Peripheral and surface dose assessment using diode and Gafchromic EBT3 films dosimeters for different radiotherapy techniques

被引:3
|
作者
AL-Shareef, Jamal M. [1 ,2 ,3 ]
Attalla, Ehab M. [2 ]
El-Gebaly, Reem H. [1 ]
Deiab, Nashaat A. [2 ]
Mokhtar, Maha H. [2 ]
Fathy, Mohamed M. [1 ]
机构
[1] Cairo Univ, Fac Sci, Dept Biophys, Giza, Egypt
[2] Cairo Univ, NCI, Dept Radiotherapy & Nucl Med, Giza, Egypt
[3] Sanaa Univ, Dept Phys, Sanaa, Yemen
关键词
Peripheral dose; Surface dose; Diode; EBT3; films; Second cancer; INTENSITY-MODULATED RADIOTHERAPY; HIGH-ENERGY PHOTON; RADIATION-THERAPY; MONTE-CARLO; HEAD; OPTIMIZATION; VERIFICATION; ALGORITHMS; ACCURACY; ENTRANCE;
D O I
10.1016/j.radphyschem.2022.110237
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Purpose: Treatment planning systems (TPSs) still face challenges in determining the peripheral and surface doses specially when using thermoplastic mask and bolus. This study aims to assess the peripheral and surface dose using diode and Gafchromic EBT3 films dosimeters. In addition to investigate the impact of thermoplastic mask with different thickness of bolus on these doses.Methods: The measurements were performed for an Alderson Rando phantom. The maxilla and prostate were delineated as the target, because the maxilla is close to the surface, to determine the impact of thermoplastic mask and bolus on the skin dose.Results: The measured surface dose for IMRT was higher compared to TPS with an average difference of 21.5%, 20.2%, 18.7%, and 13.4% in the absence of thermoplastic mask and bolus, with mask, with bolus 5 mm and with bolus 10 mm respectively. For 3D-CRT, the measured surface dose was also higher compared to TPS with an average difference of 20.7%, 19.6%, 16.1%, and 13.3% in the absence of thermoplastic mask and bolus, with mask, with bolus 5 mm and with bolus 10 mm respectively. For the prostate, the PD that is close to the target was in good agreement between peripheral dose measured with diode and acquired from TPS. Whereas TPS underestimated the PD for both 3D-CRT and IMRT at locations that are away from the target. Moreover, when comparing 3D-CRT to IMRT for PD, 3D-CRT resulted in higher for distances close to the target, whereas the IMRT indeed delivered higher than 3D-CRT to locations that are away from the target.Conclusions: This study showed that diode dosimeter is recommended to estimate both peripheral and surface doses whereas EBT3 films dosimeter is recommended to estimate surface dose and not for PD.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] A novel method for EBT3 Gafchromic films read-out at high dose levels
    Vaiano, Patrizio
    Consales, Marco
    Casolaro, Pierluigi
    Campajola, Luigi
    Fienga, Francesco
    Di Capua, Francesco
    Breglio, Giovanni
    Buontempo, Salvatore
    Cutolo, Antonello
    Cusano, Andrea
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 61 : 77 - 84
  • [22] Verification of high dose rate brachytherapy dose distributions with EBT3 Gafchromic film quality control techniques
    Palmer, Antony L.
    Nisbet, Andrew
    Bradley, David
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (03): : 497 - 511
  • [23] Absolute Film Dosimetry for Stereotactic Radiosurgery and Stereotactic Body Radiotherapy Quality Assurance Using Gafchromic EBT3 Films
    Wen, N.
    Lu, S.
    Qin, Y.
    Huang, Y.
    Zhao, B.
    Liu, C.
    Chetty, I.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3593 - 3593
  • [24] GafChromic® EBT3 films for patient specific IMRT QA using amultichannel approach
    Marrazzo, Livia
    Zani, Margherita
    Pallotta, Stefania
    Arilli, Chiara
    Casati, Marta
    Compagnucci, Antonella
    Talamonti, Cinzia
    Bucciolini, Marta
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (08): : 1035 - 1042
  • [25] Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
    Ning Wen
    Siming Lu
    Jinkoo Kim
    Yujiao Qin
    Yimei Huang
    Bo Zhao
    Chang Liu
    Indrin J. Chetty
    [J]. Radiation Oncology, 11
  • [26] Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
    Wen, Ning
    Lu, Siming
    Kim, Jinkoo
    Qin, Yujiao
    Huang, Yimei
    Zhao, Bo
    Liu, Chang
    Chetty, Indrin J.
    [J]. RADIATION ONCOLOGY, 2016, 11
  • [27] Absolute dosimetry with EBT3 Gafchromic films in a pulsed electron beam at high dose-rate
    Jaccard, M.
    Petersson, K.
    Buchillier, T.
    Bailat, C.
    Germond, J. F.
    Moeckli, R.
    Bourhis, J.
    Vozenin, M. C.
    Bochud, F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S690 - S690
  • [28] On the use of Gafchromic EBT3 films for validating a commercial electron Monte Carlo dose calculation algorithm
    Chan, EuJin
    Lydon, Jenny
    Kron, Tomas
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (05): : 2091 - 2102
  • [29] Evaluation of air cavities on dose distributions with air-filled apparatuses having different volumes using Gafchromic EBT3 films in brachytherapy
    Celik, Mehmet Bahadir
    Olacak, Nezahat
    Us, Songul Barlaz
    Tavlayan, Emin
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (04) : 411 - 416
  • [30] Dose Verification of Spot-Scanning Proton Beam Using GafChromic EBT3 Film
    Chen, C.
    Tang, S.
    Mah, D.
    Chan, M.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3399 - 3399