Accuracy and Reliability of Whole Blood Bilirubin Measurements Using a Roche Blood Gas Analyzer for Neonatal Hyperbilirubinemia Screening and Risk Stratification

被引:3
|
作者
Wang, Qing [1 ]
Zhang, Tianyi [2 ]
Lin, Yuanxi [2 ]
Jiang, Li [1 ]
Zhou, Wenlong [3 ]
Zong, Xiaolong [4 ]
机构
[1] Gen Hosp Tianjin Med Univ, Dept Clin Lab, Tianjin, Peoples R China
[2] Tianjin Med Univ, Dept Emergency Med, Hosp 2, Tianjin, Peoples R China
[3] Tianjin Med Univ, Coll Med Lab Sci, Tianjin, Peoples R China
[4] Tianjin Med Univ, Dept Clin Lab, Hosp 2, Tianjin, Peoples R China
来源
FRONTIERS IN PEDIATRICS | 2022年 / 10卷
关键词
neonatal hyperbilirubinemia; bilirubin measurement; whole-blood bilirubin; accuracy; clinical reliability; SERUM BILIRUBIN;
D O I
10.3389/fped.2022.910566
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
BackgroundAccurate bilirubin measurements are essential for appropriate management of neonatal hyperbilirubinemia. This study aimed to evaluate the accuracy and reliability of whole blood bilirubin measurements obtained using a Roche blood gas analyzer (Roche TBiL), with total serum bilirubin (TSB) measurements determined by the Ortho VITROS 4600 chemistry system (Ortho TSB) serving as a reference. Materials and MethodsMedical records of hospitalized neonates that underwent simultaneous Roche TBiL and Ortho TSB measurements were reviewed for eligibility selection and data collection. The correlations and differences between two sets of results were determined using Passing-Bablok regression analysis and a Bland-Altman plot, respectively. For eligible newborns, the risk of developing severe hyperbilirubinemia was assessed using the Bhutani nomogram. Weighted kappa analysis was used to evaluate the agreement between risk prediction by the two methods. ResultsWe obtained 618 paired Roche TBiL and Ortho TSB results from 309 neonates. Roche TBiL and Ortho TSB measurements showed a good correlation (r = 0.923; 95% CI: 0.905-0.938). Passing-Bablok regression analysis yielded the following equation: Roche TBiL = 0.794 x Ortho TSB + 1.255 mg/dL, with a slope of 0.794 (95% CI: 0.763-0.825) and intercept of 1.255 (95% CI: 1.042-1.417). The average difference between the two methods was 0.1 +/- 1.448 mg/dL. A total of 207 neonates were eligible for evaluation of the agreement between the risk-grading methods. Although kappa analysis showed good agreement between the methods, with a weighted kappa of 0.681 (95% CI: 0.610-0.751) across all populations, the values for approximately half of the neonates at intermediate and high risk of hyperbilirubinemia (33/72) were underestimated by Roche TBiL. ConclusionOur results indicate that Roche TBiL and Ortho TSB measurements in the neonatal population are not consistent. As a point-of-care and trace blood assay, Roche blood gas bilirubin measurements can facilitate primary screening of neonatal hyperbilirubinemia, but it seems to lack accuracy regarding risk stratification, particularly for high-risk newborn individuals.
引用
收藏
页数:6
相关论文
共 38 条
  • [1] Reliability of Total Bilirubin Measurements in Whole Blood from Preterm Neonates Using a Blood Gas Analyzer
    Nambara, Toshihiko
    Katayama, Yoshinori
    Enomoto, Masahiro
    Kikuchi, Shin
    Takei, Atsuko
    Ikegami, Hitoshi
    Minami, Hirotaka
    Lee, Yong Kye
    CLINICAL LABORATORY, 2016, 62 (11) : 2285 - 2289
  • [2] Evaluation of ABL90® whole blood bilirubin as a screening tool for neonatal hyperbilirubinemia
    Trowski, Leslie
    Wang, Li
    CLINICAL BIOCHEMISTRY, 2016, 49 (18)
  • [3] Accuracy of neonatal venous blood glucose measurements using blood gas analyzer compared with central laboratory chemistry analyzer
    Janjindamai, Waricha
    Tiwawatpakorn, Nichanan
    Thatrimontrichai, Anucha
    Dissaneevate, Supaporn
    Maneenil, Gunlawadee
    Phatigomet, Manapat
    JOURNAL OF CLINICAL NEONATOLOGY, 2022, 11 (01) : 7 - 12
  • [4] Blood Gas Analyzer Accuracy of Glucose Measurements
    Liang, Yafen
    Wanderer, Jonathan
    Nichols, James H.
    Klonoff, David
    Rice, Mark J.
    MAYO CLINIC PROCEEDINGS, 2017, 92 (07) : 1030 - 1041
  • [5] Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
    Huang, Yun
    Dean, Robert
    Dubbelman, Yvonne
    Vincent, Anne
    Khurshid, Faiza
    PRACTICAL LABORATORY MEDICINE, 2021, 25
  • [6] Evaluation of neonatal bilirubin method on RapidLab 500 blood gas analyzer
    Vranken, L.
    Pieltain, C.
    Cavalier, E.
    ACTA CLINICA BELGICA, 2017, 72 : 8 - 8
  • [7] Measurement of neonatal total bilirubin using the Radiometer ABL90® blood gas analyzer
    Lano, Ian Marie
    Lyon, Martha
    CLINICAL BIOCHEMISTRY, 2015, 48 (15) : 1010 - 1010
  • [8] Cord blood bilirubin and prediction of neonatal hyperbilirubinemia and perinatal infection in newborns at risk of hemolysis
    Kardum, Darjan
    Serdarusic, Ivana
    Biljan, Borna
    Santic, Kresimir
    Zivkovic, Vinko
    Kos, Martina
    JORNAL DE PEDIATRIA, 2021, 97 (04) : 440 - 444
  • [9] Evaluation of Total Bilirubin Determination in Neonatal Whole-Blood Samples by Multiwavelength Photometry on the Roche OMNI S Point-of-Care Analyzer
    Rolinski, Boris
    Okorodudu, Anthony O.
    Kost, Gerald
    Roser, Markus
    Wu, Jiaxi
    Goerlach-Graw, Ada
    Kuester, Helmut
    POINT OF CARE, 2005, 4 (01): : 3 - 8
  • [10] Accuracy of Potassium Measurement Using Blood Gas Analyzer
    Mahmoud, Hatim
    Jaffar, Zied
    Al Alawi, Yousef M.
    Al Alsuhaimi, Fatimah
    Khoja, Mohammed A. A.
    Al-Ahmadi, Muath A.
    Alattas, Abdullah M.
    Alhusayni, Mohammed F.
    Mahroos, Mohammed E.
    Alrehaili, Muath A.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (03)