Optimization of the Spatial Resolution for the GE Discovery PET/CT 710 by Using NEMA NU 2-2007 Standards

被引:15
|
作者
Yoon, Hyun Jin [1 ]
Jeong, Young Jin [1 ]
Son, Hye Joo [1 ]
Kang, Do-Young [1 ]
Hyun, Kyung-Yae [2 ]
Lee, Min-Kyung [3 ]
机构
[1] Dong A Univ, Dept Nucl Med, Pusan 602715, South Korea
[2] Dong Eui Univ, Dept Clin Lab Sci, Pusan 614714, South Korea
[3] Dong Eui Univ, Dept Dent Hyg, Pusan 614714, South Korea
关键词
Electrical Manufacturers Association (NEMA); Positron emission tomography (PET); Spatial resolution; Iterative reconstruction algorithm; VARIANT SYSTEM MODEL; PERFORMANCE-MEASUREMENTS; ATTENUATION CORRECTION; RECONSTRUCTION;
D O I
10.3938/jkps.66.287
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The spatial resolution in positron emission tomography (PET) is fundamentally limited by the geometry of the detector element, the positron's recombination range with electrons, the acollinearity of the positron, the crystal decoding error, the penetration into the detector ring, and the reconstruction algorithms. In this paper, optimized parameters are suggested to produce high-resolution PET images by using an iterative reconstruction algorithm. A phantom with three point sources structured with three capillary tubes was prepared with an axial extension of less than 1 mm and was filled with F-18-fluorodeoxyglucose (F-18-FDG) with concentrations above 200 MBq/cc. The performance measures of all the PET images were acquired according to the National Electrical Manufacturers Association (NEMA) NU 2-2007 standards procedures. The parameters for the iterative reconstruction were adjusted around the values recommended by General Electric GE, and the optimized values of the spatial resolution and the full width at half maximum (FWHM) or the full width at tenth of maximum (FWTM) values were found for the best PET resolution. The axial and the transverse spatial resolutions, according to the filtered back-projection (FBP) at 1 cm off-axis, were 4.81 and 4.48 mm, respectively. The axial and the transaxial spatial resolutions at 10 cm off-axis were 5.63 mm and 5.08 mm, respectively, and the trans-axial resolution at 10 cm was evaluated as the average of the radial and the tangential measurements. The recommended optimized parameters of the spatial resolution according to the NEMA phantom for the number of subsets, the number of iterations, and the Gaussian post-filter are 12, 3, and 3 mm for the iterative reconstruction VUE Point HD without the SharpIR algorithm (HD), and 12, 12, and 5.2 mm with SharpIR (HD.S), respectively, according to the Advantage Workstation Volume Share 5 (AW4.6). The performance measurements for the GE Discovery PET/CT 710 using the NEMA NU 2-2007 standards from our results will be helpful in the quantitative analysis of PET scanner images. The spatial resolution was modified more by using an improved algorithm such as HD.S, than by using HD and FBP. The use of the optimized parameters for iterative reconstructions is strongly recommended for qualitative images from the GE Discovery PET/CT 710 scanner.
引用
收藏
页码:287 / 294
页数:8
相关论文
共 36 条
  • [1] Optimization of the spatial resolution for the GE discovery PET/CT 710 by using NEMA NU 2-2007 standards
    Hyun Jin Yoon
    Young Jin Jeong
    Hye Joo Son
    Do-Young Kang
    Kyung-Yae Hyun
    Min-Kyung Lee
    Journal of the Korean Physical Society, 2015, 66 : 287 - 294
  • [2] Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards
    De Ponti, E.
    Morzenti, S.
    Guerra, L.
    Pasquali, C.
    Arosio, M.
    Bettinardi, V.
    Crespi, A.
    Gilardi, M. C.
    Messa, C.
    MEDICAL PHYSICS, 2011, 38 (02) : 968 - 974
  • [3] NEMA NU 2-2007 performance characteristics of GE Signa integrated PET/MR for different PET isotopes
    Caribe, Paulo R. R. V.
    Koole, M.
    D'Asseler, Yves
    Deller, Timothy W.
    Van Laere, K.
    Vandenberghe, S.
    EJNMMI PHYSICS, 2019, 6 (1)
  • [4] NEMA NU 2-2007 Performance Characteristics of GE Signa Integrated PET/MR: Impact of Using Different PET Isotopes
    Caribe, P.
    Koole, M.
    Vandenberghe, S.
    Deller, T.
    van Laere, K.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S339 - S339
  • [5] GE Signa Integrated PET/MR: NEMA NU 2-2007 Performance Characteristics for 68Ga PET Imaging
    Caribe, P.
    Koole, M.
    Vandenberghe, S.
    Deller, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S507 - S508
  • [6] 基于NEMA NU 2-2007标准测试国产uEXPLORER PET系统性能
    蒋梨
    黄衍超
    李洪生
    韩彦江
    王巧愚
    王静怡
    王全师
    中国医学影像技术, 2021, 37 (11) : 1725 - 1730
  • [7] NEMA NU 2-2007 performance measurements of the Siemens Inveon™ preclinical small animal PET system
    Kemp, Brad J.
    Hruska, Carrie B.
    McFarland, Aaron R.
    Lenox, Mark W.
    Lowe, Val J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (08): : 2359 - 2376
  • [8] NEMA NU 2-2007 Measurements and GATE Monte Carlo Simulations of GE Signa integrated PET/MR for Pure and Non-Pure Positron Emitters
    Caribe, P. R. R. V.
    Koole, M.
    Diogo, A.
    D'Asseler, Y.
    Deller, T.
    Vandenberghe, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S352 - S353
  • [9] NEMA NU 2–2007 performance characteristics of GE Signa integrated PET/MR for different PET isotopes
    Paulo R. R. V. Caribé
    M. Koole
    Yves D’Asseler
    Timothy W. Deller
    K. Van Laere
    S. Vandenberghe
    EJNMMI Physics, 6
  • [10] Evaluation of PET/CT alignment accuracy of the Discovery MI PET/CT according to the NEMA NU2-2018 standard
    Nie, Xingyu
    Schmidtlein, Ross
    Kirov, Assen
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61