Determination of serum uric acid using high-performance liquid chromatography (HPLC)/isotope dilution mass spectrometry (ID-MS) as a candidate reference method

被引:173
|
作者
Dai, Xinhua [1 ]
Fang, Xiang [1 ]
Zhang, Chunmei [1 ]
Xu, Ruifeng [1 ]
Xu, Bei [1 ]
机构
[1] Natl Inst Metrol, Natl Res Ctr CRMs, Beijing 100013, Peoples R China
关键词
uric acid; serum; LC/ID-MS; candidate reference method;
D O I
10.1016/j.jchromb.2007.07.035
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Uric acid is an important diagnostic marker of catabolism of the purine nucleosides, and accurate measurements of serum uric acid are necessary for proper diagnosis of gout or renal disease appearance. A candidate reference method involving isotope dilution coupled with liquid chromatography/mass spectrometry (LC/MS) has been described. An isotopically labeled internal standard, [1,3-N-15(2)] uric acid, was added to serum, followed by equilibration and protein removal clean up to prepare samples for liquid chromatography/mass spectrometry electrospray ionization (LC/MS-ESI) analyses. (M-H)(-) ions at m/z 167 and 169 for uric acid and its labeled internal standard were monitored for LC/MS. The accuracy of the measurement was evaluated by a comparison of results of this candidate reference method on lyophilized human serum reference materials for uric acid (Standard Reference Materials SRM909b) with the certified values determined by gas chromatography/mass spectrometry reference methods and by a recovery study for the added uric acid. The method performed well against the established reference method of ion-exchange followed by derivatization isotope dilution (ID) gas chromatography mass spectrometry (ID-GC/MS). The results of this method for uric acid agreed well with the certified values and were within 0.10%. The amounts of uric acid recovered and added were in good agreement for the three concentrations. This method was applied to determine uric acid in samples of frozen serum pools. Excellent precision was obtained with within-set CVs of 0.08-0.18% and between-set CVs of 0.02-0.07% for LC/MS analyses. Liquid chromatography/tandem mass spectrometry electrospray ionization (LC/MS/MS-ESI) analysis was also performed. The LC/MS and LC/MS/MS results were in very good agreement (within 0.14%). This LC/MS method, which demonstrates good accuracy and precision, and is in the speed of analysis without the need for a derivatization stage, qualifies as a candidate reference method. This method can be used as an alternative reference method to provide an accuracy base to which the routine methods can be compared. (c) 2007 Elsevier B.V. All fights reserved.
引用
收藏
页码:287 / 295
页数:9
相关论文
共 50 条
  • [1] Determination of serum cortisol using isotope dilution–liquid chromatography–mass spectrometry as a candidate reference method
    Pyong Gil Jung
    Byungjoo Kim
    Sang-Ryoul Park
    Hun-Young So
    Lian Hua Shi
    Yongseong Kim
    Analytical and Bioanalytical Chemistry, 2004, 380 : 782 - 788
  • [2] A CANDIDATE REFERENCE METHOD FOR THE DETERMINATION OF URIC-ACID IN SERUM BASED ON HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, COMPARED WITH AN ISOTOPE DILUTION GAS CHROMATOGRAPHY-MASS SPECTROMETER METHOD
    KOCK, R
    DELVOUX, B
    TILLMANNS, U
    GREILING, H
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1989, 27 (03): : 157 - 162
  • [3] ISOTOPE DILUTION MASS-SPECTROMETRY (ID-MS) AS A DEFINITIVE ANALYTICAL METHOD FOR GLUCOSE IN SERUM
    SUN, T
    COHEN, A
    HERTZ, H
    SCHAFFER, R
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1977, 173 (MAR20): : 165 - 165
  • [4] Determination of serum cortisol using isotope dilution-liquid chromatography-mass spectrometry as a candidate reference method
    Jung, PG
    Kim, B
    Park, SR
    So, HY
    Shi, LH
    Kim, Y
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 380 (5-6) : 782 - 788
  • [5] Determination of serum creatinine by liquid chromatography isotope dilution tandem mass spectrometry (LC-IDMS/MS): A candidate reference method
    Rasoul, Alikhani-Koupaei
    Mehrdad, Yazdanpanah
    Eleftherios, Diamandis
    Barry, Hoffman
    Khosrow, Adeli
    Jeffrey, Tong
    CLINICAL BIOCHEMISTRY, 2008, 41 (14-15) : 1273 - 1273
  • [6] Development and evaluation of a candidate reference method for the determination of total cortisol in human serum using isotope dilution liquid chromatography/mass spectrometry and liquid chromatography/tandem mass spectrometry
    Tai, SSC
    Welch, MJ
    ANALYTICAL CHEMISTRY, 2004, 76 (04) : 1008 - 1014
  • [7] Development of a candidate reference method for determination of tyrosine in serum by isotope dilution liquid chromatography-tandem mass spectrometry
    Lee, Hwashim
    Park, Sangryoul
    Kim, Byungjoo
    Lee, Gaeho
    ANALYTICAL SCIENCES, 2007, 23 (03) : 299 - 303
  • [8] Development of a Candidate Reference Method for Determination of Tyrosine in Serum by Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry
    Hwashim Lee
    Sangryoul Park
    Byungjoo Kim
    Gaeho Lee
    Analytical Sciences, 2007, 23 : 299 - 303
  • [9] Measurement of serum uric acid by isotope dilution liquid chromatography tandem mass spectrometry: Modification of a candidate reference measurement method and its clinical application
    Yan, Jun
    Lin, Haibiao
    Xu, Yuzhu
    Yu, Lintao
    Chen, Song
    Zhang, Qiaoxuan
    Han, Liqiao
    Ou, Yangfen
    Zhan, Min
    Wang, Jianbing
    Ke, Peifeng
    Wang, Yunxiu
    Huang, Xianzhang
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2022, 482
  • [10] Development of isotope dilution-liquid chromatography/tandem mass spectrometry as a candidate reference method for the determination of folic acid
    Jung, Minyoung
    Kim, Byungjoo
    Boo, Doo Wan
    So, Hun-Young
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2007, 28 (05): : 745 - 750