Validation of CT brain perfusion methods using a realistic dynamic head phantom

被引:39
|
作者
Riordan, Alan J. [1 ]
Prokop, Mathias [1 ,2 ]
Viergever, Max A. [1 ]
Dankbaar, Jan Willem [1 ]
Smit, Ewoud J. [1 ]
de Jong, Hugo W. A. M. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3584 CX Utrecht, Netherlands
[2] Radboud Univ Nijmegen, Dept Radiol, NL-6525 GA Nijmegen, Netherlands
关键词
CT perfusion; brain CT; phantom; SINGULAR-VALUE DECOMPOSITION; HUMAN CEREBRAL-CORTEX; ACUTE STROKE; BLOOD-FLOW; COMPUTED-TOMOGRAPHY; VOLUME; THRESHOLDS; HEMATOCRIT; SIMULATION; RESOLUTION;
D O I
10.1118/1.3592639
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Development and evaluation of a realistic hybrid head phantom for the validation of quantitative CT brain perfusion methods. Methods: A combination, or hybrid, of CT images of an anthropomorphic head phantom together with clinically acquired MRI brain images was used to construct a dynamic hybrid head phantom. Essential CT imaging parameters such as spatially dependent noise, effects of resolution, tube settings, and reconstruction parameters were intrinsically included by scanning a skull phantom using CT perfusion (CTP) protocols with varying mAs. These data were combined with processed high resolution 7T clinical MRI images to include healthy and diseased brain parenchyma, as well as the cerebral vascular system. Time attenuation curves emulating contrast bolus passage based on perfusion as observed in clinical studies were added. Using the phantom, CTP images were generated using three brain perfusion calculation methods: bcSVD, sSVD, and fit-based deconvolution, and the linearity and accuracy of the three calculation methods was assessed. Dependency of perfusion outcome on calculation method was compared to clinical data. Furthermore, the potential of the phantom to optimize brain perfusion packages was investigated. Results: All perfusion calculation methods showed overestimation of low perfusion values and underestimation of high perfusion values. Good correlation in behavior between phantom and clinical data was found (R-2 = 0.84). Conclusions: A dynamic hybrid head phantom constructed from CT and MRI data was demonstrated to realistically represent clinical CTP studies, which is useful for assessing CT brain perfusion acquisition, reconstruction, and analysis. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3592639]
引用
收藏
页码:3212 / 3221
页数:10
相关论文
共 50 条
  • [31] Simulation of contrast agent dynamics in digital brain phantom for CT perfusion optimization
    Divel, Sarah E.
    Christensen, Sren
    Lansberg, Maarten G.
    Pelc, Norbert J.
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [32] QUANTITATIVE DYNAMIC PERFUSION CT BRAIN IN PATIENTS WITH STROKE
    Chidambaram, Prem Kumar
    Kumar, Ram S.
    Vadanika, V.
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2015, 4 (87): : 15147 - 15152
  • [33] Validation of a source tracking method for brachytherapy in vivo dosimetry using a realistic phantom
    van Wagenberg, T.
    Fonseca, G.
    Voncken, R.
    Van Beveren, C.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S57 - S58
  • [34] A Long-Lasting Textile-Based Anatomically Realistic Head Phantom for Validation of EEG Electrodes
    Tseghai, Granch Berhe
    Malengier, Benny
    Fante, Kinde Anlay
    Van Langenhove, Lieva
    SENSORS, 2021, 21 (14)
  • [35] Microwave System to Detect Traumatic Brain Injuries Using Compact Unidirectional Antenna and Wideband Transceiver With Verification on Realistic Head Phantom
    Mobashsher, Ahmed Toaha
    Abbosh, Amin M.
    Wang, Yifan
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2014, 62 (09) : 1826 - 1836
  • [36] In-line spectroscopy for the quantification of iodine in a breast perfusion phantom for dynamic contrast-enhanced dedicated breast CT imaging validation
    Goris, Liselot
    Mikerov, Mikhail
    Manohar, Srirang
    Sechopoulos, Ioannis
    MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1, 2024, 12925
  • [37] Multicenter evaluation of a quantitative SPECT reconstruction program using a realistic brain phantom
    Iida, Hidehiro
    Takahashi, Masaaki
    Imabayashi, Etsuko
    Ishida, Kenji
    Zeniya, Tsutomu
    Matsuda, Hiroshi
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [38] A 4D CT digital phantom of an individual human brain for perfusion analysis
    Manniesing, Rashindra
    Brune, Christoph
    van Ginneken, Bram
    Prokop, Mathias
    PeerJ, 2016, 4
  • [39] Validation of the RayStation Monte Carlo dose calculation algorithm using a realistic lung phantom
    Schreuder, Andries N.
    Bridges, Daniel S.
    Rigsby, Lauren
    Blakey, Marc
    Janson, Martin
    Hedrick, Samantha G.
    Wilkinson, John B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (12): : 127 - 137
  • [40] Whole-brain dynamic CT angiography and perfusion imaging
    Orrison, W. W., Jr.
    Snyder, K. V.
    Hopkins, L. N.
    Roach, C. J.
    Ringdahl, E. N.
    Nazir, R.
    Hanson, E. H.
    CLINICAL RADIOLOGY, 2011, 66 (06) : 566 - 574