Improved electrochemical analysis of neuropathy target esterase activity by a tyrosinase carbon paste electrode modified by 1-methoxyphenazine methosulfate

被引:9
|
作者
Sokolovskaya, LG
Sigolaeva, LV
Eremenko, AV
Gachok, IV
Makhaeva, GF
Strakhova, NN
Malygin, VV
Richardson, RJ
Kurochkin, IN [1 ]
机构
[1] Moscow MV Lomonosov State Univ, Fac Chem, Moscow 119899, Russia
[2] Russian Acad Sci, Inst Physiol Act Cpds, Chernogolovka 142432, Moscow Region, Russia
[3] Univ Michigan, Toxicol Program, Ann Arbor, MI 48109 USA
关键词
biosensor; mediator; 1-methoxyphenazine methosulfate; neuropathy target esterase; tyrosinase;
D O I
10.1007/s10529-005-0020-1
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A graphite-paste tyrosinase biosensor was improved by adding 1-methoxyphenazine methosulfate as a mediator. Mediator modification enhanced sensitivity to phenol 4-fold and long-term stability 3-fold. Phenol could be detected at 25 nM (S/N=2) using an Ag/AgCl reference electrode. The biosensor was used to measure the activity of a toxicologically significant enzyme, neuropathy target esterase (NTE), which yields phenol by hydrolysis of the substrate, phenyl valerate. Using the new biosensor, blood and brain NTE inhibition by organophosphorus (OP) compounds with different neuropathic potencies were well correlated (r=0.990, n=7), supporting the use of blood NTE as a biochemical marker of exposure to neuropathic OP compounds.
引用
收藏
页码:1211 / 1218
页数:8
相关论文