Liquid chromatography-electron ionization mass spectrometry: Fields of application and evaluation of the performance of a Direct-EI interface

被引:23
|
作者
Cappiello, A [1 ]
Famiglini, G [1 ]
Palma, P [1 ]
Siviero, A [1 ]
机构
[1] Univ Urbino, Ist Sci Chim F Bruner, I-61029 Urbino, Italy
关键词
nano-column liquid chromatography; electron ionization mass spectrometry; Direct-EI; trace analysis;
D O I
10.1002/mas.20054
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A comprehensive evaluation and a thorough discussion of the fields of possible applications of the Direct-EI interface are described in this review. Direct-EI allows the direct introduction of the effluent from a capillary HPLC column into the electron ionization (EI) ion source of a mass spectrometer Thanks to the reduced liquid intake and an in-source nebulizer the interfacing process occurs smoothly and entirely into the ion source. No intermediate interfacing mechanisms of any sort are interposed between the column and the mass spectrometer, thus circumventing any undesired sample loss and minimizing the effort for instrument modification. Theoretically, any GC-MS system can be converted into an LC-MS for EI amenable compounds. Several parameters, crucial for a successful integration of liquid chromatography and mass spectrometry, have been considered in the evaluation of the functioning of such an interface: limit of detection, linearity of response, reproducibility, and chromatographic compatibility. Different mobile phases, also containing non-volatile buffers, were taken into account, demonstrating an outstanding separation flexibility. The entire set of experiments was carried out at different flow rates and temperatures of the ion source. The interface behavior was also tested in real world applications, with mixtures of pesticides, hormones, nitro-PAH, and endocrine-disrupting compounds, allowing picogram level detection and the possibility to record library-matchable, readily interpretable electron ionization mass spectra, for prompt compound characterization and confirmation. (c) 2005 Wiley Periodicals, Inc.
引用
收藏
页码:978 / 989
页数:12
相关论文
共 50 条
  • [21] Current prospects on nano liquid chromatography coupled to electron ionization mass spectrometry (nanoLC-EI-MS)
    Vargas Medina, Deyber Arley
    Pereira dos Santos, Natalia Gabrielly
    Soares Maciel, Edvaldo Vasconcelos
    Lancas, Fernando Mauro
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2021, 44 (17-20) : 862 - 871
  • [22] Gas chromatography-electron ionization mass spectrometry and liquid chromatography-electrospray tandem mass spectrometry for determination of impurities in the anti-cancer drug isophosphoramide mustard
    Cole, RB
    Chou, CW
    Boué, SM
    LeBlanc, BW
    Rodgers, AH
    Struck, RF
    Morgan, LR
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2004, 231 (2-3) : 147 - 155
  • [23] Profiling of non-esterified fatty acids in human plasma using liquid chromatography-electron ionization mass spectrometry
    Trufelli, Helga
    Famiglini, Giorgio
    Termopoli, Veronica
    Cappiello, Achille
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (09) : 2933 - 2941
  • [24] New trends in the application of electron ionization to liquid chromatography-mass spectrometry interfacing
    Cappiello, A
    Famiglini, G
    Mangani, F
    Palma, P
    MASS SPECTROMETRY REVIEWS, 2001, 20 (02) : 88 - 104
  • [25] Profiling of non-esterified fatty acids in human plasma using liquid chromatography-electron ionization mass spectrometry
    Helga Trufelli
    Giorgio Famiglini
    Veronica Termopoli
    Achille Cappiello
    Analytical and Bioanalytical Chemistry, 2011, 400 : 2933 - 2941
  • [26] High-performance liquid chromatography combined with electron ionization mass spectrometry: A review
    Rigano, Francesca
    Tranchida, Peter Q.
    Dugo, Paola
    Mondello, Luigi
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2019, 118 : 112 - 122
  • [27] Analysis of Boronic Acids by Nano Liquid Chromatography-Direct Electron Ionization Mass Spectrometry
    Flender, Cornelia
    Leonhard, Peter
    Wolf, Christian
    Fritzsche, Matthias
    Karast, Michael
    ANALYTICAL CHEMISTRY, 2010, 82 (10) : 4194 - 4200
  • [28] Determination of 29 pesticide residues in tobacco by gas chromatography-electron impact ionization mass spectrometry
    Zou Ximei
    Lin Zhuguang
    Peng Shunu
    Chen Zhaobin
    CHINESE JOURNAL OF CHROMATOGRAPHY, 2009, 27 (02) : 169 - 175
  • [29] ANALYSIS OF DEOXYNIVALENOL AS ITS TRIFLUOROACETYL ESTER BY GAS CHROMATOGRAPHY-ELECTRON IONIZATION MASS-SPECTROMETRY
    WREFORD, BJ
    SHAW, KJ
    FOOD ADDITIVES AND CONTAMINANTS, 1988, 5 (02): : 141 - 147
  • [30] LIQUID CHROMATOGRAPHY MASS SPECTROMETRY INTERFACE .1. DIRECT INTRODUCTION OF LIQUID SOLUTIONS INTO A CHEMICAL IONIZATION MASS-SPECTROMETER
    BALDWIN, MA
    MCLAFFER.FW
    ORGANIC MASS SPECTROMETRY, 1973, 7 (09): : 1111 - 1112