Robust analysis of space-, time-, and energy-resolved soft x-ray measurements of magnetically confined fusion plasmas (invited)

被引:2
|
作者
VanMeter, P. D. [1 ,3 ]
Delgado-Aparicio, L. F. [2 ]
Reusch, L. M. [1 ,4 ]
Den Hartog, D. J. [1 ]
机构
[1] Univ Wisconsin Madison, Dept Phys, Madison, WI 53706 USA
[2] Princeton Plasma Phys Lab PPPL, Princeton, NJ 08540 USA
[3] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[4] Edgewood Coll, Madison, WI 53711 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2021年 / 92卷 / 05期
关键词
REVERSED-FIELD-PINCH; TOMOGRAPHY; MODEL;
D O I
10.1063/5.0043787
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A novel compact multi-energy soft x-ray (ME-SXR) diagnostic based on the PILATUS3 100K x-ray detector has been developed in collaboration between the Princeton Plasma Physics Laboratory and the University of Wisconsin-Madison and tested on the Madison Symmetric Torus (MST) reversed-field pinch. This solid-state photon-counting detector consists of a two-dimensional array of similar to 100 000 pixels for which the lower photon absorption cutoff energy can be independently set, allowing it to be configured for a unique combination of simultaneous spatial, spectral, and temporal resolution of similar to 1 cm, 100 eV, and 500 Hz, respectively. The diagnostic is highly versatile and can be readily adapted to diverse plasma operating conditions and scientific needs without any required downtime. New results from improved-confinement and quasi-single helicity plasmas in the MST demonstrate how the detector can be applied to study multiple aspects of the evolution of magnetically confined fusion-grade plasmas. These include observing the evolution of thermal emissivity, characterizing the energy of mid-Z excitation lines, extracting the T-e profile, and observing the evolution of non-thermal populations. A technique for integrating the ME-SXR diagnostic into an integrated data analysis framework based on Bayesian inference is also presented. This allows ME-SXR measurements to be combined with data for complementary diagnostics in order to simultaneously infer T-e and n(Z) from all available information. Published under license by AIP Publishing.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] TIME-RESOLVED STUDIES OF SOFT X-RAY EMISSIONS AND SPATIAL UNIFORMITY OF LINEAR PLASMAS
    王晓方
    徐至展
    陈时胜
    范品忠
    张正泉
    钱爱娣
    余加进
    龚美霞
    高胜琛
    山冰
    Chinese Science Bulletin, 1992, (15) : 1254 - 1257
  • [42] Observation of femtosecond-laser-induced ablation plumes of aluminum using space- and time-resolved soft x-ray absorption spectroscopy
    Okano, Yasuaki
    Oguri, Katsuya
    Nishikawa, Tadashi
    Nakano, Hidetoshi
    APPLIED PHYSICS LETTERS, 2006, 89 (22)
  • [43] Watching the Methanol-to-Olefin Process with Time- and Space-Resolved High-Energy Operando X-ray Diffraction
    Wragg, David S.
    Brien, Matthew G. O.
    Bleken, Francesca L.
    Di Michiel, Marco
    Olsbye, Unni
    Fjellvag, Helmer
    ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2012, 51 (32) : 7956 - 7959
  • [44] Space- and time-resolved small angle X-ray scattering to probe assembly of silver nanocrystal superlattices
    Yu, Yixuan
    Yu, Dian
    Sadigh, Babak
    Orme, Christine A.
    NATURE COMMUNICATIONS, 2018, 9
  • [45] Space- and time-resolved small angle X-ray scattering to probe assembly of silver nanocrystal superlattices
    Yixuan Yu
    Dian Yu
    Babak Sadigh
    Christine A. Orme
    Nature Communications, 9
  • [46] Observation of population inversion in recombining Al plasma by time- and space-resolved x-ray spectroscopy
    Katsuragawa, M
    Itatani, J
    Orimo, S
    Ozaki, T
    Kuroda, H
    Sasaki, A
    APPLIED PHYSICS LETTERS, 1996, 68 (11) : 1479 - 1481
  • [47] A scattered X-ray correction method for transmission measurements and its verification using energy-resolved computed tomography
    Kuniwake, Hiroya
    Nishikawa, Junichiro
    Kanno, Ikuo
    Shimomura, Kohei
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2023, 60 (04) : 385 - 393
  • [48] High-speed tangentially viewing soft x-ray camera to study magnetohydrodynamic fluctuations in toroidally confined plasmas (invited)
    Ohdachi, S
    Toi, K
    Fuchs, G
    von Goeler, S
    Yamamoto, S
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (03): : 2136 - 2143
  • [49] DMX: An absolutely calibrated time-resolved broadband soft x-ray spectrometer designed for MJ class laser-produced plasmas (invited)
    Bourgade, JL
    Villette, B
    Bocher, JL
    Boutin, JY
    Chiche, S
    Dague, N
    Gontier, D
    Jadaud, JP
    Savale, B
    Wrobel, R
    Turner, RE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01): : 1173 - 1182
  • [50] RECENT PROGRESS IN SPACE-RESOLVED AND TIME-RESOLVED X-RAY SPECTROSCOPY OF LASER-PRODUCED PLASMAS
    YOUNG, BKF