Beam modeling and commissioning for Monte Carlo photon beam on an Elekta Versa HD LINAC

被引:2
|
作者
Can, Sumeyra [1 ]
Karacetin, Didem [2 ]
Meric, Niyazi [1 ]
机构
[1] Ankara Univ, Inst Nucl Sci, TR-06100 Ankara, Turkey
[2] Istanbul Basaksehir Cam & Sakura City Hosp, Istanbul, Turkey
关键词
Beam data measurements; Beam matching; Elekta VersaHD LINAC; Monte Carlo simulation; Monaco Treatment planning system; Virtual source model; TREATMENT PLANNING SYSTEM; DOSE CALCULATION TOOL; VIRTUAL SOURCE MODEL; CLINICAL IMPLEMENTATION; VALIDATION; IMRT; THERAPY;
D O I
10.1016/j.apradiso.2021.110054
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: This study aims at analyzing beam data commissioning along with verifying Monte Carlo-based treatment planning system on the basis of the manufacturer guidelines for Elekta Versa HD Linear Accelerator. Moreover, evaluating the beam match process in terms of quality index, photon profile and multi leaf collimator (MLC) offset is aimed as well. Materials and methods: The process of collecting beam data for Monaco 5.51 Treatment Planning System commissioning was done based on the instructions provided by the manufacturer as well as AAPM TG-106. Monte Carlo analysis was done for output factors in water, percent depth dose and beam profiles. A set of eight static and intensity modulated radiation therapy fields were used to verify the MLC parameters. Results: The difference between the measured and modeled penetrative quality (D-10) was achieved to be 0.54%. The output factors for 6 MV photon energy were measured and the difference between the measured and Monte Carlo output results was smaller than 1% for all the fields. The average percentage of passing the gamma criteria for commissioning test fields was (95+-4)%, however, the minimum agreement was 87.5% belonging to "7SEGA". Additionally, the agreement between both LINAC is 96%, however, the second LINAC reveals a positive offset in the point of approximately -4 cm on the x-axis at the crossplane. Conclusion: Test commissioning was successfully verified using a homogeneous phantom for point dose measurements, post modelling MLC parameters and patient QA plans. All plan parameters pass the gamma criteria. 6 MV photon beam was successfully commissioned for Elekta VersaHD LINAC and is ready for clinical implementation.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Commissioning of the FFF beam on the ELEKTA VERSA HD
    Pachoud, M.
    Ghandour, S.
    Matzinger, O.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (01) : 104 - 104
  • [2] Characteristics of the electron beam outside the applicator in an Elekta Versa HD Linac
    Maharaj, Kapil Dev
    Goodall, Simon
    Sabet, Mahsheed
    Dass, Joshua
    Ibrahim, Mounir
    Mahmood, Talat
    Rowshanfarzad, Pejman
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2025,
  • [3] Photon Spectra calculation for an Elekta linac beam using experimental scatter measurements and Monte Carlo techniques
    Juste, B.
    Miro, R.
    Campayo, J. M.
    Diez, S.
    Verdu, G.
    2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 3289 - +
  • [4] Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc
    Paschal, Holly M.
    Kabat, Christopher N.
    Papaconstadopoulos, Pavlos
    Kirby, Neil A.
    Myers, Pamela A.
    Wagner, Timothy D.
    Stathakis, Sotirios
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (09):
  • [5] Monte Carlo photon beam modeling and commissioning for radiotherapy dose calculation algorithm
    Toutaoui, A.
    Chikh, S. Ait
    Khelassi-Toutaoui, N.
    Hattali, B.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (07): : 833 - 837
  • [6] Beam Commissioning An Elekta Versa HD Linear Accelerator Using Accelerated Go Live Data
    Bassiri, N.
    Park, P.
    Yamamoto, T.
    Sensoy, L.
    Palefsky, S.
    Keohane, M.
    Benedict, S.
    MEDICAL PHYSICS, 2022, 49 (06) : E627 - E627
  • [7] Measurement of Photon Beam Energy Changes On An Elekta Versa HD Using 5 Methods
    Pavord, D.
    Yun, K.
    Berzanske, J.
    Denhoff, S.
    Goss, M.
    MEDICAL PHYSICS, 2022, 49 (06) : E936 - E936
  • [8] Commissioning an Elekta Versa HD linear accelerator
    Narayanasamy, Ganesh
    Saenz, Daniel
    Cruz, Wilbert
    Ha, Chul S.
    Papanikolaou, Niko
    Stathakis, Sotirios
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 179 - 191
  • [9] Validation of the Elekta Synergy Platform Linac at 6 MV Photon Beam using PRIMO Monte Carlo Software
    Marzon, Jan Risty Lucido
    Convicto, Vernie C.
    Lapening, Melbagrace A.
    Berondo, Andelson L.
    Bacala, Angelina M.
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (03) : 410 - 418
  • [10] Commissioning of a surface imaging system for beam gating at an Elekta Linac
    Batista, V.
    Albert, S.
    Rhein, B.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1105 - S1105