Prostate Image-Guided Radiotherapy by Megavolt Cone-Beam CT

被引:1
|
作者
Zucca, Sergio
Carau, Barbara
Solla, Ignazio
Garibaldi, Elisabetta
Farace, Paolo [1 ]
Lay, Giancarlo
Meleddu, Gianfranco
Gabriele, Pietro [1 ]
机构
[1] Reg Oncol Hosp, Dept Radiooncol, I-09121 Cagliari, Italy
关键词
Megavolt cone-beam CT; IGRT; Prostate; Radiotherapy; INTENSITY-MODULATED RADIOTHERAPY; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; FIDUCIAL MARKERS; GOLD MARKERS; BONY ANATOMY; LOCALIZATION; ONLINE; CANCER; SETUP;
D O I
10.1007/s00066-011-2241-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To test megavolt cone-beam CT (MV-CBCT) in order to evaluate setup errors in prostate radiotherapy. Patients and Methods: The setup of 9 patients was verified weekly by electronic portal imaging (EPI) and MV-CBCT, both performed in the same treatment session. EPI were compared with digitally reconstructed radiographies (DRRs). MV-CBCTs were matched to simulation CTs by manual registration based on bone markers (BMR), by manual registration based on soft tissues (STR) - rectum, bladder, and seminal vesicles - and by automatic registration (AR) performed by a mutual information algorithm. Shifts were evaluated along the three main axes: anteroposterior (AP), craniocaudal (CC), and laterolateral (LL). Finally, in 4 additional patients showing intraprostatic calcifications, the calcification mismatch error was used to evaluate the three MV-CBCT matching methods. Results: A total of 50 pairs of orthogonal EPIs and 50 MV-CBCTs were analyzed. Assuming an overall tolerance of 2 mm, no significant differences were observed comparing EPI vs BMR in any axis. A significant difference (p<0.001) was observed along the AP axis comparing EPI vs AR and EPI vs STR. On the calcification data set (22 measures), the calcification mismatch along the AP direction was significantly lower (p<0.05) after STR than after BMR or AR. Conclusion: Bone markers were not an effective surrogate of the target position and significant differences were observed comparing EPI or BMR vs STR, supporting the assessment of soft tissue position by MVCBs to verify and correct patient setup in prostate radiotherapy.
引用
收藏
页码:473 / 478
页数:6
相关论文
共 50 条
  • [1] Prostate Image-Guided Radiotherapy by Megavolt Cone-Beam CTBildgeführte Radiotherapie der Prostata mit Megavolt-Cone-Beam-CT
    Sergio Zucca
    Barbara Carau
    Ignazio Solla
    Elisabetta Garibaldi
    Paolo Farace
    Giancarlo Lay
    Gianfranco Meleddu
    Pietro Gabriele
    [J]. Strahlentherapie und Onkologie, 2011, 187 (8) : 473 - 478
  • [2] Cone-beam CT: Applications in image-guided external beam radiotherapy and brachytherapy
    Jaffray, DA
    Siewerdsen, JH
    Edmundson, GE
    Wong, JW
    Martinez, A
    [J]. PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2044 - 2044
  • [3] Towards optimization of megavoltage cone-beam CT for image-guided radiotherapy
    Pang, G
    O'Brien, P
    Mei, X
    Tambasco, M
    Yeboah, C
    Beachey, D
    Rowlands, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S623 - S623
  • [4] Patient dose in KV cone-beam CT image-guided radiotherapy
    Islam, M
    Purdie, T
    Moseley, D
    Alasti, H
    Norrlinger, B
    Sharpe, M
    Siewerdsen, J
    Jaffray, D
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1830 - 1830
  • [5] Automatic prostate localization on cone-beam CT scans for high precision image-guided radiotherapy
    Smitsmans, MHP
    de Bois, J
    Sonke, JJ
    Betgen, A
    Zijp, LJ
    Jaffray, DA
    Lebesque, JV
    van Herk, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (04): : 975 - 984
  • [6] Deformable image registration in cone-beam CT images for image-guided adaptive radiotherapy
    Zhang, L.
    Dong, L.
    Zhu, X.
    Garden, A.
    Ahamad, A.
    Ang, K.
    Mohan, R.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 1995 - 1995
  • [7] Hypofractionated online image-guided radiotherapy (IGRT) via cone-beam CT (CBCT) for prostate cancer
    Kestin, L.
    Yan, D.
    Ghilezan, M.
    Wu, Q.
    Wloch, J.
    Vicini, F.
    Martinez, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S169 - S169
  • [8] A personal perspective on the development and future of cone-beam CT for image-guided radiotherapy
    Jaffray, David A.
    [J]. MEDICAL PHYSICS, 2023, 50 (S1) : 54 - 57
  • [9] Cone-beam CT based image-guided RapidArc treatment of Prostate Cancer
    Hentschel, B.
    Strauss, D.
    Berger, K.
    Dorn, U.
    Oehler, W.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2013, 189 : 60 - 60
  • [10] Image artifacts caused by incorrect bowtie filters in cone-beam CT image-guided radiotherapy
    Cao, Yanan
    Ma, Tianjun
    de Boer, Steven F.
    Wang, Iris Z.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (07): : 153 - 159