Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover During Negative Pressure Application

被引:2
|
作者
Brambilla, Enrico [1 ]
Ferguson, Tanner J. [2 ]
Chu, Nathan [1 ]
Ammar, David [3 ]
Yoo, Paul [1 ]
机构
[1] Equinox Ophthalm Inc, Newport Beach, CA USA
[2] Vance Thompson Vis, 3101 W 57th St, Sioux Falls, SD 57108 USA
[3] Lions Eye Inst Transplant & Res, Tampa, FL USA
来源
CLINICAL OPHTHALMOLOGY | 2022年 / 16卷
关键词
tonometry; intraocular pressure; TOP; multi-pressure dial; MPD; ophthalmic techniques; diagnostics; GLAUCOMA PROGRESSION;
D O I
10.2147/OPTH.S359605
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: This is a 2-part study to investigate the agreement between pneumatonometry and direct pressure transducer intraocular pressure ( IOP) measurements in a perfusion organ culture (POC) model where (1) the perfusion fluid column is open to atmospheric pressure, holding IOP constant to permit evaluation of the impact of negative pressure (NP) on IOP measurements, and (2) the perfusion fluid column is a closed system, allowing IOP to vary with NP application. Methods: The first part incorporated a fluid column open to atmospheric pressure, maintaining IOP constant to permit evaluation of the effect of applied NP on IOP measurement accuracy. In the second part, the POC column was closed, allowing IOP to vary with NP application and permit evaluation of agreement between pneumatonometry and pressure transducer measurements. In each part, four perfused tissues were used in thirteen paired pre-set IOP (10, 20, 25, 30 mmHg) and NP (0, 5, 10, 15, 20 mmHg) combinations, resulting in a total of 1040 paired measurements (520 per study). The difference in IOP measurements (Delta IOP = Excursion tonometry - pressure transducer) was calculated at each paired configuration. Results: During the first part, the mean Delta IOP was -0.7 +/- 1.6 mmHg across all measurements. During the second part, the mean Delta IOP across all measurements was +0.7 +/- 1.4 mmHg. At NP settings of -5, -10, -15, and -20 mmHg, across all pre-set IOPs, the mean TOP reduction via Excursion tonometry was 3.1 +/- 0.3, 5.6 +/- 1.3, 8.5 +/- 1.7 and 11.2 +/- 1.8 mmHg, respectively. Conclusion: Measurement of IOP via Excursion tonometry yields results within the accuracy range of the pneumatonometry device (per manufacturer) and is minimally impacted by NP application. The IOP-lowering results are consistent with previous studies and further support the effectiveness of the Multi-Pressure Dial in lowering IOP relative to atmospheric pressure.
引用
收藏
页码:1289 / 1300
页数:12
相关论文
共 50 条
  • [1] Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover
    Tanner J. Ferguson
    Catherine G. Knier
    Uttio Roy Chowdhury
    Kjerseten J. Monson
    Michael Greenwood
    Russell J. Swan
    Richard Gorham
    John P. Berdahl
    Michael P. Fautsch
    Ophthalmology and Therapy, 2020, 9 : 127 - 137
  • [2] Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover
    Ferguson, Tanner J.
    Knier, Catherine G.
    Chowdhury, Uttio Roy
    Monson, Kjerseten J.
    Greenwood, Michael
    Swan, Russell J.
    Gorham, Richard
    Berdahl, John P.
    Fautsch, Michael P.
    OPHTHALMOLOGY AND THERAPY, 2020, 9 (01) : 127 - 137
  • [3] Rebound Tonometer: Ideal Tonometer for Measurement of Accurate Intraocular Pressure
    Suman, Suwarna
    Agrawal, Ajai
    Pal, Virendra K.
    Pratap, Vir B.
    JOURNAL OF GLAUCOMA, 2014, 23 (09) : 633 - 637
  • [4] Measurement of intraocular pressure with a modern noncontact tonometer
    Koçak, I
    Orgül, S
    Saruhan, A
    Haefliger, I
    Hendrickson, P
    Flammer, J
    OPHTHALMOLOGICA, 1998, 212 (02) : 81 - 87
  • [5] Intraocular pressure during scleral lens wear using pneumatonometry
    Nau, Cherie B.
    Harthan, Jennifer
    Nau, Amy Catherine
    Shorter, Ellen
    Schornack, Muriel
    Fogt, Jennifer Swingle
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [6] Measurement of Intraocular Pressure with a Dynamic Tonometer and Presentation of Measurement Results
    M. Yu. Danichenko
    A. A. Mikheev
    Biomedical Engineering, 2022, 55 : 342 - 345
  • [7] Measurement of Intraocular Pressure with a Dynamic Tonometer and Presentation of Measurement Results
    Danichenko, M. Yu
    Mikheev, A. A.
    BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2022, 55 (05): : 342 - 345
  • [8] Noninvasive measurement of rodent intraocular pressure with a rebound tonometer
    Wang, WH
    Millar, JC
    Pang, IH
    Wax, MB
    Clark, AF
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (12) : 4617 - 4621
  • [9] Accuracy of Intraocular Pressure Measurement With the Icare Tonometer in Children
    Chan, Wai Hong
    Lloyd, I. Chris
    Symes, Richard J.
    Ashworth, Jane L.
    Cosgrove, Edel
    Pilling, Rachel
    Biswas, Susmito
    ASIA-PACIFIC JOURNAL OF OPHTHALMOLOGY, 2015, 4 (06): : 357 - 359
  • [10] MEASUREMENT OF INTRAOCULAR-PRESSURE WITH A PNEUMATIC APPLANATION TONOMETER
    WEST, CE
    CAPELLA, JA
    KAUFMAN, HE
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1972, 74 (03) : 505 - &