A comparison of microwave ablation and bipolar radiofrequency ablation both with an internally cooled probe: Results in ex vivo and in vivo porcine livers

被引:147
|
作者
Yu, Jie [1 ]
Liang, Ping [1 ]
Yu, Xiaoling [1 ]
Liu, Fangyi [1 ]
Chen, Lei [2 ]
Wang, Yang [3 ]
机构
[1] Chinese Peoples Liberat Army Gen Hosp, Dept Intervent Ultrasound, Beijing 100853, Peoples R China
[2] Chinese Peoples Liberat Army Gen Hosp, Dept Otolaryngol, Beijing 100853, Peoples R China
[3] Chinese Peoples Liberat Army Gen Hosp, Dept Ultrasound, Beijing 100853, Peoples R China
关键词
Animals; Liver; Interventional procedure; Microwave; Radiofrequency; Ablation; HEPATOCELLULAR-CARCINOMA; THERMAL ABLATION; COAGULATION; THERAPY; ANTENNA; COMPLICATIONS; PRINCIPLES; ELECTRODES; SURVIVAL; LARGER;
D O I
10.1016/j.ejrad.2009.12.009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to compare the effectiveness of microwave (MW) ablation and radiofrequency (RF) ablation using a single internally cooled probe in a hepatic porcine model. Materials and methods: In the ex vivo experiment, MW ablations (n = 40) were performed with a 2450 MHz and 915 MHz needle antenna, respectively at 60 W, 70 W power settings. Bipolar RF ablations (n = 20) were performed with a 3-cm (T30) and 4-cm (T40) active tip needle electrodes, respectively at a rated power 30 W and 40 W according to automatically systematic power setting. In the in vivo experiment, the 2450 MHz and 915 MHz MW ablation both at 60 W and T30 bipolar RF ablation at 30 W were performed (n = 30). All of the application time were 10 min. Long-axis diameter (Dl), short-axis diameter (Ds), ratio of Ds/Dl, the temperature data 5 mm from the needle and the time of temperature 5 mm from the needle rising to 54 degrees C were measured. Results: Both in ex vivo and in vivo models, Ds and Dl of 915 MHz MW ablations were significantly larger than all the RF ablations (P < 0.05); the Ds for all the 2450 MHz MW ablations were significantly larger than that of T30 RF ablations (P < 0.05). 2450 MHz MW and T30 RF ablation tended to produce more elliptical-shaped ablation zone. Tissue temperatures 5 mm from the needle were considerably higher with MW ablation, meanwhile MW ablation achieved significantly faster rate of temperature rising to 54 degrees C than RF ablation. For in vivo study after 10 min of ablation, the Ds and Dl of 2450 MHz MW, 915 MHz MW and Bipolar RF were 2.35 +/- 0.75, 2.95 +/- 0.32, 1.61 +/- 0.33 and 3.86 +/- 0.81, 5.79 +/- 1.03, 3.21 +/- 0.51, respectively. Highest tissue temperatures 5 mm from the needle were 80.07 +/- 12.82 degrees C, 89.07 +/- 3.52 degrees C and 65.56 +/- 15.31 degrees C and the time of temperature rising to 54 degrees C were respectively 37.50 +/- 7.62 s, 24.50 +/- 4.09 s and 57.29 +/- 23.24 s for three applicators. Conclusion: MW ablation may have higher potential for complete destruction of liver tumors than RF ablation. (C) 2009 Elsevier Ireland Ltd. All rights reserved.
引用
下载
收藏
页码:124 / 130
页数:7
相关论文
共 50 条
  • [31] No-Touch Radiofrequency Ablation: A Comparison of Switching Bipolar and Switching Monopolar Ablation in Ex Vivo Bovine Liver
    Chang, Won
    Lee, Jeong Min
    Lee, Sang Min
    Han, Joon Koo
    KOREAN JOURNAL OF RADIOLOGY, 2017, 18 (02) : 279 - 288
  • [32] Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling
    Clasen, S
    Schmidt, D
    Boss, A
    Dietz, K
    Kröber, SM
    Claussen, CD
    Pereira, PL
    RADIOLOGY, 2006, 238 (03) : 881 - 890
  • [33] Microwave ablation: Results with a 2.45-GHz applicator in ex vivo bovine and in vivo porcine liver
    Hines-Peralta, AU
    Pirani, N
    Clegg, P
    Cronin, N
    Ryan, TP
    Liu, ZJ
    Goldberg, SN
    RADIOLOGY, 2006, 239 (01) : 94 - 102
  • [34] Microwave ablation: Results with three different diameters of antennas in ex vivo bovine and in vivo porcine liver
    Song, Ze
    Qi, Han
    Zhang, Hao
    Xie, Lin
    Cao, Fei
    Fan, Weijun
    Wan, Chao
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (05) : 737 - 741
  • [35] Hepatic radiofrequency ablation: in vivo and ex vivo comparisons of 15-gauge (G) and 17-G internally cooled electrodes
    Song, K. D.
    Lee, M. W.
    Park, H. J.
    Cha, D. I.
    Kang, T. W.
    Lee, J.
    Moon, J. Y.
    Rhim, H.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1050):
  • [36] Switching monopolar radiofrequency ablation technique using multiple, internally cooled electrodes and a multichannel generator - Ex vivo and in vivo pilot study
    Lee, Jeong Min
    Han, Joon Koo
    Kim, Hyo Cheol
    Choi, Young Hun
    Kim, Se Hyung
    Choi, Jin Young
    Choi, Byung Ihn
    INVESTIGATIVE RADIOLOGY, 2007, 42 (03) : 163 - 171
  • [37] Ex vivo human bile duct radiofrequency ablation with a bipolar catheter
    Mustafa Atar
    Abdurrahman Kadayifci
    Ebubekir Daglilar
    Catherine Hagen
    Carlos Fernandez-del Castillo
    William R. Brugge
    Surgical Endoscopy, 2018, 32 : 2808 - 2813
  • [38] Ex vivo human bile duct radiofrequency ablation with a bipolar catheter
    Atar, Mustafa
    Kadayifci, Abdurrahman
    Daglilar, Ebubekir
    Hagen, Catherine
    Fernandez-del Castillo, Carlos
    Brugge, William R.
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2018, 32 (06): : 2808 - 2813
  • [39] Comparison of bipolar and unipolar radiofrequency ablation in an in vivo experimental model
    Bugge, E
    Nicholson, IA
    Thomas, SP
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 28 (01) : 76 - 80
  • [40] Comparison of ablation performance between dual internally cooled wet tip and conventional dual internally cooled tip radiofrequency electrodes: an experimental study in ex vivo bovine liver
    Cha, Dong Ik
    Lee, Min Woo
    Jeong, Woo Kyoung
    Ha, Sang Yun
    Ahn, Soo Hyun
    Rhim, Hyunchul
    Lim, Hyo Keun
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2021, 38 (01) : 332 - 340