Validation of CT brain perfusion methods using a realistic dynamic head phantom

被引:39
|
作者
Riordan, Alan J. [1 ]
Prokop, Mathias [1 ,2 ]
Viergever, Max A. [1 ]
Dankbaar, Jan Willem [1 ]
Smit, Ewoud J. [1 ]
de Jong, Hugo W. A. M. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3584 CX Utrecht, Netherlands
[2] Radboud Univ Nijmegen, Dept Radiol, NL-6525 GA Nijmegen, Netherlands
关键词
CT perfusion; brain CT; phantom; SINGULAR-VALUE DECOMPOSITION; HUMAN CEREBRAL-CORTEX; ACUTE STROKE; BLOOD-FLOW; COMPUTED-TOMOGRAPHY; VOLUME; THRESHOLDS; HEMATOCRIT; SIMULATION; RESOLUTION;
D O I
10.1118/1.3592639
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Development and evaluation of a realistic hybrid head phantom for the validation of quantitative CT brain perfusion methods. Methods: A combination, or hybrid, of CT images of an anthropomorphic head phantom together with clinically acquired MRI brain images was used to construct a dynamic hybrid head phantom. Essential CT imaging parameters such as spatially dependent noise, effects of resolution, tube settings, and reconstruction parameters were intrinsically included by scanning a skull phantom using CT perfusion (CTP) protocols with varying mAs. These data were combined with processed high resolution 7T clinical MRI images to include healthy and diseased brain parenchyma, as well as the cerebral vascular system. Time attenuation curves emulating contrast bolus passage based on perfusion as observed in clinical studies were added. Using the phantom, CTP images were generated using three brain perfusion calculation methods: bcSVD, sSVD, and fit-based deconvolution, and the linearity and accuracy of the three calculation methods was assessed. Dependency of perfusion outcome on calculation method was compared to clinical data. Furthermore, the potential of the phantom to optimize brain perfusion packages was investigated. Results: All perfusion calculation methods showed overestimation of low perfusion values and underestimation of high perfusion values. Good correlation in behavior between phantom and clinical data was found (R-2 = 0.84). Conclusions: A dynamic hybrid head phantom constructed from CT and MRI data was demonstrated to realistically represent clinical CTP studies, which is useful for assessing CT brain perfusion acquisition, reconstruction, and analysis. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3592639]
引用
收藏
页码:3212 / 3221
页数:10
相关论文
共 50 条
  • [21] Validation of advanced hybrid SPECT/CT system using dynamic anthropomorphic cardiac phantom
    Gelbart, Elad
    Krakovich, Alexander
    Sherm, Yigal
    Rabin, Gilad
    Ratner, Hagit
    Scheinowitz, Mickey
    Goldkorn, Ronen
    ANNALS OF NUCLEAR MEDICINE, 2024, 38 (11) : 919 - 926
  • [22] Quantification and Validation of DCE-CT Using a Novel Dynamic Flow Imaging Phantom
    Driscoll, B.
    Keller, H.
    Coolens, C.
    MEDICAL PHYSICS, 2010, 37 (06) : 3406 - 3407
  • [23] The effect of head movement on CT perfusion summary maps: simulations with CT hybrid phantom data
    Fahmi, F.
    Riordan, A.
    Beenen, L. F. M.
    Streekstra, G. J.
    Janssen, N. Y.
    de Jong, H. W.
    Majoie, C. B. L.
    van Bavel, E.
    Marquering, H. A.
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2014, 52 (02) : 141 - 147
  • [24] The effect of head movement on CT perfusion summary maps: simulations with CT hybrid phantom data
    F. Fahmi
    A. Riordan
    L. F. M. Beenen
    G. J. Streekstra
    N. Y. Janssen
    H. W. de Jong
    C. B. L. Majoie
    E. van Bavel
    H. A. Marquering
    Medical & Biological Engineering & Computing, 2014, 52 : 141 - 147
  • [25] Detection and Localization of Brain Strokes in Realistic 3-D Human Head Phantom
    Mobashsher, A. T.
    Nguyen, P. T.
    Abbosh, A.
    2013 IEEE MTT-S INTERNATIONAL MICROWAVE WORKSHOP SERIES ON RF AND WIRELESS TECHNOLOGIES FOR BIOMEDICAL AND HEALTHCARE APPLICATIONS (IMWS-BIO), 2013, : 65 - 67
  • [26] Realistic CT simulation using the 4D XCAT phantom
    Segars, W. P.
    Mahesh, M.
    Beck, T. J.
    Frey, E. C.
    Tsui, B. M. W.
    MEDICAL PHYSICS, 2008, 35 (08) : 3800 - 3808
  • [27] Validation of Computational Studies for Electrical Brain Stimulation With Phantom Head Experiments
    Kim, Donghyeon
    Jeong, Jinmo
    Jeong, Sangdo
    Kim, Sohee
    Jun, Sung Chan
    Chung, Euiheon
    BRAIN STIMULATION, 2015, 8 (05) : 914 - 925
  • [28] The Relationship Between Cerebral Perfusion CT Dose and MDCT Collimation Using TLD Chips in An Anthropomorphic Head Phantom
    Al-Senan, R.
    Goerner, F.
    Giantsoudi, D.
    Hatab, M.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [29] Standardized regional Perfusion Parameters in Dynamic Whole Brain Perfusion CT
    Feyen, Ludger
    Minnerup, Jens
    Wersching, Heike
    Duning, Thomas
    Heindel, Walter
    Hesselmann, Volker
    Kemmling, Andre
    STROKE, 2011, 42 (03) : E117 - E118
  • [30] A direct comparison of the sensitivity of CT and MR cardiac perfusion using a myocardial perfusion phantom
    Otton, James
    Morton, Geraint
    Schuster, Andreas
    Bigalke, Boris
    Marano, Riccardo
    Olivotti, Luca
    Nagel, Eike
    Chiribiri, Amedeo
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2013, 7 (02) : 117 - 124