Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films

被引:71
|
作者
Wen, Ning [1 ]
Lu, Siming [1 ]
Kim, Jinkoo [1 ]
Qin, Yujiao [1 ]
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Liu, Chang [1 ]
Chetty, Indrin J. [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 West Grand Blvd, Detroit, MI 48202 USA
来源
RADIATION ONCOLOGY | 2016年 / 11卷
关键词
Stereotactic radiosurgery; Stereotactic body radiation therapy; Gafchromic films; Quality assurance; IONIZATION-CHAMBER ARRAYS; RADIOCHROMIC FILM; PHOTON BEAMS; SCANNERS; VERIFICATION;
D O I
10.1186/s13014-016-0709-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic (TM) (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). Method and materials: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference ion chamber array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to 4 days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 83 SRS/SBRT treatment plans, and analysis of the measurements and calculations are presented here. Results: The scanner response varied within 1 % for the field sizes less than 5 x 5 cm(2), and up to 5 % for the field sizes of 10 x 10 cm(2) for all color channels. The scanner correction method was able to remove visually evident, irregular detector responses for larger field sizes. The dose response of the film changed rapidly (similar to 10 %) in the first two hours and became smooth plateaued afterwards, similar to 3 % change between 2 and 24 h. The uncertainties were approximately 1.5, 1.7 and 4.8 % over the dose range of 3 similar to 15 Gy for the red, green and blue channels. The green channel showed very high sensitivity and low uncertainty in the dose range between 10 and 15 Gy, which is suitable for SRS/SBRT commissioning and PSQA. The difference between the calculated dose and measured dose of ion chamber measurement at isocenter was -0.64 +/- 2.02 for all plans, corresponding to a 95 % confidence interval of (-1.09, -0.26). The percentage of points passing the 3 %/1 mm gamma criteria in absolute dose, averaged over all tests was 95.0 +/- 4.2. Conclusion: We have developed the EBT3 films based dosimetry protocol to obtain absolute dose values. The overall uncertainty has been established to be 1.5 % for SRS and SBRT PSQA.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
    Ning Wen
    Siming Lu
    Jinkoo Kim
    Yujiao Qin
    Yimei Huang
    Bo Zhao
    Chang Liu
    Indrin J. Chetty
    Radiation Oncology, 11
  • [2] Absolute Film Dosimetry for Stereotactic Radiosurgery and Stereotactic Body Radiotherapy Quality Assurance Using Gafchromic EBT3 Films
    Wen, N.
    Lu, S.
    Qin, Y.
    Huang, Y.
    Zhao, B.
    Liu, C.
    Chetty, I.
    MEDICAL PHYSICS, 2016, 43 (06) : 3593 - 3593
  • [3] Dosimetric verification of stereotactic radiosurgery/stereotactic radiotherapy dose distributions using Gafchromic EBT3
    Cusumano, Davide
    Fumagalli, Maria L.
    Marchetti, Marcello
    Fariselli, Laura
    De Martin, Elena
    MEDICAL DOSIMETRY, 2015, 40 (03) : 226 - 231
  • [4] Gafchromic EBT3 film provides equivalent dosimetric performance to EBT-XD film for stereotactic radiosurgery dosimetry
    Smyth, Lloyd
    Alves, Andrew
    Collins, Katherine
    Beveridge, Sabeena
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (03) : 1095 - 1106
  • [5] Evaluation of Dual-Channel Compound Method On EBT3 Film Dosimetry for Stereotactic Body Radiotherapy Quality Assurance
    Chung, J.
    Kang, S.
    Suh, T.
    Cho, W.
    Lee, J.
    MEDICAL PHYSICS, 2017, 44 (06) : 2880 - 2881
  • [6] Precise EBT3 Gafchromic film dosimetry for Grid therapy
    Gholami, Somayeh
    Nedaie, Hassan Ali
    Meigooni, Ali S.
    RADIATION MEASUREMENTS, 2019, 121 : 69 - 76
  • [7] Monte Carlo calculated output correction factors for Gafchromic EBT3 film for relative dosimetry in small stereotactic radiosurgery fields
    Johnny E. Morales
    Martin Butson
    Robin Hill
    Scott B. Crowe
    J. V. Trapp
    Physical and Engineering Sciences in Medicine, 2020, 43 : 609 - 616
  • [8] Monte Carlo calculated output correction factors for Gafchromic EBT3 film for relative dosimetry in small stereotactic radiosurgery fields
    Morales, Johnny E.
    Butson, Martin
    Hill, Robin
    Crowe, Scott B.
    Trapp, J., V
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) : 609 - 616
  • [9] Multichannel absolute film dosimetry for EBT3 GafChromic films
    Ventura, T.
    Lopes, M. C.
    Ferreira, B. C.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S169 - S169
  • [10] Clinical Comparison of Gafchromic EBT-XD Film to Gafchromic EBT3 Film When Utilized as An Absolute Dose Verification Method in Stereotactic Radiotherapy
    Goodman, R.
    Murphy, T.
    Siergiej, D.
    Schroeder, T.
    MEDICAL PHYSICS, 2018, 45 (06) : E280 - E280