Inference and sample size calculation in the fit assessment of filtering facepiece respirators

被引:3
|
作者
Zhang, Zhiwei [1 ]
Kotz, Richard M. [1 ]
机构
[1] US FDA, Div Biostat, OSB, Ctr Devices & Radiol Hlth, Rockville, MD 20850 USA
关键词
analysis of variance; delta method; random effect; respirator; sample size; variance component;
D O I
10.1080/10543400802071394
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Filtering facepiece respirators have recently been cleared by the U.S. Food and Drug Administration (FDA) for use by the general public in public health medical emergencies such as pandemic influenza. In the fit assessment of these devices it is important to distinguish between the two sources of variability: population heterogeneity and random fluctuations over repeated donnings. The FDA Special Controls Guidance Document (SCGD) which describes these devices and their evaluation, recommends that the fit performance of a filtering facepiece respirator be evaluated in terms of the proportion of users who will receive a specified level of protection 95% of the time. A point estimator of this proportion is easily obtained under an analysis of variance model, and the SCGD suggests bootstrap as one possible approach to interval estimation. This paper describes a closed-form procedure to obtain confidence intervals and provides sample size formulas. Simulation results suggest that the proposed procedure performs well in realistic settings and compares favorably to two simple bootstrap procedures.
引用
收藏
页码:713 / 723
页数:11
相关论文
共 50 条
  • [41] Assessing the Fit of N95 Filtering Facepiece Respirators Fitted with an Ear Loop Strap System: A Pilot Study
    Niu, Xinyi
    Koehler, Richard H.
    Yermakov, Michael
    Grinshpun, Sergey A.
    ANNALS OF WORK EXPOSURES AND HEALTH, 2023, 67 (01) : 50 - 58
  • [42] Physiological effects of filtering facepiece respirators based on age and exercise intensity
    Go, Sulbee
    Yang, Yeram
    Park, Suhong
    Moon, Hyo Youl
    Yoon, Chungsik
    PLOS ONE, 2024, 19 (08):
  • [43] The influence of head strap elasticity on the protective properties of filtering facepiece respirators
    Tretiakova, Larisa
    Cheberyachko, Yurii
    Sharovatova, Olena
    Nehrii, Tetiana
    Nehrii, Serhii
    Kravchenko, Bohdan
    Zolotarova, Oksana
    RUDARSKO-GEOLOSKO-NAFTNI ZBORNIK, 2024, 39 (01): : 131 - 140
  • [44] Nanoparticle Filtration Performance of Filtering Facepiece Respirators and Canister/cartridge Filters
    Rengasamy, Samy
    BerryAnn, Roland
    Szalajda, Jonathan
    JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2013, 10 (09) : 519 - 525
  • [45] Multicycle Autoclave Decontamination of N95 Filtering Facepiece Respirators
    Bopp, Nathen E.
    Bouyer, Donald H.
    Gibbs, Christopher M.
    Nichols, Joan E.
    Ntiforo, Corrie A.
    Grimaldo, Miguel A.
    APPLIED BIOSAFETY, 2020, 25 (03) : 150 - 156
  • [46] Recovery of Bacteria in Filtering Facepiece Respirators and Effects of Artificial Saliva/Perspiration on Bacterial Survival and Performance of Respirators
    Lin, Tzu-Hsien
    Tang, Feng-Cheng
    Chiang, Chia-Hsiu
    Chang, Cheng-Ping
    Lai, Chane-Yu
    AEROSOL AND AIR QUALITY RESEARCH, 2017, 17 (01) : 187 - 197
  • [47] Comparing mask fit and usability of traditional and nanofibre N95 filtering facepiece respirators before and after nursing procedures
    Suen, L. K. P.
    Guo, Y. P.
    Ho, S. S. K.
    Au-Yeung, C. H.
    Lam, S. C.
    JOURNAL OF HOSPITAL INFECTION, 2020, 104 (03) : 336 - 343
  • [48] Temporal changes in filtering-facepiece respirator fit
    Zhuang, Ziqing
    Bergman, Michael
    Brochu, Elizabeth
    Palmiero, Andrew
    Niezgoda, George
    He, Xinjian
    Roberge, Raymond
    Shaffer, Ronald
    JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2016, 13 (04) : 269 - 278
  • [49] Effect of Fit Testing on the Protection Offered by N95 Filtering Facepiece Respirators Against Fine Particles in a Laboratory Setting
    Reponen, Tiina
    Lee, Shu-An
    Grinshpun, Sergey A.
    Johnson, Erik
    Mckay, Roy
    ANNALS OF OCCUPATIONAL HYGIENE, 2011, 55 (03): : 264 - 271
  • [50] Implementing fit testing for N95 filtering facepiece respirators: Practical information from a large cohort of hospital workers
    McMahon, Elizabeth
    Wada, Koji
    Dufresne, Andre
    AMERICAN JOURNAL OF INFECTION CONTROL, 2008, 36 (04) : 298 - 300