Storage Phosphors for Medical Imaging

被引:157
|
作者
Leblans, Paul [1 ]
Vandenbroucke, Dirk [1 ]
Willems, Peter [2 ]
机构
[1] Agfa HealthCare NV, B-2640 Mortsel, Belgium
[2] Ind Sci & Comp BVBA, B-9190 Stekene, Belgium
关键词
computed radiography; storage phosphor; X-ray; optically stimulated lumi-nescence; BaFBr:Eu2+; CsBr:Eu2+; needle imaging plate; image quality; COMPUTED RADIOGRAPHY SYSTEMS; RADIATION-INDUCED DEFECTS; PHOTOSTIMULATED LUMINESCENCE; STIMULATED LUMINESCENCE; HOLE-CENTERS; F-CENTERS; DIGITAL RADIOGRAPHY; ELECTRON; PERFORMANCE; MECHANISM;
D O I
10.3390/ma4061034
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Computed radiography (CR) uses storage phosphor imaging plates for digital imaging. Absorbed X-ray energy is stored in crystal defects. In read-out the energy is set free as blue photons upon optical stimulation. In the 35 years of CR history, several storage phosphor families were investigated and developed. An explanation is given as to why some materials made it to the commercial stage, while others did not. The photo stimulated luminescence mechanism of the current commercial storage phosphors, BaFBr:Eu2+ and CsBr:Eu2+ is discussed. The relation between storage phosphor plate physical characteristics and image quality is explained. It is demonstrated that the morphology of the phosphor crystals in the CR imaging plate has a very significant impact on its performance.
引用
收藏
页码:1034 / 1086
页数:53
相关论文
共 50 条
  • [1] Scintillator phosphors for medical imaging
    General Electric Corporate Research, and Development Cent, Schenectady, United States
    Electrochem Soc Interface, 2 (34-38):
  • [2] A review report on medical imaging phosphors
    Kaur, Jagjeet
    Parganiha, Yogita
    Dubey, Vikas
    Singh, Deepti
    RESEARCH ON CHEMICAL INTERMEDIATES, 2014, 40 (08) : 2837 - 2858
  • [3] A review report on medical imaging phosphors
    Jagjeet Kaur
    Yogita Parganiha
    Vikas Dubey
    Deepti Singh
    Research on Chemical Intermediates, 2014, 40 : 2837 - 2858
  • [4] Thermal neutron imaging with CsBr storage phosphors
    Winch, N. M.
    Edgar, A.
    Bartle, C. M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 763 : 394 - 398
  • [5] RADIOGRAPHIC IMAGING WITH STORAGE PHOSPHORS - IMAGE QUALITY
    KLINGENBECK, K
    CONRAD, B
    SIEMENS FORSCHUNGS-UND ENTWICKLUNGSBERICHTE-SIEMENS RESEARCH AND DEVELOPMENT REPORTS, 1987, 16 (05): : 192 - 198
  • [6] HIGH-RESOLUTION DIGITAL IMAGING WITH STORAGE PHOSPHORS
    FUHRMAN, CR
    GUR, D
    SCHAETZING, R
    JOURNAL OF THORACIC IMAGING, 1990, 5 (01) : 21 - 30
  • [7] MEDICAL IMAGING APPLICATIONS OF X-RAY PHOSPHORS
    ISSLER, SL
    BEUTEL, J
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1992, 204 : 66 - IEC
  • [8] Transparent caesium bromide storage-phosphors for radiation imaging
    Winch, Nicola M.
    Edgar, Andrew
    PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 2012, 209 (12): : 2427 - 2432
  • [9] PHYSICAL-MECHANISMS OF PHOTOSTIMULATION IN MEDICAL X-RAY STORAGE PHOSPHORS
    VONSEGGERN, H
    VOIGT, T
    SCHWARZMICHEL, K
    SIEMENS FORSCHUNGS-UND ENTWICKLUNGSBERICHTE-SIEMENS RESEARCH AND DEVELOPMENT REPORTS, 1988, 17 (03): : 125 - 130
  • [10] EFFICIENCY OF STORAGE PHOSPHORS
    CARUTHERS, EB
    HSU, WY
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1986, 133 (08) : C339 - C339