摘要:
Selectivity is very important for stationary phase applications. However, selectivity is usually quite difficult to adjustonce the ligands are fixed. Departing from the reported methods, dodecyl methacrylate (DOMA) and hydroxyethyl methacrylate (HEMA) were grafted onto a silica surface via two-step surface initiated-atom transfer radical polymerization (SI-ATRP) to synthesize reverse-phase/hydrophilic interaction mixed-mode stationary phases. The grafted amounts of C12 and OH functional groups were controlled by varying the ratios of DOMA to HEMA in the polymerization system. The resulting stationary phases were characterized by transmission electron microscopy (TEM), solid-state C-13 NMR spectroscopy, elemental analysis (EA), Fourier transform infrared (FT-IR) spectrometry, and thermogravimetric analysis (TGA). The chromatographic performance and separation selectivity of the packed columns were investigated in different chromatographic modes using a wide range of analytes, including a non-polar benzene series, moderately polar -agonists as well as organic acids, and strongly polar nucleosides. Exceptionally, the retention of the tested analytes displayed an obvious dependence on the ratio of the two functional monomers in the polymerization. In conclusion, the proposed strategy for the development of a mixed-mode chromatographic stationary phase can provide flexible selectivity by simply tuning the ratios of functional monomers.
引用
收藏
页码:5387 / 5397
页数:11
相关论文