Analytical performance validation and clinical application of blood gas analyzer on the detection of neonatal bilirubin

被引:3
|
作者
Kuang, Zhaozhong [1 ]
Zong, Xiaolong [2 ]
Xing, Shuang [3 ]
Zhao, Fuguang [3 ]
Guo, Shanshan [3 ]
Li, Huiqiang [1 ]
Wei, Dianjun [4 ]
机构
[1] Tianjin Med Univ, Sch Med Lab Sci, Tianjin, Peoples R China
[2] Tianjin Med Univ, Hosp 2, Dept Lab, Tianjin, Peoples R China
[3] Tanggu Obstet & Gynecol Hosp, Dept Lab, Binhai New Area, Tianjin, Peoples R China
[4] Yanda Hosp, Dept Lab, Langfang, Hebei, Peoples R China
关键词
Newborn; bilirubin; blood gas analyzer; biochemical analyzer; consistency; HYPERBILIRUBINEMIA;
D O I
10.21037/tp-21-541
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background: It is not yet clear whether the trace blood gas analyzer can be used for biochemical detection of newborns. This study aimed to evaluate the reliability of the method for the detection of bilirubin in infants. Method: Based on the CLSI EP15-A2 document, the analytical performance of the blood gas analyzer method for bilirubin detection in neonates was validated. The resulting data of 363 simultaneous bilirubin detection with blood gas analyzer (optical method) and biochemical analyzer (enzymatic method) were reviewed. According to the CLSI EP9-A3 document, the relevance and consistency of the measurement results were evaluated by Pearson correlation analysis, Passing-Bablok regression, and Bland-Altman deviation analysis. Results: The precision and accuracy of the Werfen GEM 4000 blood gas analyzer for the detection of different levels of bilirubin samples adhered to the manufacturer's statement and industry quality standards. The bilirubin detection values of the 2 methods showed a good correlation, and both of them were significantly correlated (P<0.001). Passing-Bablok regression results showed that the regression equation of the bilirubin detection value of the 2 methods is y = -21.00 + 1.17x, with the slope as 1.17 [95% confidence interval (CI): 1.15 to 1.19], and the intercept was -21.00 (95% CI: -23.62 to -18.71), the data of the 2 sets were not consistent in each concentration range. The Bland-Altman plot demonstrated that the bilirubin detection value of 16/363 cases (4.4%) for the 2 methods exceeded the 95% limits of agreement (95% LoA); of which the maximum bias was -30.34 (95% CI: -38.48 to -22.26) and there were 5/76 cases (6.6%) outside the 95% LoA in the >300 mu mol/L group. Conclusions: The method for detecting neonatal total bilirubin by trace blood gas analyzer basically meets the clinical requirements and can be used for the preliminary screening of neonatal jaundice. However, for severe hyperbilirubinemia that requires close monitoring of dynamics, a precise enzymatic quantification is required.
引用
收藏
页码:3175 / 3183
页数:9
相关论文
共 50 条
  • [1] Evaluation of neonatal bilirubin method on RapidLab 500 blood gas analyzer
    Vranken, L.
    Pieltain, C.
    Cavalier, E.
    ACTA CLINICA BELGICA, 2017, 72 : 8 - 8
  • [2] Point-of-care determination of neonatal bilirubin with the blood gas analyzer RapidLab 1265
    Mielsch, Caroline
    Zimmermann, Andrea
    Wagner, Dieter
    Matthes, Bianca
    Schlebusch, Harald
    Luppa, Peter B.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (10) : 1455 - 1461
  • [3] CLINICAL AND ANALYTICAL EVALUATION OF THE EPOC BLOOD-GAS ANALYZER
    Carraro, P.
    Vigan, S.
    Spadaro, S.
    Zilio, M.
    Navoni, F.
    Vettore, G.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S673 - S673
  • [4] Analytical performance of a portable critical care blood gas analyzer
    Schlebusch, H
    Paffenholz, I
    Zerback, R
    Leinberger, R
    CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 107 - 112
  • [5] Measurement of neonatal total bilirubin using the Radiometer ABL90® blood gas analyzer
    Lano, Ian Marie
    Lyon, Martha
    CLINICAL BIOCHEMISTRY, 2015, 48 (15) : 1010 - 1010
  • [6] Analytical Performance of the ABL90 FLEX Blood Gas Analyzer
    Seeger, Corina
    Kawiecki, Ronald W.
    Kristensen, Hans Bjarne
    POINT OF CARE, 2011, 10 (03): : 108 - 115
  • [7] Total bilirubin measurement by photometry on a blood gas analyzer:: Potential for use in neonatal testing at the point of care
    Rolinski, B
    Küster, H
    Ugele, B
    Gruber, R
    Horn, K
    CLINICAL CHEMISTRY, 2001, 47 (10) : 1845 - 1847
  • [8] Analytical performance assessment of a novel cartridge-based blood gas analyzer
    Cervera, Jose
    Mansouri, Sohrab
    Pamidi, Prasad V. A.
    CLINICAL BIOCHEMISTRY, 2019, 63 : 154 - 155
  • [9] Accuracy and Reliability of Whole Blood Bilirubin Measurements Using a Roche Blood Gas Analyzer for Neonatal Hyperbilirubinemia Screening and Risk Stratification
    Wang, Qing
    Zhang, Tianyi
    Lin, Yuanxi
    Jiang, Li
    Zhou, Wenlong
    Zong, Xiaolong
    FRONTIERS IN PEDIATRICS, 2022, 10
  • [10] Reply to "Analytical performance assessment of a novel cartridge-based blood gas analyzer"
    Aloisio, Elena
    Dolci, Alberto
    Panteghini, Mauro
    CLINICAL BIOCHEMISTRY, 2019, 63 : 156 - 157