Accuracy and Reliability of Whole Blood Bilirubin Measurements Using a Roche Blood Gas Analyzer for Neonatal Hyperbilirubinemia Screening and Risk Stratification

被引:3
|
作者
Wang, Qing [1 ]
Zhang, Tianyi [2 ]
Lin, Yuanxi [2 ]
Jiang, Li [1 ]
Zhou, Wenlong [3 ]
Zong, Xiaolong [4 ]
机构
[1] Gen Hosp Tianjin Med Univ, Dept Clin Lab, Tianjin, Peoples R China
[2] Tianjin Med Univ, Dept Emergency Med, Hosp 2, Tianjin, Peoples R China
[3] Tianjin Med Univ, Coll Med Lab Sci, Tianjin, Peoples R China
[4] Tianjin Med Univ, Dept Clin Lab, Hosp 2, Tianjin, Peoples R China
来源
FRONTIERS IN PEDIATRICS | 2022年 / 10卷
关键词
neonatal hyperbilirubinemia; bilirubin measurement; whole-blood bilirubin; accuracy; clinical reliability; SERUM BILIRUBIN;
D O I
10.3389/fped.2022.910566
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
BackgroundAccurate bilirubin measurements are essential for appropriate management of neonatal hyperbilirubinemia. This study aimed to evaluate the accuracy and reliability of whole blood bilirubin measurements obtained using a Roche blood gas analyzer (Roche TBiL), with total serum bilirubin (TSB) measurements determined by the Ortho VITROS 4600 chemistry system (Ortho TSB) serving as a reference. Materials and MethodsMedical records of hospitalized neonates that underwent simultaneous Roche TBiL and Ortho TSB measurements were reviewed for eligibility selection and data collection. The correlations and differences between two sets of results were determined using Passing-Bablok regression analysis and a Bland-Altman plot, respectively. For eligible newborns, the risk of developing severe hyperbilirubinemia was assessed using the Bhutani nomogram. Weighted kappa analysis was used to evaluate the agreement between risk prediction by the two methods. ResultsWe obtained 618 paired Roche TBiL and Ortho TSB results from 309 neonates. Roche TBiL and Ortho TSB measurements showed a good correlation (r = 0.923; 95% CI: 0.905-0.938). Passing-Bablok regression analysis yielded the following equation: Roche TBiL = 0.794 x Ortho TSB + 1.255 mg/dL, with a slope of 0.794 (95% CI: 0.763-0.825) and intercept of 1.255 (95% CI: 1.042-1.417). The average difference between the two methods was 0.1 +/- 1.448 mg/dL. A total of 207 neonates were eligible for evaluation of the agreement between the risk-grading methods. Although kappa analysis showed good agreement between the methods, with a weighted kappa of 0.681 (95% CI: 0.610-0.751) across all populations, the values for approximately half of the neonates at intermediate and high risk of hyperbilirubinemia (33/72) were underestimated by Roche TBiL. ConclusionOur results indicate that Roche TBiL and Ortho TSB measurements in the neonatal population are not consistent. As a point-of-care and trace blood assay, Roche blood gas bilirubin measurements can facilitate primary screening of neonatal hyperbilirubinemia, but it seems to lack accuracy regarding risk stratification, particularly for high-risk newborn individuals.
引用
收藏
页数:6
相关论文
共 38 条
  • [31] Clinical Impact of Point-of-Care Testing Using the OMNI-S Blood Gas Analyzer in a Neonatal Intensive Care Setting
    Arthurs, Owen
    Pattnayak, Santosh
    Bewley, Barbara
    Kelsall, Wilf
    POINT OF CARE, 2010, 9 (01): : 21 - 24
  • [32] Accuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients under General Anesthesia in the Operating Room
    Karon, Brad S.
    Donato, Leslie J.
    Larsen, Chelsie M.
    Siebenaler, Lindsay K.
    Wells, Amy E.
    Wood-Wentz, Christina M.
    Shirk-Marienau, Mary E.
    Curry, Timothy B.
    ANESTHESIOLOGY, 2017, 127 (03) : 466 - 474
  • [33] Lung cancer screening in high-risk subjects: early detection with LDCT and risk stratification using miRNA-based blood test
    Sestini, Stefano
    Boeri, Mattia
    Marchiano, Alfonso
    Silva, Mario
    Calareso, Giuseppina
    Galeone, Carlotta
    Sozzi, Gabriella
    Pastorino, Ugo
    EPIDEMIOLOGIA & PREVENZIONE, 2016, 40 (01): : 42 - 50
  • [34] Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function
    Lim, Ha-Jin
    Lee, Seung-Yeob
    Choi, Hyun-Jung
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (05)
  • [35] Rapid and reliable screening method for detection of 70 pesticides in whole blood by gas chromatography-mass spectrometry using a constructed calibration-locking database
    Kudo, Keiko
    Nagamatsu, Kumi
    Umehara, Takahiro
    Usumoto, Yosuke
    Sameshima, Naomi
    Tsuji, Akiko
    Ikeda, Noriaki
    LEGAL MEDICINE, 2012, 14 (02) : 93 - 100
  • [36] SIMULTANEOUS AND INDIVIDUAL MEASUREMENTS OF T4 AND TSH FOR SCREENING FOR NEONATAL HYPO-THYROIDISM USING A SINGLE 3-MM DISK OF DRIED BLOOD ON FILTER-PAPER
    IKEDA, T
    ISHIBASHI, K
    OHSAWA, R
    NAGATAKI, S
    ANNALES D ENDOCRINOLOGIE, 1979, 40 (05) : A9 - A9
  • [37] Screening for cytomegalovirus (CMV) infection in allogeneic bone marrow transplantation using a quantitative whole blood polymerase chain reaction (PCR) method: analysis of potential risk factors for CMV infection
    Qamruddin, AO
    Oppenheim, BA
    Guiver, M
    Mutton, KJ
    Chopra, R
    BONE MARROW TRANSPLANTATION, 2001, 27 (03) : 301 - 306
  • [38] Screening for cytomegalovirus (CMV) infection in allogeneic bone marrow transplantation using a quantitative whole blood polymerase chain reaction (PCR) method: analysis of potential risk factors for CMV infection
    AO Qamruddin
    BA Oppenheim
    M Guiver
    KJ Mutton
    R Chopra
    Bone Marrow Transplantation, 2001, 27 : 301 - 306