Radiobiological Comparison of Acuros External Beam and Anisotropic Analytical Algorithm on Esophageal Carcinoma Radiotherapy Treatment Plans

被引:0
|
作者
Wang, Lin [1 ,2 ,3 ,4 ]
Zhang, Jianping [1 ,2 ,3 ,4 ,5 ]
Huang, Miaoyun [1 ,3 ,4 ]
Xu, Benhua [1 ,2 ,3 ,4 ,5 ]
Li, Xiaobo [1 ,2 ,3 ,4 ,5 ]
机构
[1] Fujian Med Univ, Dept Radiat Oncol, Union Hosp, Fuzhou, Peoples R China
[2] Fujian Med Univ, Coll Med Technol & Engn, Dept Med Imaging Technol, Fuzhou, Peoples R China
[3] Fujian Med Univ, Fujian Key Lab Intelligent Imaging & Precis Radio, Fuzhou, Peoples R China
[4] Clin Res Ctr Radiol & Radiotherapy Fujian Prov Di, Fuzhou, Peoples R China
[5] Fujian Med Univ, Union Clin Med Coll, Fuzhou, Peoples R China
来源
DOSE-RESPONSE | 2022年 / 20卷 / 03期
关键词
radiobiological assessment; anisotropic analytical algorithm and acuros external beam; gamma passing rate; composition of the structures; esophageal carcinoma; DOSE-TO-MEDIUM; MONTE-CARLO; DOSIMETRIC VALIDATION; REPORTING MODES; XB ALGORITHM; WATER; AAA; DIFFERENCE; IMPACT; HEAD;
D O I
10.1177/15593258221105678
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Objective: The present study aimed to investigate the dose differences and radiobiological assessment between Anisotropic Analytical Algorithm (AAA) and Acuros External Beam (AXB) with its 2 calculation models, namely, dose-to-water (AXB-Dw) and dose-to-medium (AXB-Dm), on esophageal carcinoma radiotherapy treatment plans. Materials and methods: The AXB-Dw and AXB-Dm plans were generated by recalculating the initial 66 AAA plans using the AXB algorithm with the same monitor units and beam parameters as those in the original plan. The dosimetric and radiobiological assessment parameters were calculated for the planning target volume (PTV) and organs at risk (OARs). The gamma agreement for the PTV and the correlation between it and the volume of the air cavity and bone among the different algorithms were compared simultaneously. The dose discrepancy between the theoretical calculation and treatment planning system (TPS) when switching from AXB-Dm to AXB-Dw was analyzed according to the composition of the structures. Results: The PTV dose of AXB-Dm plans was significantly smaller than that of the AAA and AXB-Dw plans (P < .05), except for D-2. The difference values for AAA vs AXB-Dm ( increment D-x,D-(AAA-AXB,D-Dm)) and AXB-Dw vs AXB-Dm ( increment D-x,D-(AXB,D-Dw-AXB,D-Dm)) were 1.94% [1.27%, 2.64%] and 1.95% [1.56%, 2.27%], respectively. For the spinal cord and heart, there were obvious differences between the AAA vs AXB-Dm (spinal cord: 1.15%, heart: 2.89%) and AXB-Dw vs AXB-Dm (spinal cord: 1.88%, heart: 3.25%) plans. For the lung, the differences between AAA vs AXB-Dm and AAA vs AXB-Dw were significantly larger than those of AXB-Dm vs AXB-Dw. Compared to the case of AAA and AXB-Dw, the decrease in biologically effective dose (BED10, alpha beta=10 ) of AXB-Dm due to dose non-uniformity exceeded 6.5%, even for a small sigma. The average values of equivalent uniform dose in the AAA, AXB-Dw, and AXB-Dm plans were 52.03 +/-.39 Gy, 52.24 +/- .81 Gy, and 51.13 +/- .47 Gy, respectively. The tumor control probability (TCP) results for PTV in the AAA, AXB-Dw, and AXB-Dm plans were 62.29 +/- 1.57%, 62.82 +/- 1.69%, and 58.68 +/- 1.88%, respectively. With the 2%/2 mm and 3%/3 mm acceptance criteria, the mean values of Delta gamma(AXB-Dw)(AAA), Delta gamma(AXB-Dm)(AAA), and Delta gamma(AXB-Dw)(AXB-Dm) were 87.24, 63.3, and 64.81% vs 97.86, 91.77, and 89.25%, respectively. The dose discrepancy between the theoretical calculation and TPS when switching from AXB-Dm to AXB-Dw was approximately 1.63%. Conclusions: The AAA and AXB-Dw algorithms overestimated the radiobiological parameters when the tumor particularly consisted of nonuniform tissues. A relatively small dose difference could cause a significant reduction in the corresponding TCP. Dose distribution algorithms should be carefully chosen by physicists and oncologists to improve tumor control, as well as to optimize OARs protection.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] Analysis of treatment plans robustness for dynamic techniques in external beam radiotherapy
    Dabrowska-Szewczyk, E.
