A new dead-time determination method for gamma-ray detectors using attenuation law

被引:2
|
作者
Akyurek, T. [1 ]
机构
[1] Marmara Univ, Fac Art & Sci, Dept Phys, TR-34722 Istanbul, Turkey
关键词
dead-time; non-paralyzing model; paralyzing model; DMRAL method;
D O I
10.1016/j.net.2021.06.038
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
This study presents a new dead-time measurement method using the gamma attenuation law and generalized dead-time models for nuclear gamma-ray detectors. The dead-time of the Nal(TI) detection system was obtained to validate the new dead-time determination method using very thin lead and polyethylene absorbers. Non-paralyzing dead-time was found to be 8.39 mu s, and paralyzing dead-time was found to be 8.35 mu s using lead absorber for Nal(TI) scintillator detection system. These dead-time values are consistent with the previously reported dead-time values for scintillator detection systems. The gamma build-up factor's contribution to the dead-time was neglected because a very thin material was used. (C) 2021 Korean Nuclear Society, Published by Elsevier Korea LLC.
引用
收藏
页码:4093 / 4097
页数:5
相关论文
共 50 条
  • [41] DETERMINATION OF HYDROUS PROFILE OF A SOIL COLUMN BY ATTENUATION OF A GAMMA-RAY BEAM
    RAHMOUNI, O
    MESURES REGULATION AUTOMATISME, 1971, 36 (8-9): : 102 - &
  • [42] Determination of Gamma-ray Attenuation Coefficients at Different Energies in Amasya Marbles
    Mavi, B.
    Oner, F.
    Akkurt, I.
    ACTA PHYSICA POLONICA A, 2015, 128 (2B) : B395 - B396
  • [43] Study on the Correction Method for the Dead-Time Effects of Neutron Detectors Due to Pulsed Radiation
    Guo, Siming
    Wu, Jinjie
    Ma, Zhongjian
    Li, Guanjia
    Wang, Qingbin
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (04) : 485 - 491
  • [44] Study on the Correction Method for the Dead-Time Effects of Neutron Detectors Due to Pulsed Radiation
    Siming Guo
    Jinjie Wu
    Zhongjian Ma
    Guanjia Li
    Qingbin Wang
    Journal of the Korean Physical Society, 2018, 72 : 485 - 491
  • [45] A Novel Method for Dead-Time Compensation of the Inverter Using SVPWM
    Gong, YiMin
    Chen, XiaoJiao
    Huang, Ying
    ADVANCES IN MULTIMEDIA, SOFTWARE ENGINEERING AND COMPUTING, VOL 2, 2011, 129 : 563 - 568
  • [46] GAMMA-RAY SPECTROSCOPY USING GE(LI) DETECTORS - NUCL
    CAMP, DC
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1970, : 1 - &
  • [47] Determination of self-attenuation correction factor for lichen samples by using gamma-ray spectrometry
    Eke, C.
    Agar, O.
    Boztosun, I.
    Aslan, A.
    Emsen, B.
    KERNTECHNIK, 2017, 82 (01) : 136 - 139
  • [48] ACCURATE DETERMINATION OF PLUTONIUM K-ABSORPTION EDGE ENERGY USING GAMMA-RAY ATTENUATION
    CANADA, TR
    BEARSE, RC
    TAPE, JW
    NUCLEAR INSTRUMENTS & METHODS, 1977, 142 (03): : 609 - 611
  • [49] INSITU GAMMA-RAY MEASUREMENT USING GE(LI)DETECTORS
    SAKAI, E
    TERADA, H
    KATAGIRI, M
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (01) : 726 - 733
  • [50] POROSITY MEASUREMENTS OF CORE SAMPLES USING GAMMA-RAY ATTENUATION
    BODWADKAR, SV
    REIS, JC
    NUCLEAR GEOPHYSICS, 1994, 8 (01): : 61 - 78