Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner

被引:20
|
作者
Delso, Gaspar
Fuerst, Sebastian
Jakoby, Bjoern [1 ,2 ]
Ladebeck, Ralf [3 ]
Ganter, Carl [4 ]
Nekolla, Stephan G.
Schwaiger, Markus
Ziegler, Sibylle I.
机构
[1] Univ Surrey, Guildford GU2 5XH, Surrey, England
[2] Siemens Med Solut USA Inc, Knoxville, TN USA
[3] Siemens Healthcare, Erlangen, Germany
[4] Tech Univ Munich, Klinikum Rechts Isar, Inst Radiol, D-8000 Munich, Germany
关键词
PET/MR; NEMA; performance;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The recently released Biograph mMR is the first commercially available integrated whole-body PET/MR scanner. There are considerable advantages to integrating both modalities in a single scanner that enables truly simultaneous acquisition. However, there are also concerns about the possible degradation of both PET and MR performance in an integrated system. This paper evaluates the performance of the Biograph mMR during independent and simultaneous acquisition of PET and morphologic MR data. Methods: The NEMA NU 2-2007 protocol was followed for studying the PET performance. The following measurements were performed: spatial resolution; scatter fraction, count losses, and randoms; sensitivity; accuracy of the correction for count losses and randoms; and image quality. The quality control manual of the American College of Radiology was followed for studying the MR performance. The following measurements were performed: geometric accuracy, spatial resolution, low-contrast detectability, signal-to-noise ratio, static field (B-0) homogeneity, radiofrequency field (B-1) homogeneity, and radiofrequency noise. Results: An average spatial resolution of 4.3 mm in full width at half maximum was measured at 1 cm offset from the center of the field of view. The system sensitivity was 15.0 kcps/MBq along the center of the scanner. The scatter fraction was 37.9%, and the peak noise-equivalent count rate was 184 kcps at 23.1 kBq/mL. The maximum absolute value of the relative count rate error due to dead-time losses and randoms was 5.5%. The average residual error in scatter and attenuation correction was 12.1%. All MR parameters were within the tolerances defined by the American College of Radiology. B0 inhomogeneities below 1 ppm were measured in a 120-mm radius. B1 homogeneity and signal-to-noise ratio were equivalent to those of a standard MR scanner. No radiofrequency interference was detected. Conclusion: These results compare favorably with other state-of-the-art PET/CT and PET/MR scanners, indicating that the integration of the PET detectors in the MR scanner and their operation within the magnetic field do not have a perceptible impact on the overall performance. The MR subsystem performs essentially like a standalone system. However, further work is necessary to evaluate the more advanced MR applications, such as functional imaging and spectroscopy.
引用
收藏
页码:1914 / 1922
页数:9
相关论文
共 50 条
  • [1] Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
    Delso, Gaspar
    Furst, Sebastian
    Jakoby, Bjorn
    Ladebeck, Ralf
    Ganter, Carl
    Nekolla, Stephan G.
    Schwaiger, Markus
    Ziegler, Sibylle, I
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61 : 238S - 245S
  • [3] PERFORMANCE-CHARACTERISTICS OF A WHOLE-BODY PET SCANNER
    DEGRADO, TR
    TURKINGTON, TG
    WILLIAMS, JJ
    STEARNS, CW
    HOFFMAN, JM
    COLEMAN, RE
    JOURNAL OF NUCLEAR MEDICINE, 1994, 35 (08) : 1398 - 1406
  • [4] Monte Carlo simulations of the count rate performance of a clinical whole-body MR/PET scanner
    Delso, Gaspar
    Martinez, Maria-Jose
    Torres, Irene
    Ladebeck, Ralf
    Michel, Christian
    Nekolla, Stephan
    Ziegler, Sibylle I.
    MEDICAL PHYSICS, 2009, 36 (09) : 4126 - 4135
  • [5] Clinical application of whole-body hybrid PET-MR scanner in oncology
    Ratib, O.
    Willi, J.
    Wissmeyer, M.
    Steiner, C.
    Allaoua, M.
    Garibotto, V.
    Rager, O.
    Zaidi, H.
    Zaidi, H.
    Becker, M.
    Vallee, J.
    Loubeyre, P.
    Becker, C.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S220 - S220
  • [6] Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scanner
    Delso, G.
    Martinez-Moeller, A.
    Bundschuh, R. A.
    Ladebeck, R.
    Candidus, Y.
    Faul, D.
    Ziegler, S. I.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (15): : 4361 - 4374
  • [7] Performance evaluation of a whole-body PET scanner using the NEMA protocol
    Brix, G
    Zaers, J
    Adam, LE
    Bellemann, ME
    Ostertag, H
    Trojan, H
    Haberkorn, U
    Doll, J
    Oberdorfer, F
    Lorenz, WJ
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (10) : 1614 - 1623
  • [8] Integrated Whole-Body PET/MR Hybrid Imaging Clinical Experience
    Quick, Harald H.
    von Gall, Carl
    Zeilinger, Martin
    Wiesmueller, Marco
    Braun, Harald
    Ziegler, Susanne
    Kuwert, Torsten
    Uder, Michael
    Doerfler, Arnd
    Kalender, Willi A.
    Lell, Michael
    INVESTIGATIVE RADIOLOGY, 2013, 48 (05) : 280 - 289
  • [9] Integrated simultaneous whole-body MR/PET: First comparison between MR/PET and PET/CT in patients
    Drzezga, Alexander
    Souvatzoglou, Michael
    Beer, Ambros
    Ziegler, Sibylle
    Fuerst, Sebastian
    Nekolla, Stephan
    Schwaiger, Markus
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [10] Performance evaluation of the whole-body PET scanner ECAT EXACT HR+
    Adam, LE
    Zaers, J
    Ostertag, H
    Trojan, H
    Bellemann, ME
    Brix, G
    Lorenz, WJ
    1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 1270 - 1274