Effect of clothing on sphygmomanometric and oscillometric blood pressure measurement in hypertensive subjects

被引:8
|
作者
Eder, M. J. C. [1 ]
Holzgreve, H. [2 ]
Liebl, M. E. [3 ]
Bogner, J. R. [1 ]
机构
[1] Univ Munich, Klinikum Innenstadt, Med Poliklin, D-80336 Munich, Germany
[2] Kardiol Praxis, Munich, Germany
[3] Abt Orthopad, Med Klin, Berlin, Germany
关键词
hypertension; blood pressure measurement; oscillometry; sphygmomanometry; clothes;
D O I
10.1055/s-2008-1077254
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and objective: The few studies published on this subject have shown that blood pressure measurements give similar results whether the patients' arm is covered by clothing or not. But it has not been clarified whether this is also true in the case of hypertensive persons, Yet, in practice the correct measurement is of critical importance in the diagnosis of hypertension. Methods: 203 hypertensive patients were examined with the auscultatory sphygmomanometry and the automatic oscillometry by measuring the pressure three times with each method. These tests were carried out in a randomized sequence on a covered arm (the patient's own clothing of maximally 2 mm thickness), the bare arm and the arm covered with a standardized cotton sleeve (2 mm). The auscultatory sphygmomanometry was done blinded (non-sleeved, sleeved, standardized). Results: Calculation of confidence intervals for the mean differences of the three settings (bare, clothing and standard sleeve) and equivalence testing demonstrated that a garment or cloth on the arm under the manometer cuffs did not significantly effect the blood pressure within the predefined interval of equivalence of 4 mm Hg. Thus, measuring blood pressure with the cuff over the person's sleeve does not significantly effect the result. Conclusion: This study shows that measuring blood pressure in hypertensive persons with or without a cloth sleeve (maximally 2 mm thick) does not result in any statistically significant difference. This simplifies the blood pressure measurement that have to be taken frequently on hypertensive persons.
引用
收藏
页码:1288 / 1292
页数:5
相关论文
共 50 条
  • [1] The effect of clothes on sphygmomanometric and oscillometric blood pressure measurement
    Liebl, ME
    Holzgreve, H
    Schulz, M
    Crispin, A
    Bogner, JR
    BLOOD PRESSURE, 2004, 13 (05): : 279 - 282
  • [2] The effect of clothes on sphygmomanometric blood pressure measurement in hypertensive patients
    Pinar, Rukiye
    Ataalkin, Siddika
    Watson, Roger
    JOURNAL OF CLINICAL NURSING, 2010, 19 (13-14) : 1861 - 1864
  • [3] Quantitative analysis of the effect of clothing on the oscillometric waveform envelope and oscillometric blood pressure measurements
    Kim, Jeonghoon
    Lee, Jongshill
    Lee, Jeyeon
    Park, Hoon-Ki
    Kim, In Young
    BIOMEDICAL ENGINEERING LETTERS, 2025, : 525 - 536
  • [4] The effect of clothes on sphygmomanometric blood pressure measurement in hypertensive patients (vol 19, pg 1861, 2010)
    Pinar, R.
    Ataalkin, S.
    Watson, R.
    Ma, G.
    Sabin, N.
    Dawes, M.
    JOURNAL OF CLINICAL NURSING, 2016, 25 (5-6) : 886 - 886
  • [5] DOPPLER ULTRASONOGRAPHIC, OSCILLOMETRIC SPHYGMOMANOMETRIC, AND PHOTOPLETHYSMOGRAPHIC TECHNIQUES FOR NONINVASIVE BLOOD-PRESSURE MEASUREMENT IN ANESTHETIZED CATS
    BINNS, SH
    SISSON, DD
    BUOSCIO, DA
    SCHAEFFER, DJ
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 1995, 9 (06) : 405 - 414
  • [6] Significantly Reduced Blood Pressure Measurement Variability for Both Normotensive and Hypertensive Subjects: Effect of Polynomial Curve Fitting of Oscillometric Pulses
    Yang, Fangwei
    Chen, Fei
    Zhu, Mingping
    Chen, Aiqing
    Zheng, Dingchang
    BIOMED RESEARCH INTERNATIONAL, 2017, 2017
  • [7] EFFECT OF UNDERCUFFING ON AUSCULTATORY AND OSCILLOMETRIC BLOOD PRESSURE MEASUREMENT
    Albuquerque, Nila
    Padwal, Raj
    Araujo, Thelma
    JOURNAL OF HYPERTENSION, 2021, 39 : E130 - E130
  • [8] Effect of Peripheral Edema on Oscillometric Blood Pressure Measurement
    Ghaffari, Shamsi
    Malaki, Majid
    Rezaeifar, Afshin
    Fakhim, Shahin Abdollahi
    JOURNAL OF CARDIOVASCULAR AND THORACIC RESEARCH, 2014, 6 (04) : 217 - 221
  • [9] VALIDITY OF SPHYGMOMANOMETRIC DETERMINATIONS FOR THE MEASUREMENT OF BLOOD-PRESSURE
    DELAFUENTE, SR
    MEDICINA CLINICA, 1985, 84 (17): : 712 - 713
  • [10] Rapid oscillometric blood pressure measurement compared to conventional oscillometric measurement
    Yarows, SA
    Patel, K
    Brook, R
    BLOOD PRESSURE MONITORING, 2001, 6 (03) : 145 - 147