Accuracy of Implant Placement Using a CAD/CAM Surgical Guide: An In Vitro Study

被引:0
|
作者
Nokar, Saeid [1 ]
Moslehifard, Elnaz [1 ]
Bahman, Tootiaii [2 ]
Bayanzadeh, Masoud [1 ]
Nasirpouri, Farzad [3 ]
Nokar, Arash [4 ]
机构
[1] Tabriz Univ Med Sci, Sch Dent, Dept Prosthodont, Tabriz, Iran
[2] Malek Ashtar Univ, Tehran, Iran
[3] Sahand Univ Technol, Dept Mat Engn, Tabriz, Iran
[4] SSO IRI, Tehran, Iran
关键词
computed tomography; computer-aided design; coordinate measuring machine; dental implants; stereolithography; surgical templates; DEFINITIVE PROSTHESIS; DENTAL IMPLANTS; TEMPLATE;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To determine and compare the accuracy of an advanced surgical template based on computer-aided design/computer-assisted manufacture (CAD/CAM) with the conventional surgical template in different respects such as entry point, length, and osteotomy angle. Materials and Methods: Computed tomography (CT) scanning of a dentate epoxy mandible was performed and its three-dimensional computerized model was simulated. Sixteen rapid-prototyped models were fabricated and divided into two groups. In the first group, a radiographic template was fabricated and placed on the model during CT scanning and then was modified to the conventional surgical template form. In the second group, a coordinate measuring machine was used to reformat a nonanatomic radiographic template fabricated by a stereolithographic machine, and four implants were planned and then placed in the jaw. The differences between planned and actual mesiodistal and buccolingual entry points, lengths, and angles of the implants were measured. Statistical analysis was performed with the Mann-Whitney and Friedman tests to detect differences between groups. Results: The average differences between the planned and actual entry points in the mesiodistal and buccolingual directions, lengths, and angles of the implants and the osteotomy showed a considerable reduction in the CAD/CAM group versus the conventional group (P < .005). Conclusion: The accuracy of implant placement was improved using an innovative CAD/CAM surgical template. INT J ORAL MAXILLOFAC IMPLANTS 2011;26:520-526
引用
收藏
页码:520 / 526
页数:7
相关论文
共 50 条
  • [1] Reliability of a CAD/CAM Surgical Guide for Implant Placement: An In Vitro Comparison of Surgeons' Experience Levels and Implant Sites
    Park, Su-Jung
    Leesungbok, Richard
    Cui, Taixing
    Lee, Suk Won
    Ahn, Su-Jin
    [J]. INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2017, 30 (04) : 367 - 369
  • [2] Accuracy of implant placement using CAD-CAM tooth-supported surgical guides for an auricular prosthesis in vitro
    Fan, Yiping
    Wei, Hongbo
    Zhao, Ruifeng
    Lin, Xiuwen
    Bai, Shizhu
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2024, 33 (01): : 70 - 76
  • [3] ACCURACY OF THREE DIFFERENT TYPES OF STEREOLITHOGRAPHIC SURGICAL GUIDE IN IMPLANT PLACEMENT: AN IN VITRO STUDY
    Turbush, Sarah Katherine
    Turkyilmaz, Ilser
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2012, 108 (03): : 181 - 188
  • [4] Accuracy assessment of implant placement with versus without a CAD/CAM surgical guide by novices versus specialists via the digital registration method: an in vitro randomized crossover study
    Li, Sha
    Yi, Chun
    Yu, Ziyang
    Wu, Aozhou
    Zhang, Yu
    Lin, Ye
    [J]. BMC ORAL HEALTH, 2023, 23 (01)
  • [5] Accuracy assessment of implant placement with versus without a CAD/CAM surgical guide by novices versus specialists via the digital registration method: an in vitro randomized crossover study
    Sha Li
    Chun Yi
    Ziyang Yu
    Aozhou Wu
    Yu Zhang
    Ye Lin
    [J]. BMC Oral Health, 23
  • [6] Accuracy of implant placement with a stereolithographic surgical guide
    Sarment, DP
    Sukovic, P
    Clinthorne, N
    [J]. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2003, 18 (04) : 571 - 577
  • [7] Reliability of a chairside CAD-CAM surgical guide for dental implant surgery on the anterior maxilla: An in vitro study
    Htay, Phyo Ei Ei
    Leesungbok, Richard
    Lee, Suk Won
    Jee, Yu-Jin
    Kang, Kyung Lhi
    Hong, Sung Ok
    [J]. JOURNAL OF ADVANCED PROSTHODONTICS, 2023, 15 (05): : 259 - 270
  • [8] Accuracy of a milled digital implant surgical guide: An in vitro study
    Liu, Xiaoqian
    Liu, Jianzhang
    Feng, Hailan
    Pan, Shaoxia
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2022, 127 (03): : 453 - 461
  • [9] Accuracy Assessment of Immediate and Delayed Implant Placements Using CAD/CAM Surgical Guides
    Alzoubi, Fawaz
    Massoomi, Nima
    Nattestad, Anders
    [J]. JOURNAL OF ORAL IMPLANTOLOGY, 2016, 42 (05) : 391 - 398
  • [10] Accuracy of Implant Placement Guided with Surgical Template: An In Vitro and In Vivo Study
    Zhou, Miao
    Zhou, Hui
    Li, Shu-Yi
    Yang, Xiao-Bin
    Geng, Yuan-Ming
    Che, Yue-Juan
    [J]. INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2021, 41 (02) : E55 - E62