Torsional Resistance of Heat-Treated Nickel-Titanium Instruments under Different Temperature Conditions

被引:7
|
作者
Jo, Hyo Jin [1 ]
Kwak, Sang Won [2 ]
Kim, Hyeon-Cheol [2 ]
Kim, Sung Kyo [1 ]
Ha, Jung-Hong [1 ]
机构
[1] Kyungpook Natl Univ, Sch Dent, Dept Conservat Dent, Daegu 41940, South Korea
[2] Pusan Natl Univ, Sch Dent, Dent Res Inst, Dept Conservat Dent, Yangsan 50612, South Korea
基金
新加坡国家研究基金会;
关键词
body temperature; C-wire; differential scanning calorimetry; electrical discharge machining; gold-wire; heat-treated nickel-titanium file; torsional resistance; ROTARY ENDODONTIC INSTRUMENTS; PHASE-TRANSFORMATION BEHAVIOR; CYCLIC FATIGUE; MECHANICAL-PROPERTIES; SEM OBSERVATIONS; FILES;
D O I
10.3390/ma14185295
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This study compared the torsional resistance of heat-treated nickel-titanium (NiTi) instruments under different temperature conditions. Four thermomechanically treated single-use NiTi rotary instruments were selected for this study: OneShape (OS), OneCurve (OC), WaveOne Gold (WOG) and HyFlex EDM (HFE). Each instrument was further subdivided by temperature into 2 subgroups. Maximum torque and the distortion angle until fracture occurred were evaluated. Differential scanning calorimetry analysis was performed to measure the phase transformation temperature. Statistical analysis was performed using a two-way ANOVA and t-test (p < 0.05). Fractured fragments were observed using scanning electron microscopy (SEM). The two-way ANOVA showed no significant differences for different temperature conditions. At both room (RT) and body temperature (BT), OS was predominantly austenite while HFE was martensite. OC and WOG were predominantly martensite at RT and mixed phase at BT. At BT, more than half of WOG was martensite, while half of OC was austenite. SEM examination showed no topographical differences between instruments in different temperature groups. In relation to a limitation of this study, there was no difference in torsional resistance of NiTi rotary instruments between the BT and RT conditions. This implies that clinicians do not need to consider a decrease of torsional resistance of heat-treated NiTi instruments at BT.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Influence of Cyclic Flexural Deformation on the Torsional Resistance of Controlled Memory and Conventional Nickel-titanium Instruments
    Pelaez Acosta, Eufemia Carolina
    Resende, Pedro Damas
    da Cunha Peixoto, Isabella Faria
    Joviano Pereira, Erika Sales
    Lopes Buono, Vicente Tadeu
    de Azevedo Bahia, Maria Guiomar
    JOURNAL OF ENDODONTICS, 2017, 43 (04) : 613 - 618
  • [42] Changes in Geometry and Transportation of Root Canals with Severe Curvature Prepared by Different Heat-treated Nickel-titanium Instruments: A Micro-computed Tomographic Study
    Filizola de Oliveira, Daniel Jose
    Leoni, Graziela Bianchi
    Goulart, Rafael da Silva
    de Sousa-Neto, Manoel Damiao
    Correa Silva Sousa, Yara Teresinha
    Silva, Ricardo Gariba
    JOURNAL OF ENDODONTICS, 2019, 45 (06) : 768 - 773
  • [43] Influence of Different Heat Treatments on Torsional and Cyclic Fatigue Resistance of Nickel-Titanium Rotary Files: A Comparative Study
    Gambarini, Gianluca
    Cicconetti, Andrea
    Di Nardo, Dario
    Miccoli, Gabriele
    Zanza, Alessio
    Testarelli, Luca
    Seracchiani, Marco
    APPLIED SCIENCES-BASEL, 2020, 10 (16):
  • [44] Evaluation of dynamic and static torsional resistances of nickel-titanium rotary instruments
    Abu-Tahun, Ibrahim H.
    Ha, Jung-Hong
    Kwak, Sang Won
    Kim, Hyeon-Cheol
    JOURNAL OF DENTAL SCIENCES, 2018, 13 (03) : 207 - 212
  • [45] Effect of Fatigue on Torsional Failure of Nickel-Titanium Controlled Memory Instruments
    Campbell, Les
    Shen, Ya
    Zhou, Hui-min
    Haapasalo, Markus
    JOURNAL OF ENDODONTICS, 2014, 40 (04) : 562 - 565
  • [46] A Multimethod Assessment of a New Customized Heat-Treated Nickel-Titanium Rotary File System
    Silva, Emmanuel J. N. L.
    Martins, Jorge N. R.
    Ajuz, Natasha C.
    Antunes, Henrique S.
    Vieira, Victor T. L.
    Fernandes, Francisco M. Braz
    Belladonna, Felipe G.
    Versiani, Marco A.
    MATERIALS, 2022, 15 (15)
  • [47] ULTRASONIC FATIGUE BEHAVIOR OF NICKEL-TITANIUM ENDODONTIC INSTRUMENTS UNDER FLEXURAL CONDITIONS
    Zhou, H. M.
    Li, Q. F.
    Li, L.
    Zheng, Y. F.
    SMST-2007: PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON SHAPE MEMORY AND SUPERELASTIC TECHNOLOGIES, 2008, : 473 - +
  • [48] Fracture resistances of heat-treated nickel-titanium files used for minimally invasive instrumentation
    Kang, Yoo Jung
    Kwak, Sang Won
    Ha, Jung-Hong
    Gambarini, Gianluca
    Kim, Hyeon-Cheol
    BMC ORAL HEALTH, 2025, 25 (01):
  • [49] Influence of shaft length on torsional behavior of endodontic nickel-titanium instruments
    Gambarini, Gianluca
    Seracchiani, Marco
    Zanza, Alessio
    Miccoli, Gabriele
    Del Giudice, Andrea
    Testarelli, Luca
    ODONTOLOGY, 2021, 109 (03) : 568 - 573
  • [50] Physico-mechanical properties of two different heat treated Nickel-Titanium instruments: in-vitro study
    Schiavon, Riccardo
    Pelliccioni, Gian Andrea
    Zamparini, Fausto
    Bolelli, Giovanni
    Spinelli, Andrea
    Pirani, Chiara
    Iacono, Francesco
    Gandolfi, Maria Giovanna
    Prati, Carlo
    Generali, Luigi