi
    GAIT & POSTURE, 2018, 61 : 282 - 286
  • [9] Test-retest reliability of temporal and spatial gait characteristics measured with an instrumented walkway system (GAITRite)
    Van Uden C.J.T.
    Besser M.P.
    BMC Musculoskeletal Disorders, 5 (1)
  • [10] Test-retest reliability of temporal and spatial gait characteristics measured with an instrumented walkway system (GAITRite®)
    van Ulden, CJT
    Besser, MP
    BMC MUSCULOSKELETAL DISORDERS, 2004, 5