Multiresolution iterative reconstruction in high-resolution extremity cone-beam CT

被引:24
|
作者
Cao, Qian [1 ]
Zbijewski, Wojciech [1 ]
Sisniega, Alejandro [1 ]
Yorkston, John [2 ]
Siewerdsen, Jeffrey H. [1 ,3 ]
Stayman, J. Webster [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Carestream Hlth, Rochester, NY 14608 USA
[3] Johns Hopkins Univ, Russell H Morgan Dept Radiol, Baltimore, MD 21205 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2016年 / 61卷 / 20期
基金
美国国家卫生研究院;
关键词
CBCT; model-based reconstruction; multiresolution; bone; extremity; IMAGE QUALITY; TOMOGRAPHIC RECONSTRUCTION; BREAST CT; ALGORITHMS; REGION; SYSTEM; DESIGN; MODELS;
D O I
10.1088/0031-9155/61/20/7263
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Application of model-based iterative reconstruction (MBIR) to high resolution cone-beam CT (CBCT) is computationally challenging because of the very fine discretization (voxel size <100 mu m) of the reconstructed volume. Moreover, standard MBIR techniques require that the complete transaxial support for the acquired projections is reconstructed, thus precluding acceleration by restricting the reconstruction to a region-of-interest. To reduce the computational burden of high resolution MBIR, we propose a multiresolution penalized-weighted least squares (PWLS) algorithm, where the volume is parameterized as a union of fine and coarse voxel grids as well as selective binning of detector pixels. We introduce a penalty function designed to regularize across the boundaries between the two grids. The algorithm was evaluated in simulation studies emulating an extremity CBCT system and in a physical study on a test-bench. Artifacts arising from the mismatched discretization of the fine and coarse sub-volumes were investigated. The fine grid region was parameterized using 0.15 mm voxels and the voxel size in the coarse grid region was varied by changing a downsampling factor. No significant artifacts were found in either of the regions for downsampling factors of up to 4x. For a typical extremities CBCT volume size, this downsampling corresponds to an acceleration of the reconstruction that is more than five times faster than a brute force solution that applies fine voxel parameterization to the entire volume. For certain configurations of the coarse and fine grid regions, in particular when the boundary between the regions does not cross high attenuation gradients, downsampling factors as high as 10x can be used without introducing artifacts, yielding a similar to 50x speedup in PWLS. The proposed multiresolution algorithm significantly reduces the computational burden of high resolution iterative CBCT reconstruction and can be extended to other applications of MBIR where computationally expensive, high-fidelity forward models are applied only to a sub-region of the field-of-view.
引用
收藏
页码:7263 / 7281
页数:19
相关论文
共 50 条
  • [21] Efficacy of high-resolution cone-beam CT in the evaluation of perforators in vertebral artery dissection
    Shimada, Kenji
    Tanaka, Michihiro
    Kadooka, Keisuke
    Hadeishi, Hiromu
    INTERVENTIONAL NEURORADIOLOGY, 2017, 23 (04) : 350 - 356
  • [22] High-resolution helical cone-beam micro-CT with theoretically-exact reconstruction from experimental data
    Varslot, T.
    Kingston, A.
    Myers, G.
    Sheppard, A.
    MEDICAL PHYSICS, 2011, 38 (10) : 5459 - 5476
  • [23] Reconstruction of a cone-beam CT image via forward iterative projection matching
    Brock, R. Scott
    Docef, Alen
    Murphy, Martin J.
    MEDICAL PHYSICS, 2010, 37 (12) : 6212 - 6220
  • [24] Reconstruction of a Cone-Beam CT Image Via Forward Iterative Projection Matching
    Brock, R.
    Docef, A.
    Murphy, M.
    MEDICAL PHYSICS, 2010, 37 (06) : 3112 - +
  • [25] Motion compensated iterative reconstruction of a region of interest in cardiac cone-beam CT
    Isola, A. A.
    Ziegler, A.
    Schaefer, D.
    Koehler, T.
    Niessen, W. J.
    Grass, M.
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2010, 34 (02) : 149 - 159
  • [26] Feasibility of iterative cone-beam CT reconstruction for adaptive radiotherapy: a preliminary study
    Martin, Peter
    Thind, Kundan
    MEDICAL PHYSICS, 2019, 46 (11) : 5391 - 5391
  • [27] Cone-beam iterative reconstruction of a segment of a long object
    Zeng, GSL
    Gullberg, GT
    Christian, PE
    Gagnon, D
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) : 37 - 41
  • [28] Dedicated Cone-Beam CT System for Extremity Imaging
    Carrino, John A.
    Al Muhit, Abdullah
    Zbijewski, Wojciech
    Thawait, Gaurav K.
    Stayman, J. Webster
    Packard, Nathan
    Senn, Robert
    Yang, Dong
    Foos, David H.
    Yorkston, John
    Siewerdsen, Jeffrey H.
    RADIOLOGY, 2014, 270 (03) : 816 - 824
  • [29] A FAST ITERATIVE RECONSTRUCTION ALGORITHM FOR CONE-BEAM SPECT
    LI, J
    JASZCZAK, RJ
    GREER, KL
    COLEMAN, RE
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (05) : P80 - P80
  • [30] Exact Interior Reconstruction with Cone-Beam CT
    Ye, Yangbo
    Yu, Hengyong
    Wang, Ge
    INTERNATIONAL JOURNAL OF BIOMEDICAL IMAGING, 2007, 2007