Quantitative assessment of four-dimensional computed tomography image acquisition quality

被引:19
|
作者
Starkschall, George
Desai, Neil
Balter, Peter
Prado, Karl
Luo, Dershan
Cody, Dianna
Pan, Tinsu
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas, MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
来源
关键词
computed tomography; 4D imaging; image assessment; organ motion;
D O I
10.1120/jacmp.v8i3.2362
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of the present work was to describe the development and validation of a series of tests to assess the quality of four-dimensional ( 4D) computed-tomography ( CT) imaging as it is applied to radiation treatment planning. Using a commercial respiratory motion phantom and a programmable moving platform with a CT phantom, we acquired 4D CT datasets on two commercial multislice helical CT scanners that use different approaches to 4D CT image reconstruction. Datasets were obtained as the platform moved in various patterns designed to simulate breathing. Known inserts in the phantom were contoured, and statistics were generated to evaluate properties important to radiation therapy-namely, accuracy of phase-binning, shape, volume, and CT number. Phase-binning accuracy varied by as much as 5% for a 4D procedure in which images were reconstructed and then binned, but exhibited no variation for a 4D procedure in which projections were binned before reconstruction. The magnitude of geometric distortion was found to be small for both approaches, as was the magnitude of volume error. Partial-volume effects in the direction perpendicular to the transverse planes of reconstruction affected volume accuracy, however. Computed tomography numbers were reproduced accurately, but 4D images exhibited more variation in CT number than static CT images did. Characterization of such properties can be used to better understand and optimize the various parameters that affect 4D CT image quality.
引用
收藏
页码:1 / 20
页数:20
相关论文
共 50 条
  • [1] Assessment of four-dimensional CT image acquisition quality
    Starkschall, G.
    Desai, N.
    Balter, P.
    Prado, K.
    Luo, D.
    Cody, D.
    Pan, T.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 1986 - 1987
  • [2] Effect of respiratory cycle irregularities on image quality in four-dimensional computed tomography
    Thorndyke, B
    Xing, L
    Graves, E
    Loo, BW
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S506 - S506
  • [3] Quality assurance of four-dimensional CT image acquisition
    Fitzpatrick, MJ
    Starkschall, G
    Batter, P
    Pan, T
    Antolak, J
    Guerrero, T
    Stevens, C
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1780 - 1780
  • [4] Four-dimensional computed tomography: Image formation and clinical protocol
    Rietzel, E
    Pan, TS
    Chen, GTY
    [J]. MEDICAL PHYSICS, 2005, 32 (04) : 874 - 889
  • [5] Exploring correlation information for image compression of four-dimensional computed tomography
    Yan, Hui
    Li, Yexiong
    Dai, Jianrong
    [J]. QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (07) : 1270 - 1277
  • [6] A comprehensive quality assurance program for four-dimensional computed tomography in radiotherapy
    Tahiri, Jinane Bakkali
    Kyndt, Martin
    Dhont, Jennifer
    Gulyban, Akos
    Szkitsak, Juliane
    Bogaert, Evelien
    Reynaert, Nick
    Burghelea, Manuela
    [J]. PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 27
  • [7] Quality assurance of a breathing controlled four-dimensional computed tomography algorithm
    Szkitsak, Juliane
    Karius, Andre
    Hofmann, Christian
    Fietkau, Rainer
    Bert, Christoph
    Speer, Stefan
    [J]. PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 23 : 85 - 91
  • [8] Quality assurance of four-dimensional computed tomography for stereotactic ablative radiotherapy
    Nailon, W. H.
    Vaque, J. Puxeu
    McLaren, D. B.
    Erridge, S. C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S261 - S261
  • [9] Quantitation of the four-dimensional computed tomography process
    Lu, W
    Parikh, P
    El Naqa, I
    Nystrom, M
    Hubenschmidt, J
    Wahab, S
    Singh, A
    Christensen, G
    Mutic, S
    Bradley, J
    Low, D
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1780 - 1780
  • [10] Quantitation of the four-dimensional computed tomography process
    Lu, W
    Parikh, P
    ElNaqa, I
    Nystrom, M
    Hubenschmidt, J
    Wahab, S
    Mutic, S
    Singh, A
    Christensen, G
    Bradley, JD
    Low, DA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S286 - S286