Mechanical resistance of screwless morse taper and screw-retained implant-abutment connections

被引:33
|
作者
Ugurel, Cihan Sadi [1 ]
Steiner, Martin [2 ]
Isik-Ozkol, Gul [1 ]
Kutay, Omer [1 ]
Kern, Matthias [2 ]
机构
[1] Istanbul Univ, Fac Dent, Dept Prosthodont, TR-34093 Istanbul, Turkey
[2] Univ Kiel, Sch Dent, Dept Prosthodont Propaedeut & Dent Mat, Kiel, Germany
关键词
dental implant-abutment connection; mechanical strength; morse taper; PROSPECTIVE CLINICAL-EVALUATION; SINGLE-TOOTH REPLACEMENTS; FIXED PARTIAL DENTURES; LEAST; 5; YEARS; COMPLICATION RATES; OBSERVATION PERIOD; IN-VITRO; BRANEMARK IMPLANTS; INTERFACE DESIGN; DENTAL IMPLANTS;
D O I
10.1111/clr.12303
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
ObjectiveThe screwless morse taper (SMT) implant-abutment connection is an alternative to conventional external or internal screw-retained (ISR) connections. The aim of this in vitro study was to evaluate mechanical resistance of the SMT connection and to compare it with three different ISR connections. Materials and MethodsFour implant systems were tested in this study: SMT system; Tasarimmed Octo (Istanbul, Turkey), ISR systems; Straumann Bone Level (Basel, Switzerland), Biohorizons Internal (Birmingham, AL, USA), and Dentsply Friadent Xive (Mannheim, Germany). Overall, 64 specimens with stylized single crowns were prepared: 32 for dynamic loading (DL) and 32 for static loading (SL). DL was carried out using a chewing simulator with 120N at 1.75Hz for 1.2x10(6) cycles, and SL was performed with a universal testing machine at a crosshead speed of 2mm/min with an angulation of 30 degrees. Cycles until failure from DL and fracture/bending loads at SL were recorded. Statistical analyses were made with Dunn's multiple comparison. ResultsMedian cycles until failure in DL were as follows: Octo 86,354 (24,810-153,875), Straumann 1,200,000 (1,156,618-1,200,000), Biohorizons 539,719 (437,224-858,732), Xive 139,411 (139,411-139,411). Median fracture/bending loads in Newton at SL were as follows: Octo 429.6 (404.5-482.7), Straumann 574.6 (544.6-629.9), Biohorizons 548.7 (532.9-567.3), Xive 431.5 (412.5-520.5). There were significant differences between the implant systems under both loading conditions (P0.05) revealing that the Octo implant system's SMT connection showed significantly lower cycles to failures and lower fracture/bending loads compared with the ISR connections of the Straumann and Biohorizons implant systems. However, there was no significant difference compared with the Xive implant system. ConclusionThe mechanical resistance of the screwless morse taper implant system is lower than that of the ISR implant systems that might result in more frequent clinical complications.
引用
收藏
页码:137 / 142
页数:6
相关论文
共 50 条
  • [31] Evaluation of contiguous implants with cement-retained implant-abutment connections. A minipig study
    Martins De Barros, Raquel Rezende
    Novaes Junior, Arthur Belem
    Papalexiou, Vula
    Luczyszyn, Sonia Mara
    Papalexiou Neto, Stylianos Nicolas
    Garcez Ramos, Cassiana Maria
    Goncalves De Almeida, Adriana Luisa
    Piattelli, Adriano
    Assenza, Bartolomeo
    JOURNAL OF OSSEOINTEGRATION, 2014, 6 (01) : 3 - 10
  • [32] Mechanical integrity of cement- and screw-retained zirconium-lithium silicate glass-ceramic crowns to Morse taper implants
    Vahey, Brendan R.
    Sordi, Mariane B.
    Stanley, Kyle
    Magini, Ricardo S.
    de Oliveira, Antonio P. Novaes
    Fredel, Marcio C.
    Henriques, Bruno
    Souza, Julio C. M.
    JOURNAL OF PROSTHETIC DENTISTRY, 2018, 120 (05): : 721 - 731
  • [33] Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit
    Velez, Jorge
    Pelaez, Jesus
    Lopez-Suarez, Carlos
    Agustin-Panadero, Ruben
    Tobar, Celia
    Suarez, Maria J.
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (08) : 1 - 13
  • [34] The influence of implant-abutment connection on the screw loosening and microleakage
    Tsuruta, Katsuhiro
    Ayukawa, Yasunori
    Matsuzaki, Tatsuya
    Kihara, Masafumi
    Koyano, Kiyoshi
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2018, 4
  • [35] A new design of a multifunctional abutment to morse taper implant connection: Experimental mechanical analysis
    Gehrke, Sergio Alexandre
    Dedavid, Berenice Anina
    Fernandes, Gustavo Vicentis de Oliveira
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2021, 116
  • [36] Assessment of lipopolysaccharide microleakage at conical implant-abutment connections
    Sönke Harder
    Elgar Susanne Quabius
    Lars Ossenkop
    Matthias Kern
    Clinical Oral Investigations, 2012, 16 : 1377 - 1384
  • [37] Fatigue induced changes in conical implant-abutment connections
    Blum, Kai
    Wiest, Wolfram
    Fella, Christian
    Balles, Andreas
    Dittmann, Jonas
    Rack, Alexander
    Maier, Dominik
    Thomann, Ralf
    Spies, Benedikt Christopher
    Kohal, Ralf Joachim
    Zabler, Simon
    Nelson, Katja
    DENTAL MATERIALS, 2015, 31 (11) : 1415 - 1426
  • [38] Microbial contamination of the implant-abutment connections: Review of the literature
    Baixe, S.
    Tenenbaum, H.
    Etienne, O.
    REVUE DE STOMATOLOGIE DE CHIRURGIE MAXILLO-FACIALE ET DE CHIRURGIE ORALE, 2016, 117 (01) : 20 - 25
  • [39] Assessment of lipopolysaccharide microleakage at conical implant-abutment connections
    Harder, Soenke
    Quabius, Elgar Susanne
    Ossenkop, Lars
    Kern, Matthias
    CLINICAL ORAL INVESTIGATIONS, 2012, 16 (05) : 1377 - 1384
  • [40] The influence of stiffness of implant-abutment connection on load-deflection ratios of a screw-retained stiff cantilever beam. 3-D measurements in vitro
    Cassel, Bjorn
    Lundgren, Dan
    Karlsson, Dan
    CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 (11) : 1251 - 1256