High-accuracy detector calibration in the 3-1500 eV spectral range at the PTB radiometry laboratory

被引:7
|
作者
Scholze, F [1 ]
Henneken, H [1 ]
Kuschnerus, P [1 ]
Rabus, H [1 ]
Richter, M [1 ]
Ulm, G [1 ]
机构
[1] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
来源
关键词
photodiodes; detector calibration; soft X-rays; primary detector standards;
D O I
10.1107/S0909049597014076
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
State-of-the-art detector calibration in the UV/VUV and soft Xray spectral ranges at the Physikalisch-Technische Bundesanstalt (PTB) is based on the primary detector standard SYRES, a cryogenic electrical substitution radiometer capable of measuring radiant power of a few mu W. At the PTB radiometry laboratory at the synchrotron radiation facility BESSY, two dedicated beamlines are operated, providing monochromatic radiation of high spectral purity, high radiant power and tunable photon energy in the 3-1500eV range. The spectral responsivity of detectors, e.g. photodiodes, can be measured with a relative uncertainty of about 1% by direct comparison with SYRES, as will be demonstrated for PtSi/Si and GaAsP/Au Schottky and silicon n-on-p photodiodes. The calibration of photon-counting detectors traceable to SYRES can by accomplished by exploiting the unique capability to scale the spectral photon flux over several orders of magnitude by changing the stored electron current. Calibrations of CCDs and photomultipliers are presented as examples.
引用
收藏
页码:866 / 868
页数:3
相关论文
共 50 条
  • [31] A high-accuracy electronic pressure transducer with a wide temperature range and single iteration temperature calibration
    Burke, M.J.
    WSEAS Transactions on Circuits and Systems, 2010, 9 (09): : 577 - 586
  • [32] Calibration of a pyroelectric detector at 10.6 μm with the National Institute of Standards and Technology high-accuracy cryogenic radiometer
    Gentile, T.R.
    Houston, J.M.
    Eppeldauer, G.
    Migdall, A.L.
    Cromer, C.L.
    Applied Optics, 1997, 36 (16):
  • [33] High-Accuracy Calibration for a Multiview Microscopic 3-D Measurement System
    Dai, Rong
    Li, Wenpan
    Tang, Xueyan
    Li, Yucheng
    Liu, Yun-Hui
    IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2024, 29 (04) : 2475 - 2486
  • [34] High-speed, high-accuracy large range 3D measurement
    An, Yatong
    Zhang, Song
    DIMENSIONAL OPTICAL METROLOGY AND INSPECTION FOR PRACTICAL APPLICATIONS VI, 2017, 10220
  • [35] High-accuracy and rapid azimuth calibration for polarizing elements in ellipsometry by differential spectral analysis on the ellipse azimuth
    Tu, Huatian
    Zheng, Yuxiang
    Shan, Yao
    Chen, Yao
    Zhang, Haotian
    Zhang, Rongjun
    Wang, Songyou
    Lee, YoungPak
    Chen, Liangyao
    APPLIED OPTICS, 2021, 60 (06) : 1609 - 1614
  • [36] High-accuracy real time calibration of tunable Fabry-Pérot filter on large range
    Qiao, Xueguang
    Wang, Yu
    Fu, Haiwei
    Zhao, Dazhuang
    Wang, Wei
    Zhang, Jing
    Guangxue Xuebao/Acta Optica Sinica, 2008, 28 (05): : 852 - 855
  • [37] Calibration of a pyroelectric detector at 10.6 mu m with the National Institute of Standards and Technology high-accuracy cryogenic radiometer
    Gentile, TR
    Houston, JM
    Eppeldauer, G
    Migdall, AL
    Cromer, CL
    APPLIED OPTICS, 1997, 36 (16) : 3614 - 3621
  • [38] High-Accuracy Patternless Calibration of Multiple 3-D LiDARs for Autonomous Vehicles
    de Miguel, Miguel Angel
    Guindel, Carlos
    Al-Kaff, Abdulla
    Garcia, Fernando
    IEEE SENSORS JOURNAL, 2023, 23 (11) : 12200 - 12208
  • [39] A high-accuracy online calibration method for structured light 3D measurement
    Xing, Chao
    Huang, Junhui
    Wang, Zhao
    Duan, Qiongqiong
    Li, Zijun
    Qi, Miaowei
    MEASUREMENT, 2023, 210
  • [40] A VIPA spectrograph with ultra-high spectral resolution, ultra-high wavelength calibration accuracy and wide spectral range
    Zhang, Zhongnan
    Wang, Yutao
    Zhu, Xiaoming
    Lin, Dong
    He, Jinping
    JOURNAL OF OPTICS, 2024, 26 (12)