Estimation of hematocrit by means of dual-gate power Doppler

被引:0
|
作者
Nowicki, A [1 ]
Secomski, W [1 ]
机构
[1] Polish Acad Sci, Inst Fundamental Technol Res, Dept Ultrasound, PL-00049 Warsaw, Poland
关键词
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Noninvasive estimation of the blood hematocrit is important for the medical diagnosis. At present, there is only one reliable method based on measurement of speed of sound. It is mainly used during patient's dialysis. We propose a novel approach to solve this problem. The instrumentation developed is based on the dependence of the absorption of ultrasonic wave in blood. The instrument is based on pulsed Doppler ultrasonic blood flowmeter, operating at frequency 20 MHz. Doppler signal is simultaneously recorded in two gates at two distinct depths. The ratio of signal power from two gates depends on hematocrit only. The preliminary measurements were done in vitro using porcine blood. The Doppler signal was recorded in two gates separated by 1.2 mm. and 3.6 mm. A good agreement of hematocrit values calculated from Doppler signal with laboratory measurements was obtained: for sample volumes separated by 3.6 mrn correlation coefficients were equal to 0.96 for p <0.001. This novel method offers the possibility of direct monitoring of hematocrit during dialysis.