Improving the accuracy of motion quantification using area detector computed tomography for real-time tumor-tracking irradiation in stereotactic ablative radiotherapy

被引:1
|
作者
Suzuki, Kentaro [1 ]
Usui, Keisuke [2 ]
Sasai, Keisuke [3 ]
机构
[1] Juntendo Univ, Grad Sch Med, Dept Radiat Oncol, Tokyo 1138421, Japan
[2] Juntendo Univ, Fac Hlth Sci, Fac Med, Dept Radiol Technol, Tokyo 1138421, Japan
[3] Juntendo Univ, Fac Med, Dept Radiat Oncol, Tokyo 1138421, Japan
关键词
Four-dimensional computed tomography; Real-time tumor-tracking irradiation; Stereotactic body radiotherapy; Respiratory motion; Tumor delineation; BODY RADIOTHERAPY;
D O I
10.1016/j.meddos.2022.02.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
CyberKnife radiotherapy enables tumor-tracking irradiation using positional information regarding the tumor and a fiducial marker in a patient's body. This positional information acts as a surrogate of tumor motion. Therefore, deviations in these movements should be quantitatively estimated and included as an internal margin for radiation treatment planning. This study aimed to investigate variations between the positions of fiducial markers and tumor regions using 320-row area detector computed tomography and to analyze the effectiveness of our proposed method in contouring tumor regions based on the fiducial marker position. To determine the moving tumor volume, a typical single-phase image was selected, and pixel values in other phase images were accumulated. Moreover, a maximum-intensity projection image was created to clarify motion deviations in the tumor volume. To evaluate the delineation accuracy, the dice similarity coefficient and mean distance to agreement were calculated in phase-selected and breath-holding computed tomography. Moving chest phantom images were acquired using helical scanning 4-dimensional computed tomography (H-4DCT) and volumetric scanning 4-dimensional computed tomography (V-4DCT), and the delineation accuracies were compared for each scanning type. The average dice similarity coefficient and mean distance to agreement were degraded in limited-phase images, which cannot represent the hysteretic motion of a tumor. Moreover, deviations in tumor volume with unstable motion reached 71.6% in H-4DCT but only 1.6% in V-4DCT. Our proposed method with V-4DCT using area detector computed tomography can achieve accurate moving tumor delineation and can clarify positional associations between the fiducial marker and tumor under respiratory motion. (C) 2022 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:166 / 172
页数:7
相关论文
共 50 条
  • [31] Optimization of fluoroscopy parameters using pattern matching prediction in the real-time tumor-tracking radiotherapy system
    Miyamoto, Naoki
    Ishikawa, Masayori
    Bengua, Gerard
    Sutherland, Kenneth
    Suzuki, Ryusuke
    Kimura, Suguru
    Shimizu, Shinichi
    Onimaru, Rikiya
    Shirato, Hiroki
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (15): : 4803 - 4813
  • [32] Stereotactic Body Radiation Therapy Using a Real-time Tumor-Tracking Radiation Therapy System for Hepatocellular Carcinomas
    Uchinami, Y.
    Katoh, N.
    Abo, D.
    Harada, K.
    Inoue, T.
    Taguchi, H.
    Onimaru, R.
    Shimizu, S.
    Sakuhara, Y.
    Ogawa, K.
    Kamiyama, T.
    Shirato, H.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E155 - E155
  • [33] Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy
    Shirato, H
    Suzuki, K
    Sharp, GC
    Fujita, K
    Onimaru, R
    Fujino, M
    Kato, N
    Osaka, Y
    Kinoshita, R
    Taguchi, H
    Onodera, S
    Miyasaka, K
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (04): : 1229 - 1236
  • [34] Application of real-time tumor-tracking and gated radiotherapy system for unresectable pancreatic cancer
    Ahn, YC
    Shimizuz, S
    Shirato, H
    Hashimoto, T
    Osaka, Y
    Zhang, XQ
    Abe, T
    Hosokawa, M
    Miyasaka, K
    [J]. YONSEI MEDICAL JOURNAL, 2004, 45 (04) : 584 - 590
  • [35] Feasibility Study of Markerless Tracking Using Dual Energy Fluoroscopic Images for Real-Time Tumor-Tracking Radiotherapy System
    Shiinoki, T.
    Sawada, A.
    Uehara, T.
    Yuasa, Y.
    Koike, M.
    Kawamura, S.
    Shibuya, K.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3650 - 3650
  • [36] A motion-compensated image filter for low-dose fluoroscopy in a real-time tumor-tracking radiotherapy system
    Miyamoto, Naoki
    Ishikawa, Masayori
    Sutherland, Kenneth
    Suzuki, Ryusuke
    Matsuura, Taeko
    Toramatsu, Chie
    Takao, Seishin
    Nihongi, Hideaki
    Shimizu, Shinichi
    Umegaki, Kikuo
    Shirato, Hiroki
    [J]. JOURNAL OF RADIATION RESEARCH, 2015, 56 (01) : 186 - 196
  • [37] Stereotactic body radiotherapy using a new real-time tumor tracking system and fiducial markers
    Hiroshima, Y.
    Nitta, K.
    Saitoh, T.
    Ohno, T.
    Shinoda, K.
    Tamaki, Y.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S758 - S758
  • [38] Three-dimensional intrafractional motion of breast during tangential breast irradiation monitored with high-sampling frequency using a real-time tumor-tracking radiotherapy system
    Kinoshita, Rumiko
    Shimizu, Shinichi
    Taguchi, Hiroshi
    Katoh, Norio
    Fujino, Masaharu
    Onimaru, Rikiya
    Aoyama, Hidefumi
    Katoh, Fumi
    Omatsu, Tokuhiko
    Ishikawa, Masayori
    Shirato, Hiroki
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (03): : 931 - 934
  • [39] Real-time tumor-tracking radiotherapy with SyncTraX for primary liver tumors requiring isocenter shift
    Uchinami, Yusuke
    Miyamoto, Naoki
    Abo, Daisuke
    Morita, Ryo
    Ogawa, Koji
    Kakisaka, Tatsuhiko
    Suzuki, Ryusuke
    Miyazaki, Tomohiko
    Taguchi, Hiroshi
    Katoh, Norio
    Aoyama, Hidefumi
    [J]. JOURNAL OF RADIATION RESEARCH, 2024, 65 (01) : 92 - 99
  • [40] Lung stereotactic body radiotherapy with electromagnetic transponders for real-time tumor tracking
    Jaccard, Maud
    Jaccard, M.
    Champion, A.
    Dubouloz, A.
    Plojoux, J.
    Soccal, P.
    Miralbell, R.
    Dipasquale, G.
    Caparrotti, F.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 (12) : 1126 - 1126