    Bukat, M.
    Kukolowicz, P.
    Szalkowski, D.
    Paciorkiewicz, A.
    Zawadzka, A.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S989 - S989
  • [22] Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning
    Bragg, Christopher M.
    Conway, John
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 (03) : 315 - 323
  • [23] Comparison of Acuros (AXB) and Anisotropic Analytical Algorithm (AAA) for dose calculation in treatment of oesophageal cancer: effects on modelling tumour control probability
    Padmanaban, Sriram
    Warren, Samantha
    Walsh, Anthony
    Partridge, Mike
    Hawkins, Maria A.
    RADIATION ONCOLOGY, 2014, 9
  • [24] Combination external beam radiotherapy and brachytherapy for radical treatment of esophageal carcinoma in patients not suitable for surgery or chemoradiation
    Aggarwal, Ajay
    Harrison, Mark
    Sinharay, Ricky
    Cooper, Derek
    Glynne-Jones, Robert
    Hoskin, Peter
    JOURNAL OF CLINICAL ONCOLOGY, 2013, 31 (04)
  • [25] Dosimetric comparison of analytical anisotropic algorithm and the two dose reporting modes of Acuros XB dose calculation algorithm in volumetric modulated arc therapy of carcinoma lung and carcinoma prostate
    Gopalakrishnan, Zhenia
    Bhasi, Saju
    Raghukumar, P.
    V. Menon, Sharika
    Sarin, B.
    Thayil, Anna George
    Nair, Raghuram K.
    MEDICAL DOSIMETRY, 2022, 47 (03) : 280 - 287
  • [26] RADIOBIOLOGICAL COMPARISON OF EXTERNAL BEAM IRRADIATION AND RADIOIMMUNOTHERAPY IN RENAL-CELL CARCINOMA XENOGRAFTS
    WESSELS, BW
    VESSELLA, RL
    PALME, DF
    BERKOPEC, JM
    SMITH, GK
    BRADLEY, EW
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (06): : 1257 - 1263
  • [27] Comparison of Acuros (AXB) and Anisotropic Analytical Algorithm (AAA) for dose calculation in treatment of oesophageal cancer: effects on modelling tumour control probability
    Sriram Padmanaban
    Samantha Warren
    Anthony Walsh
    Mike Partridge
    Maria A Hawkins
    Radiation Oncology, 9
  • [28] Radiobiological model comparison of 3D conformal radiotherapy and IMRT plans for the treatment of prostate cancer
    Deb, P.
    Fielding, A.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (02) : 51 - 61
  • [29] Radiobiological model comparison of 3D conformal radiotherapy and IMRT plans for the treatment of prostate cancer
    P. Deb
    A. Fielding
    Australasian Physics & Engineering Sciences in Medicine, 2009, 32 : 51 - 61
  • [30] Breast Cancer Radiation Treatment Planning: A Comparison of Anisotropic Analytical Algorithm (AAA) and Pencil Beam Convolution (PBC) Algorithm
    Yoo, S.
    Wu, Q.
    O'Daniel, J.
    Horton, J.
    Yin, F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S240 - S240