Precise 13CO2/12CO2 isotopic ratio measurements for breath diagnosis with a 2 μm diode laser

被引:0
|
作者
Sun Mingguo [1 ,2 ]
Ma Hongliang [1 ]
Cao Zhensong [1 ]
Liu Kun [2 ]
Wang Guishi [2 ]
Wang Lei [2 ]
Liu Qiang [1 ]
Gao Xiaoming [2 ]
机构
[1] Chinese Acad Sci, Key Lab Atmospher Composit & Opt Radiat, Hefei 230031, Peoples R China
[2] Chinese Acad Sci, Anhui Inst Optic Fine Mech, Lab Atmospher Physico Chem, Hefei 230031, Peoples R China
关键词
Stable isotopes; Laser spectroscopy; Human breath; Wavelet denoising; Kalman filter; RING-DOWN SPECTROSCOPY; C-13/C-12;
D O I
10.1117/12.2071221
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A laser spectrometer based on a distributed-feedback semiconductor diode laser at 2 mu m is developed to measure the changes of (CO2)-C-13/(CO2)-C-12 isotope ratio in exhaled breath sample with the CO2 concentration of similar to 4%. It is characterized by a simplified optical layout, in which a single detector and associated electronics are used to probe CO2 spectrum. The cell has 10 cm long base length with 26.4 m optical path length in total and 330 cm(3) volume. The cell pressure and temperature are controlled at 50 Torr and 28 degrees C, respectively. The best (13)delta precision of 0.06 parts per thousand was achieved by using wavelet denoising and Kalman filter.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Relative line intensity measurement in absorption spectra using a tunable diode laser at 1.6 μm: application to the determination of 13CO2/12CO2 isotope ratio
    R. Chaux
    B. Lavorel
    Applied Physics B, 2001, 72 : 237 - 240
  • [22] Relative line intensity measurement in absorption spectra using a tunable diode laser at 1.6 μm:: application to the determination of 13CO2/12CO2 isotope ratio
    Chaux, R
    Lavorel, B
    APPLIED PHYSICS B-LASERS AND OPTICS, 2001, 72 (02): : 237 - 240
  • [23] Real Time Sensing of 12CO2 and 13CO2 Using 2μm DFB-LD
    Amamoto, Kanto
    Tei, Kazuyoku
    Yamaguchi, Shigeru
    Sakai, Shunichi
    Asobe, Masaki
    Ohba, Takeshi
    2017 22ND MICROOPTICS CONFERENCE (MOC), 2017, : 210 - 211
  • [24] The ν1+5ν3 dyad of 12CO2 and 13CO2
    Campargue, A
    Bailly, D
    Teffo, JL
    Tashkun, SA
    Perevalov, VI
    JOURNAL OF MOLECULAR SPECTROSCOPY, 1999, 193 (01) : 204 - 212
  • [25] High-Precision 13CO2/12CO2 Isotopic Ratio Measurement Using Tunable Diode Laser Absorption Spectroscopy at 4.3 μm for Deep-Sea Natural Gas Hydrate Exploration
    Zhang, Hanquan
    Wen, Mingming
    Li, Yonghang
    Wan, Peng
    Chen, Chen
    APPLIED SCIENCES-BASEL, 2019, 9 (17):
  • [26] Retrieving the 13CO2/12CO2 isotopic ratio from high resolution ground-based FTIR measurement in Tsukuba
    Rokotyan, N. V.
    Imasu, R.
    Morina, I.
    Gribanov, K. G.
    Zakharov, V. I.
    24TH INTERNATIONAL SYMPOSIUM ON ATMOSPHERIC AND OCEAN OPTICS: ATMOSPHERIC PHYSICS, 2018, 10833
  • [27] Evaluating calibration strategies for isotope ratio infrared spectroscopy for atmospheric 13CO2/12CO2 measurement
    Wen, X. -F.
    Meng, Y.
    Zhang, X. -Y.
    Sun, X. -M.
    Lee, X.
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2013, 6 (06) : 1491 - 1501
  • [28] Toward the quantification of the 13CO2/12CO2 ratio in exhaled mouse breath with mid-infrared hollow waveguide gas sensors
    Andreas Wilk
    Felicia Seichter
    Seong-Soo Kim
    Erhan Tütüncü
    Boris Mizaikoff
    Josef A. Vogt
    Ulrich Wachter
    Peter Radermacher
    Analytical and Bioanalytical Chemistry, 2012, 402 : 397 - 404
  • [29] Toward the quantification of the 13CO2/12CO2 ratio in exhaled mouse breath with mid-infrared hollow waveguide gas sensors
    Wilk, Andreas
    Seichter, Felicia
    Kim, Seong-Soo
    Tuetuencue, Erhan
    Mizaikoff, Boris
    Vogt, Josef A.
    Wachter, Ulrich
    Radermacher, Peter
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 402 (01) : 397 - 404
  • [30] Spectroscopy and intensity measurements of the 3ν1+3ν3 tetrad of 12CO2 and 13CO2
    Weirauch, G
    Campargue, A
    JOURNAL OF MOLECULAR SPECTROSCOPY, 2001, 207 (02) : 263 - 268