Accurate and Rapid Measurement of Glycated Hemoglobin Using HLC-723 G11 Variant Mode

被引:4
|
作者
Park, Min-Seung [1 ]
Lee, Kyunghoon [2 ,3 ]
Song, Junghan [3 ]
Park, Hyung-Doo [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Lab Med & Genet, 81 Irwon Ro, Seoul 06351, South Korea
[2] Sungkyunkwan Univ, Kangbuk Samsung Hosp, Dept Lab Med, Sch Med, Seoul, South Korea
[3] Seoul Natl Univ, Bundang Hosp, Dept Lab Med, Coll Med, Seongnam, South Korea
关键词
HbA1c; Hb variants; Glycated Hb; Tosoh HLC-723 G11 analyzer; Performance; HBA1C; HPLC;
D O I
10.3343/alm.2019.39.3.243
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Type II diabetes mellitus causes many complications, and its prevalence continues to increase in Korea. Accurate measurement of glycated Hb (HbA1c) is important because of its usefulness in diagnosis, follow-up, and prediction of prognosis. We tested the analytical performance of the HLC-723 G11 Variant Mode (G11vr; Tosoh Bioscience, Inc., Tokyo, Japan), recently introduced to Korea, in detecting HbA1c. Methods: We evaluated precision, linearity, carry-over, and turnaround time. Using 208 samples, including 108 flagged samples, we compared HbA1c concentrations from four analyzers through correlation analysis: G11vr, HLC-723 G8 Variant Mode (G8vr, Tosoh Bioscience), HLC-723 G11 Standard Mode (G11st, Tosoh Bioscience), and HLC-723 G8 Standard Mode (G8st, Tosoh Bioscience). We used HPLC mass spectrometry (MS) and capillary electrophoresis (CE) to confirm the HbA1c concentrations of 15 additional known Hb variant samples. Results: Repeatability (% CV) in measuring low- and high-concentration controls was 0.57% and 0.35%, respectively; within-laboratory precision was 0.86% and 0.69%, respectively. In a linearity test, the coefficient of determination was 0.9999 (measurement range: 3.64% to 18.59%) for HbA1c. The correlations between G11vr and other analyzers were weaker for flagged samples than for non-flagged samples. The carry-over effect was less than 0.4%. Turnaround time for a single sample was lower in G11vr (one minute) than in G8vr (1.6 minutes). For 15 samples with Hb variants, G11vr HbA1c results were more similar than those of other analyzers to HPLC-MS and CE results. Conclusions: G11vr showed adequate performance and rapid turnaround time in measuring HbA1c.
引用
收藏
页码:243 / 250
页数:8
相关论文
共 17 条
  • [1] Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
    Yuan, Bo
    Yang, Wei
    Zhang, Na
    Shi, Hongyan
    Dong, Shuangpeng
    PRACTICAL LABORATORY MEDICINE, 2023, 37
  • [2] Evaluation of the analytic performances of the new HPLC system HLC-723 G7 for the measurement of hemoglobin A1C
    Terreni, A
    Paleari, R
    Caldini, A
    Ognibene, A
    Mosca, A
    Messeri, G
    CLINICAL BIOCHEMISTRY, 2003, 36 (08) : 607 - 610
  • [3] Analytical evaluation of the Tosoh HLC-723 G8 automated HPLC analyzer for hemoglobin analysis in beta-thalassemia mode
    Merono, Francoise
    Agouti, Imane
    Bonello-Palot, Nathalie
    Paolasso, Chantal
    Levy, Nicolas
    Badens, Catherine
    CLINICAL BIOCHEMISTRY, 2011, 44 (5-6) : 441 - 443
  • [4] Unanticipated error in HbA1c measurement on the HLC-723 G7 analyzer
    van den Ouweland, Johannes M. W.
    de Keijzer, Marinus H.
    van Daal, Henny
    CLINICAL BIOCHEMISTRY, 2010, 43 (06) : 623 - 625
  • [5] Rapid haemoglobin A and S quantification using Tosoh HLC-723G8 in variant mode for patients with sickle cell disease
    Fadel, Julien
    Noyelle, Juliette
    Maingon, Mathieu
    Homedan, Chadi
    Dieu, Xavier
    de la Barca, Juan Manuel Chao
    Reynier, Pascal
    Mallebranche, Coralie
    Brasme, Jean-Francois
    Mirebeau-Prunier, Delphine
    Orvain, Corentin
    Chabrun, Floris
    BIOMEDICAL CHROMATOGRAPHY, 2024, 38 (03)
  • [6] Analytical evaluation of the Tosoh HLC-723 G7 automated HPLC analyzer for hemoglobin A2 and F determination
    Paleari, R
    Cannata, M
    Leto, F
    Maggio, A
    Demartis, FR
    Desogus, MF
    Galanello, R
    Mosca, A
    CLINICAL BIOCHEMISTRY, 2005, 38 (02) : 159 - 165
  • [7] Interference of Thalassemias on Hemoglobin A1c Measurement by Ion-Exchange High-Performance Liquid Chromatography Method Tosoh HLC-723 G8
    Chao, Yan
    Wan, Zemin
    Wu, Xiaobin
    Qiu, Feng
    Wang, Yunxiu
    Ke, Peifeng
    Wu, Xinzhong
    Xu, Jianhua
    Zhuang, Junhua
    Huang, Xianzhang
    CLINICAL LABORATORY, 2018, 64 (7-8) : 1171 - 1175
  • [8] High consistency of the results provided by HLC-723G7 (Tosoh) and RA8140 (Menarini-Arkray) systems in glycated hemoglobin measurement.
    Dorizzi, RM
    Perobelli, L
    Dimitri, G
    Menna, C
    Bonaldo, A
    Scandola, S
    Rizzotti, P
    CLINICAL CHEMISTRY, 2004, 50 (06) : A30 - A31
  • [9] The silent hemoglobin α chain variant Hb Riccarton [α51(CE9)Gly→Ser] may affect HbA1c determination on the HLC-723 G7 analyzer
    van den Ouweland, Johannes M. W.
    van Daal, Henny
    Klaassen, Corne H.
    van Aarssen, Yvonne
    Harteveld, Cornelis L.
    Giordano, Piero C.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (06) : 827 - 830
  • [10] Hb Oegstgeest [α104(G11)Cys → Ser (α1)].: A new hemoglobin variant associated with a mild α-thalassemia phenotype
    Harteveld, CL
    Rozendaal, L
    Blom, NA
    Lo-A-Njoe, S
    Akkerman, N
    Arkestijn, S
    Van Delft, P
    Giordano, PC
    HEMOGLOBIN, 2005, 29 (03) : 165 - 169