Effect of Scala Tympani Height on Insertion Depth of Straight Cochlear Implant Electrodes

被引:17
|
作者
Morrel, William G. [1 ]
Holder, Jourdan T. [2 ]
Dawant, Benoit M. [3 ]
Noble, Jack H. [3 ]
Labadie, Robert F. [1 ,4 ]
机构
[1] Vanderbilt Univ, Dept Otolaryngol, Med Ctr, 7209 Med Ctr East,South Tower, Nashville, TN 37232 USA
[2] Vanderbilt Univ, Med Ctr, Dept Hearing & Speech Sci, Nashville, TN 37232 USA
[3] Vanderbilt Univ, Dept Elect Engn & Comp Sci, Nashville, TN 37232 USA
[4] Vanderbilt Univ, Dept Biomed Engn, Nashville, TN 37232 USA
关键词
cochlear implantation; scala tympani height; translocation; straight electrodes; HEARING; LENGTH; LOCALIZATION; LOCATION; IMPACT; DIMENSIONS; PLACEMENT; SPEECH;
D O I
10.1177/0194599820904941
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objective Studies suggest lateral wall (LW) scala tympani (ST) height decreases apically, which may limit insertion depth. No studies have investigated the relationship of LW ST height with translocation rate or location. Study Design Retrospective review. Setting Cochlear implant program at tertiary referral center. Subjects and Methods LW ST height was measured in preoperative images for patients with straight electrodes. Scalar location, angle of insertion depth (AID), and translocation depth were measured in postoperative images. Audiologic outcomes were tracked. Results In total, 177 ears were identified with 39 translocations (22%). Median AID was 443 degrees (interquartile range [IQR], 367 degrees-550 degrees). Audiologic outcomes (126 ears) showed a small, significant correlation between consonant-nucleus-consonant (CNC) word score and AID (r = 0.20, P = .027), although correlation was insignificant if translocation occurred (r = 0.11, P = .553). Translocation did not affect CNC score (P = .335). AID was higher for translocated electrodes (503 degrees vs 445 degrees, P = .004). Median translocation depth was 381 degrees (IQR, 222 degrees-399 degrees). Median depth at which a 0.5-mm electrode would not fit within 0.1 mm of LW was 585 degrees (IQR, 405 degrees-585 degrees). Median depth at which a 0.5-mm electrode would displace the basilar membrane by >= 0.1 mm was 585 degrees (IQR, 518 degrees-765 degrees); this was defined as predicted translocation depth (PTD). Translocation rate was 39% for insertions deeper than PTD and 14% for insertions shallower than PTD (P = .008). Conclusion AID and CNC are directly correlated for straight electrodes when not translocated. Translocations generally occur around 380 degrees and are more common with deeper insertions due to decreasing LW ST height. Risk of translocation increases significantly after 580 degrees.
引用
收藏
页码:718 / 724
页数:7
相关论文
共 50 条
  • [1] Effect of Embedded Dexamethasone in Cochlear Implant Array on Insertion Forces in an Artificial Model of Scala Tympani
    Yann Nguyen
    Bernardeschi, Daniele
    Kazmitcheff, Guillaume
    Miroir, Mathieu
    Vauchel, Thomas
    Ferrary, Evelyne
    Sterkers, Olivier
    OTOLOGY & NEUROTOLOGY, 2015, 36 (02) : 354 - 358
  • [2] Morphometric Analysis and Linear Measurements of the Scala Tympani and Implications in Cochlear Implant Electrodes
    Fujiwara, Rance J. T.
    Ishiyama, Gail
    Lopez, Ivan A.
    Ishiyama, Akira
    OTOLOGY & NEUROTOLOGY, 2023, 44 (05) : E343 - E349
  • [3] Bilateral cochlear implant user with a right ear scala vestibuli insertion and left ear scala tympani insertion: case report
    Leonor, A
    Luis, AS
    COCHLEAR IMPLANTS, 2004, 1273 : 455 - 458
  • [4] COCHLEAR PATHOLOGY WITH CHRONICALLY IMPLANTED SCALA TYMPANI ELECTRODES
    LEAKEJONES, PA
    REBSCHER, SJ
    ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1983, 405 (JUN) : 203 - 223
  • [5] SURGICAL APPROACH FOR INSERTION OF MULTICHANNEL ELECTRODES INTO THE SCALA TYMPANI
    MICHELSON, RP
    SCHINDLER, RA
    ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1983, 405 (JUN) : 343 - 347
  • [6] Potential distribution in the scala tympani youth cochlear implant stimulation
    Kral, A
    Mortazavi, D
    Hartmann, R
    Klinke, R
    PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1997, 433 (06): : P174 - P174
  • [7] Scala vestibuli insertion in cochlear implantation: A valuable alternative for cases with obstructed scala tympani
    Kiefer, J
    Weber, A
    Pfennigdorff, T
    von Ilberg, C
    ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 2000, 62 (05): : 251 - 256
  • [8] Durability of Hearing Preservation after Cochlear Implantation with Conventional-Length Electrodes and Scala Tympani Insertion
    Sweeney, Alex D.
    Hunter, Jacob B.
    Carlson, Matthew L.
    Rivas, Alejandro
    Bennett, Marc L.
    Gifford, Rene H.
    Noble, Jack H.
    Haynes, David S.
    Labadie, Robert F.
    Wanna, George B.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2016, 154 (05) : 907 - 913
  • [9] Insertion Depth for Optimized Positioning of Precurved Cochlear Implant Electrodes
    Banalagay, Rueben A.
    Labadie, Robert F.
    Chakravorti, Srijata
    Noble, Jack H.
    OTOLOGY & NEUROTOLOGY, 2020, 41 (08) : 1066 - 1071
  • [10] In vitro modifications of the scala tympani environment and the cochlear implant array surface
    Kontorinis, Georgios
    Scheper, Verena
    Wissel, Kirsten
    Stoever, Timo
    Lenarz, Thomas
    Paasche, Gerrit