Mass-Preserving Motion Correction of Dual Gated Cardiac PET

被引:0
|
作者
Gigengack, Fabian [1 ]
Ruthotto, Lars [2 ]
Burger, Martin [3 ]
Wolters, Carsten H. [4 ]
Jiang, Xiaoyi
Schaefers, Klaus P. [1 ,5 ]
机构
[1] Univ Munster, EIMI, Dept Math & Comp Sci, D-4400 Munster, Germany
[2] Univ Lubeck, Inst Math & Image Comp MIC, Lubeck, Germany
[3] Univ Munster, Inst Computat & Appl Math, Munster, Germany
[4] Univ Munster, Inst Biomagnetism & Biosignalanal, Munster, Germany
[5] Univ Munster, EIMI, Munster, Germany
关键词
motion correction; mass-preservation; image registration; dual gating; hyperelastic regularization; PET;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Image degradation due to motion is a known problem in Positron Emission Tomography (PET). To reduce motion artifacts in PET, gating based techniques were found applicable. In pure respiratory gated PET, each gate still contains cardiac motion. Analogously, pure cardiac gated PET images still contain respiratory motion. Hence, we make use of dual (joint respiratory and cardiac) gating to further reduce the amount of motion contained in the images. The inclusion of non-rigid cardiac motion leads to intensity modulations caused by the Partial Volume Effect (PVE). To overcome the intensity modulations we identified and considered the mass-preserving property of PET images in the Variational Algorithm for Mass-Preserving Image REgistration (VAMPIRE). The aim of this paper is the elimination of both cardiac and respiratory motion in thoracic dual gated PET without a loss of statistic. The strategy of our proposed motion correction approach is: 1) Dual gating, 2) Mass-preserving motion estimation (VAMPIRE), and 3) Averaging of the aligned images. In a patient study we showed that the proposed mass-preserving motion correction strategy significantly removes motion artifacts in cardiac PET. VAMPIRE achieved accurate results that resemble cardiac and respiratory motion across all patients.
引用
收藏
页码:3206 / 3208
页数:3
相关论文
共 50 条
  • [1] Motion Correction in Dual Gated Cardiac PET Using Mass-Preserving Image Registration
    Gigengack, Fabian
    Ruthotto, Lars
    Burger, Martin
    Wolters, Carsten H.
    Jiang, Xiaoyi
    Schaefers, Klaus P.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (03) : 698 - 712
  • [2] Motion Correction of Cardiac PET Using Mass-Preserving Registration
    Gigengack, Fabian
    Ruthotto, Lars
    Burger, Martin
    Wolters, Carsten H.
    Jiang, Xiaoyi
    Schaefers, Klaus
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3317 - 3319
  • [3] Mass-Preserving Motion Correction of PET: Displacement Field vs. Spline Transformation
    Gigengack, Fabian
    Ruthotto, Lars
    Burger, Martin
    Wolters, Carsten H.
    Jiang, Xiaoyi
    Schaefers, Klaus P.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3088 - 3090
  • [4] Correction of Respiratory Motion in Dual Gated Cardiac Imaging in PET/CT
    Lamare, F.
    Teras, M.
    Kokki, T.
    Fayad, H.
    Rimoldi, O.
    Camici, P. G.
    Knuuti, J.
    Visvikis, D.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4530 - +
  • [5] Respiratory and cardiac motion correction in dual gated PET/MR imaging
    Hadi Fayad
    Florian Monnier
    Freedy Odille
    Jacques Felblinger
    Frederic Lamare
    Dimitris Visvikis
    EJNMMI Physics, 2 (Suppl 1)
  • [6] CT based motion correction of dual gated cardiac PET images
    Klen, Riku
    Noponen, Tommi
    Koikkalainen, Juha
    Lotjonen, Jyrki
    Thielemans, Kris
    Hoppela, Erika
    Sipila, Hannu
    Teras, Mika
    Knuuti, Juhani
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [7] Pipeline for Motion Correction in Dual Gated PET
    Gigengack, Fabian
    Ruthotto, Lars
    Koesters, Thomas
    Jiang, Xiaoyi
    Modersitzki, Jan
    Burger, Martin
    Wolters, Carsten H.
    Schaefers, Klaus P.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2983 - +
  • [8] Evaluation of respiratory and cardiac motion correction schemes in dual gated PET/CT cardiac imaging
    Lamare, F.
    Le Maitre, A.
    Dawood, M.
    Schaefers, K. P.
    Fernandez, P.
    Rimoldi, O. E.
    Visvikis, D.
    MEDICAL PHYSICS, 2014, 41 (07)
  • [9] SUPER-RESOLUTION IN CARDIAC PET USING MASS-PRESERVING IMAGE REGISTRATION
    Yan, Hua
    Gigengack, Fabian
    Jiang, Xiaoyi
    Schaefers, Klaus
    2013 20TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP 2013), 2013, : 752 - 756
  • [10] Motion frozen of dual gated (cardiac and respiratory) PET images
    Le Meunier, Ludovic
    Slomka, Piotr
    Fermin, Jimmy
    Huber, Brandi
    Cheng, Victor
    Tamarappoo, Balaji
    Ramesh, Amit
    Germano, Guido
    Berman, Daniel
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50